Parameter Optimization and Effective Imaging Volume Determination of Helical Scan for a Pinhole Animal SPECT

被引:0
|
作者
Yao, Rutao [1 ]
Deng, Xiao [1 ]
Dai, Tiantian [2 ]
Wei, Qingyang [2 ]
Beaudoin, Jean-Francois
Naaman, Charles
Ma, Tianyu [2 ]
Cadorette, Jules
Lecomte, Roger
机构
[1] SUNY Buffalo, Dept Nucl Med, Buffalo, NY 14217 USA
[2] Tsinghua Univ, Dept Engn Phys, Beijing, Peoples R China
基金
加拿大健康研究院; 加拿大创新基金会;
关键词
SPECT; optimal helical pitch; multiple pinhole; COLLIMATOR;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We developed SPECT imaging capability on a LabPET T animal scanner. The add-on SPECT uses a multiple-pinhole collimator and a step-and-shoot helical scan scheme. The objective of this work was to determine I) the effective imaging volume of a given scan protocol, and 2) the helical scan parameters, that is, the angular and axial step sizes, for achieving uniform high resolution images throughout the imaging volume. Three rotational step sizes, {1 degrees, 2 degrees, 3 degrees}, and three axial step sizes, {0.3 mm, 0.6 mm, 1.2 mm}, were combined to form 9 helical scan protocols. These step sizes were derived based on the reference values analytically calculated using the spatial resolution of the SPECT system and the Nyquist sampling theorem. Two numerical indices, the sampling completeness percentage (SCP), and the directional resolution (DR), were designed to quantify the projection sampling and resolution performance of the helical protocols. Scanning of a digital sphere-grid phantom centered in the SPECT system's FOV was simulated for each of the 9 helical protocols. Aggregated RMS resolution values derived from the image of distributed spheres over the FOV were used to assess the resolution performance of the scan protocols, and for validating the predictive value of the SCP and DR indices, which were calculated over the matching grid. The RMS results showed that the effective image volume of a helical scan where sampling is adequate was defined by the start and end pinhole planes. The results also showed that using a small axial step size (about half the value derived from Nyquist theorem) achieved best imaging resolution, while the angular step size was less critical. The RMS images correlated well with that of SCP and DR. We conclude the following for designing a helical scan protocol with a multiple pinhole collimator. First, the effective imaging volume should be confined between the start and end planes of the pinholes' movement range. Second, using an axial step size that is half of what is calculated from the Nyquist theorem is desirable for optimal resolution performance. In addition, SCP and DR are useful indices for assessing scan protocols analytically.
引用
收藏
页码:3797 / 3803
页数:7
相关论文
共 50 条
  • [1] Multipinhole SPECT helical scan parameters and imaging volume
    Yao, Rutao
    Deng, Xiao
    Wei, Qingyang
    Dai, Tiantian
    Ma, Tianyu
    Lecomte, Roger
    MEDICAL PHYSICS, 2015, 42 (11) : 6599 - 6609
  • [2] Helical pinhole SPECT for small-animal imaging: A method for addressing sampling completeness
    Metzler, SD
    Greer, KL
    Jaszczak, RJ
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) : 1575 - 1583
  • [3] Fast iterative reconstruction for helical pinhole SPECT imaging
    Huang, Po-Chia
    Hsu, Ching-Han
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2015, 26 : S1371 - S1380
  • [4] Small animal imaging with multi-pinhole SPECT
    Nuyts, Johan
    Vunckx, Kathleen
    Defrise, Michel
    Vanhove, Christian
    METHODS, 2009, 48 (02) : 83 - 91
  • [5] A review of small animal imaging planar and pinhole spect Γ camera imaging
    Peremans, K
    Cornelissen, B
    Van den Bossche, B
    Audenaert, K
    Van de Wiele, C
    VETERINARY RADIOLOGY & ULTRASOUND, 2005, 46 (02) : 162 - 170
  • [6] Evaluation of A-SPECT: A desktop pinhole SPECT system for small animal Imaging
    McElroy, DP
    MacDonald, LR
    Beekman, FJ
    Wang, YC
    Patt, BE
    Iwanczyk, JS
    Tsui, BMW
    Hoffman, EJ
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1835 - 1839
  • [7] On the Use of Asymmetrical Pinhole Collimators for Small Animal SPECT Imaging
    Peter, J.
    Ruehle, H.
    Stamm, V.
    Semmler, W.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S254 - S254
  • [8] HIGH-RESOLUTION PINHOLE SPECT FOR SMALL ANIMAL IMAGING
    ERLANDSSON, K
    IVANOVIC, M
    STRAND, SE
    SJOGREN, K
    WEBER, DA
    JOURNAL OF NUCLEAR MEDICINE, 1993, 34 (05) : P9 - P9
  • [9] Finite aperture modeling in small animal pinhole SPECT Imaging
    Huang, PC
    Hsu, CH
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3454 - 3457
  • [10] Optimization of pinhole emission imaging systems for small animal imaging
    Zhou, Lili
    Khurd, Parmeshwar
    Gindi, Gene
    2007 IEEE 33RD ANNUAL NORTHEAST BIOENGINEERING CONFERENCE, 2007, : 94 - +