Lensless, reflection-based dark-field microscopy (RDFM) on a CMOS chip

被引:6
|
作者
Imanbekova, Meruyert [1 ]
Perumal, Ayyappasamy Sudalaiyadum [1 ]
Kheireddine, Sara [1 ]
Nicolau, Dan, V [1 ]
Wachsmann-Hogiu, Sebastian [1 ]
机构
[1] McGill Univ, Dept Bioengn, Montreal, PQ H3A 0E9, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
CELL-SIZE; HOLOGRAPHY; COMPACT;
D O I
10.1364/BOE.394615
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We present for the first time a lens-free, oblique illumination imaging platform for on-sensor dark- field microscopy and shadow-based 3D object measurements. It consists of an LED point source that illuminates a 5-megapixel, 1.4 mu m pixel size, back-illuminated CMOS sensor at angles between 0 degrees and 90 degrees. Analytes (polystyrene beads, microorganisms, and cells) were placed and imaged directly onto the sensor. The spatial resolution of this imaging system is limited by the pixel size (similar to 1.4 mu m) over the whole area of the sensor (3.6x2.73 mm). We demonstrated two imaging modalities: (i) shadow imaging for estimation of 3D object dimensions (on polystyrene beads and microorganisms) when the illumination angle is between 0 degrees and 85 degrees, and (ii) dark-field imaging, at >85 degrees illumination angles. In dark-field mode, a 3-4 times drop in background intensity and contrast reversal similar to traditional dark-field imaging was observed, due to larger reflection intensities at those angles. With this modality, we were able to detect and analyze morphological features of bacteria and single-celled algae clusters. (C) 2020 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
引用
收藏
页码:4942 / 4959
页数:18
相关论文
共 50 条
  • [1] Numerical autofocusing in Gabor lensless microscopy employing computational dark-field imaging
    Trusiak, M.
    Rogalski, M.
    Zdankowski, P.
    Picazo-Bueno, Jose-Angel
    Mico, Vicente
    HOLOGRAPHY: ADVANCES AND MODERN TRENDS VII, 2021, 11774
  • [2] In vivo dark-field reflection-mode photoacoustic microscopy
    Maslov, K
    Stoica, G
    Wang, LHV
    OPTICS LETTERS, 2005, 30 (06) : 625 - 627
  • [3] Dark-field Brillouin microscopy
    Antonacci, Giuseppe
    OPTICS LETTERS, 2017, 42 (07) : 1432 - 1435
  • [4] DARK-FIELD ELECTRON MICROSCOPY
    HALL, CE
    ANALYTICAL CHEMISTRY, 1948, 20 (01) : 92 - 92
  • [5] Nonlinear Dark-Field Microscopy
    Harutyunyan, Hayk
    Palomba, Stefano
    Renger, Jan
    Quidant, Romain
    Novotny, Lukas
    NANO LETTERS, 2010, 10 (12) : 5076 - 5079
  • [6] Research on Detecting Plankton Based on Dark-field Microscopy
    Guo, Buyu
    Yu, Jia
    Liu, Huiping
    Liu, Xiaoying
    Hou, Rui
    OPTICAL DESIGN AND TESTING VII, 2016, 10021
  • [7] Reflection-Based Short Pulse Generation in CMOS
    Khoeini, Farzad
    Hadidian, Bahareh
    Zhang, Keshu
    Afshari, Ehsan
    IEEE SOLID-STATE CIRCUITS LETTERS, 2020, 3 : 318 - 321
  • [8] DARK-FIELD MICROSCOPY AND PATIENT EDUCATION
    COLE, AM
    NEWCOMB, GM
    NIXON, KC
    AUSTRALIAN DENTAL JOURNAL, 1984, 29 (06) : 394 - 397
  • [9] DARK-FIELD SCANNING OPTICAL MICROSCOPY
    WILSON, T
    HAMILTON, DK
    OPTIK, 1985, 71 (01): : 23 - 26
  • [10] DARK-FIELD MICROSCOPY OF DENTAL PLAQUE
    NEWMAN, HH