Effect of computed tomography number-relative electron density conversion curve on the calculation of radiotherapy dose and evaluation of Monaco radiotherapy treatment planning system

被引:10
|
作者
Hasani, Mohsen [1 ]
Farhood, Bagher [2 ]
Ghorbani, Mahdi [3 ]
Naderi, Hamideh [4 ]
Saadatmand, Sepideh [4 ]
Zandi, Saeed Karimkhani [4 ]
Knaup, Courtney [5 ]
机构
[1] Univ Tehran Med Sci, Canc Inst, Canc Res Ctr, Dept Radiotherapy Phys, Tehran, Iran
[2] Kashan Univ Med Sci, Fac Paramed Sci, Dept Med Phys & Radiol, Kashan, Iran
[3] Shahid Beheshti Univ Med Sci, Canc Res Ctr, Tehran, Iran
[4] Qom Univ Med Sci, Canc Inst, Dept Radiotherapy Phys, Qom, Iran
[5] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
关键词
CT-RED curve; Radiotherapy; Treatment planning system; Audit; Monaco; RADIATION-THERAPY; QUALITY-ASSURANCE; CT-NUMBER; ACCURACY; PHOTON; PHANTOM; PERFORMANCE; CALIBRATION; PARAMETERS; IMPACT;
D O I
10.1007/s13246-019-00745-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The accuracy of a computed tomography (CT)-relative electron density (RED) curve may have an indirect impact on the accuracy of dose calculation by a treatment planning system (TPS). This effect has not been previously quantified for input of different CT-RED curves from different CT-scan units in the Monaco TPS. This study aims to evaluate the effect of CT-RED curve on the dose calculation by the Monaco radiotherapy TPS. Four CT images of the CIRS phantom were obtained by different CT scanners. The accuracy of the dose calculation in the three algorithms of the Monaco TPS (Monte Carlo, collapse cone, and pencil beam) is also evaluated based on TECDOC 1583. The CT-RED curves from the CT scanners were transferred to the Monaco TPS to audit the different algorithms of the TPS. The dose values were measured with an ionization chamber in the CIRS phantom. Then, the dose values were calculated by the Monaco algorithms in the corresponding points. For the Monaco TPS and based on TECDOC 1583, the accuracy of the dose calculation in all the three algorithms is within the agreement criteria for most of the points evaluated. For low dose regions, the differences between the calculated and measured dose values are higher than the agreement criteria in a number of points. For the majority of the points, the algorithms underestimate the calculated dose values. It was also found that the use of different CT-RED curves can lead to minor discrepancies in the dose calculation by the Monaco TPS, especially in low dose regions. However, it appears that these differences are not clinically significant in most of the cases.
引用
收藏
页码:489 / 502
页数:14
相关论文
共 50 条
  • [1] Effect of computed tomography number-relative electron density conversion curve on the calculation of radiotherapy dose and evaluation of Monaco radiotherapy treatment planning system
    Mohsen Hasani
    Bagher Farhood
    Mahdi Ghorbani
    Hamideh Naderi
    Sepideh Saadatmand
    Saeed Karimkhani Zandi
    Courtney Knaup
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 489 - 502
  • [2] The influence of different kVs and phantoms on computed tomography number to relative electron density calibration curve for radiotherapy dose calculation
    Jaafar, Ahmed Mousa
    Elsayed, Hussein
    Khalil, Magdy M.
    Yaseen, Mohamed Nabil
    Alshewered, Ahmed
    Ammar, Hany
    PRECISION RADIATION ONCOLOGY, 2022, 6 (04): : 289 - 297
  • [3] The Impact of Computed Tomography (CT) Reconstruction Kernels On Radiotherapy Treatment Planning Dose Calculation
    Vergalasova, I.
    McKenna, M.
    Yue, N.
    Reyhan, M.
    MEDICAL PHYSICS, 2019, 46 (06) : E397 - E397
  • [4] CALCULATION OF ELECTRON DOSE DISTRIBUTION FOR RADIOTHERAPY TREATMENT PLANNING
    KAWACHI, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (04): : 571 - 577
  • [5] Applications of simulator computed tomography number for photon dose calculations during radiotherapy treatment planning
    Chu, JCH
    Ni, B
    Kriz, R
    Saxena, VA
    RADIOTHERAPY AND ONCOLOGY, 2000, 55 (01) : 65 - 73
  • [6] EVALUATION OF THE ROLE OF COMPUTED-TOMOGRAPHY IN RADIOTHERAPY TREATMENT PLANNING
    ADAM, EJ
    BERRY, RJ
    CLITHEROW, S
    BEDFORD, A
    CLINICAL RADIOLOGY, 1984, 35 (02) : 147 - 150
  • [7] Evaluation of a cone-beam computed tomography system calibrated for accurate radiotherapy dose calculation
    Bogowicz, Marta
    Lustermans, Didier
    Taasti, Vicki Trier
    Hazelaar, Colien
    Verhaegen, Frank
    Fonseca, Gabriel Paiva
    van Elmpt, Wouter
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 29
  • [8] Relative electron density calibration of CT scanners for radiotherapy treatment planning
    Thomas, SJ
    BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (860): : 781 - 786
  • [9] Relative electron density determination for radiotherapy and proton therapy treatment planning
    Van Abbema, J. K.
    Van Goethem, M.
    Greuter, M. J. W.
    Van der Schaaf, A.
    Brandenburg, S.
    Van der Graaf, E. R.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S444 - S445
  • [10] The dependence of computed tomography number to relative electron density conversion on phantom geometry and its impact on planned dose
    Emma K. Inness
    Vaughan Moutrie
    Paul H. Charles
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 385 - 391