Atomic absorption spectrometric determination of trace amount of nickel in alloys and biological samples after preconcentration with the ion pair of 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol and ammonium tetraphenylborate on microcrystalline naphthalene or by column method

被引:0
|
作者
Taher, MA [1 ]
机构
[1] Shahid Bahonar Univ, Dept Chem, Kerman, Iran
关键词
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Nickel is quantitatively retained on 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (5-Br-PADAP)-ammonium tetraphenylborate with microcrystalline naphthalene or by a column method in the pH range 9.0-10.5 from a large volume of aqueous solutions of various samples. After filtration, the solid mass consisting of the nickel complex and naphthalene was dissolved with 5 ml of dimethylformamide and the metal was determined by atomic absorption spectrometry. The nickel complex can alternatively be quantitatively adsorbed on ammonium tetraphenylborate-naphthalene adsorbent packed in a column and determined similarly. The calibration curve is linear at 0.30-18 ppm in dimethylformamide solution. Eight replicate determinations of 5.0 ppm of nickel gave a mean absorbance of 0.096 with a relative standard deviation of 1.5%. The sensitivity for 1% absorption was 0.23 mu g ml(-1). The interference of a large number of anions and cations has been studied and the optimized conditions developed were utilized for the trace determination of nickel in various standard samples.
引用
收藏
页码:573 / 576
页数:4
相关论文
共 50 条