Comparison of blood pressure measurements of anesthetized dogs obtained noninvasively with a cylindrical blood pressure cuff and an anatomically modified conical blood pressure cuff

被引:0
|
作者
Ramos, Sara J. [1 ]
da Cunha, Anderson F. [1 ]
Domingues, Michelle [1 ]
Shelby, Amanda M. [1 ]
Stout, Rhett W. [2 ]
Acierno, Mark J. [1 ]
机构
[1] Louisiana State Univ, Sch Vet Med, Dept Vet Clin Sci, Baton Rouge, LA 70803 USA
[2] Louisiana State Univ, Sch Vet Med, Div Lab Anim Med, Baton Rouge, LA 70803 USA
关键词
SYSTEMIC HYPERTENSION; AGREEMENT; MONITOR; GUIDELINES; DIAGNOSIS; CATS;
D O I
暂无
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
OBJECTIVE To compare blood pressure measured noninvasively with an oscillometric device that involved use of a novel conical cuff and a traditional cylindrical blood pressure cuff. ANIMALS 17 adult hound-type dogs. PROCEDURES Dogs were anesthetized, and a 20-gauge, 1.5-inch catheter was inserted in the median sacral artery. The catheter was attached to a pressure transducer via fluid-filled noncompliant tubing, and direct blood pressure was recorded with a multifunction monitor. A specially fabricated conical cuff was placed on the antebrachium. Four sets of direct and indirect blood pressure measurements were simultaneously collected every 2 minutes. Four sets of measurements were then obtained by use of a cylindrical cuff. RESULTS The cylindrical cuff met American College of Veterinary Internal Medicine consensus guidelines for validation of indirect blood pressure measurements for mean arterial blood pressure (MAP), systolic arterial blood pressure (SAP), and diastolic arterial blood pressure (DAP). The conical cuff met the consensus guidelines for difference of paired measurements, SD, and percentages of measurements within 10 and 20 mm Hg of the value for the reference method, but it failed a correlation analysis. In addition, although bias for the conical cuff was less than that for the cylindrical cuff for SAP, MAP, and DAP measurements, the limits of agreement for the conical cuff were wider than those for the cylindrical cuff for SAP and MAP measurements. CONCLUSIONS AND CLINICAL RELEVANCE On the basis of results of this study, use of a conical cuff for oscillometric blood pressure measurement cannot be recommended.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 50 条
  • [1] An Experimental Study on the Blood Pressure Cuff as a Transducer for Oscillometric Blood Pressure Measurements
    Bogatu, Laura, I
    Turco, Simona
    Mischi, Massimo
    Muehlsteff, Jens
    Woerlee, Pierre
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2021, 70
  • [2] ACCURACY OF CUFF BLOOD PRESSURE AND SYSTOLIC BLOOD PRESSURE AMPLIFICATION
    Bui, Tan
    Picone, Dean
    Schultz, Martin
    Peng, Xiaoqing
    Black, Andrew
    Dwyer, Nathan
    Thomson, Philip Roberts
    Adams, Heath
    Chen, Chen Huan
    Cheng, Hao Min
    Pucci, Giacomo
    Wang, Jiguang
    Goupil, Remi
    Sharman, James
    JOURNAL OF HYPERTENSION, 2023, 41 : E203 - E203
  • [3] Accuracy of cuff blood pressure and systolic blood pressure amplification
    Bui, Tan V.
    Picone, Dean S.
    Schultz, Martin G.
    Peng, Xiaoqing
    Black, J. Andrew
    Dwyer, Nathan
    Roberts-Thomson, Philip
    Adams, Heath
    Chen, Chen-Huan
    Cheng, Hao-min
    Pucci, Giacomo
    Wang, Jiguang
    Goupil, Remi
    Sharman, James E.
    HYPERTENSION RESEARCH, 2023, 46 (08) : 1961 - 1969
  • [4] Accuracy of cuff blood pressure and systolic blood pressure amplification
    Tan V. Bui
    Dean S. Picone
    Martin G. Schultz
    Xiaoqing Peng
    J. Andrew Black
    Nathan Dwyer
    Philip Roberts-Thomson
    Heath Adams
    Chen-Huan Chen
    Hao-min Cheng
    Giacomo Pucci
    Jiguang Wang
    Remi Goupil
    James E. Sharman
    Hypertension Research, 2023, 46 : 1961 - 1969
  • [5] An improved blood pressure cuff
    Collens, WS
    Boas, LC
    AMERICAN HEART JOURNAL, 1942, 23 : 114 - 115
  • [6] BLOOD PRESSURE CUFF FOR INFANTS
    不详
    ANESTHESIOLOGY, 1956, 17 (02) : 363 - 363
  • [7] Use of a neonatal blood pressure cuff to monitor blood pressure in the adult finger comparison with a standard adult arm cuff
    Khan, SQ
    Wardlaw, JM
    Davenport, R
    Slattery, J
    Lewis, S
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1998, 14 (04) : 233 - 238
  • [8] Use of A Neonatal Blood Pressure Cuff to Monitor Blood Pressure In The Adult Finger – Comparison With A Standard Adult Arm Cuff
    Sohail Q Khan
    Joanna M. Wardlaw
    Richard Davenport
    Jim Slattery
    Stephanie Lewis
    Journal of Clinical Monitoring and Computing, 1998, 14 : 233 - 238
  • [9] Correct placement of blood pressure cuff during blood pressure measurement
    Garcia, Mary Grace Uranza
    Ang, Emily
    Ahmad, Noor Melati
    Lim, Chi Ching
    INTERNATIONAL JOURNAL OF EVIDENCE-BASED HEALTHCARE, 2012, 10 (03): : 191 - 196
  • [10] LENGTH OF BLOOD-PRESSURE CUFF
    BAILIE, MD
    CANADIAN MEDICAL ASSOCIATION JOURNAL, 1976, 114 (06) : 496 - 496