Mono-Energy Coronary Angiography with a Compact Synchrotron Source

被引:25
|
作者
Eggl, Elena [1 ,2 ]
Mechlem, Korbinian [1 ,2 ,3 ]
Braig, Eva [1 ,2 ,3 ]
Kulpe, Stephanie [1 ,2 ]
Dierolf, Martin [1 ,2 ]
Guenther, Benedikt [1 ,2 ,4 ]
Achterhold, Klaus [1 ,2 ]
Herzen, Julia [1 ,2 ]
Gleich, Bernhard [2 ]
Rummeny, Ernst [3 ]
Noel, Peter B. [1 ,3 ]
Pfeiffer, Franz [1 ,2 ,3 ]
Muenzel, Daniela [3 ]
机构
[1] Tech Univ Munich, Dept Phys, Chair Biomed Phys, James Franck Str 1, D-85748 Garching, Germany
[2] Tech Univ Munich, Munich Sch Bioengn, Boltzmannstr 11, D-85748 Garching, Germany
[3] Tech Univ Munich, Klinikum Rechts Isar, Dept Diagnost & Intervent Radiol, Ismaninger Str 22, D-81675 Munich, Germany
[4] Max Planck Inst Quantum Opt, Hans Kopfermann Str 1, D-85748 Garching, Germany
来源
SCIENTIFIC REPORTS | 2017年 / 7卷
基金
欧洲研究理事会;
关键词
GADOLINIUM-BASED CONTRAST; MEDIA;
D O I
10.1038/srep42211
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
X-ray coronary angiography is an invaluable tool for the diagnosis of coronary artery disease. However, the use of iodine-based contrast media can be contraindicated for patients who present with chronic renal insufficiency or with severe iodine allergy. These patients could benefit from a reduced contrast agent concentration, possibly achieved through application of a mono-energetic x-ray beam. While large-scale synchrotrons are impractical for daily clinical use, the technology of compact synchrotron sources strongly advanced during the last decade. Here we present a quantitative analysis of the benefits a compact synchrotron source can offer in coronary angiography. Simulated projection data from quasi-mono-energetic and conventional x-ray tube spectra is used for a CNR comparison. Results show that compact synchrotron spectra would allow for a significant reduction of contrast media. Experimentally, we demonstrate the feasibility of coronary angiography at the Munich Compact Light Source, the first commercial installation of a compact synchrotron source.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Mono-Energy Coronary Angiography with a Compact Synchrotron Source
    Elena Eggl
    Korbinian Mechlem
    Eva Braig
    Stephanie Kulpe
    Martin Dierolf
    Benedikt Günther
    Klaus Achterhold
    Julia Herzen
    Bernhard Gleich
    Ernst Rummeny
    Peter B. Noёl
    Franz Pfeiffer
    Daniela Muenzel
    Scientific Reports, 7
  • [2] Mono-Energy Coronary Angiography with a Compact Light Source
    Eggl, Elena
    Meehlem, Korbinian
    Braig, Eva
    Kulpe, Stephanie
    Dierolf, Martin
    Guenther, Benedikt
    Achterhold, Klaus
    Herzen, Julia
    Gleich, Bernhard
    Rummeny, Ernst
    Noel, Peter B.
    Pfeiffer, Franz
    Muenzel, Daniela
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [3] K-edge subtraction imaging for coronary angiography with a compact synchrotron X-ray source
    Kulpe, Stephanie
    Dierolf, Martin
    Braig, Eva
    Guenther, Benedikt
    Achterhold, Klaus
    Gleich, Bernhard
    Herzen, Julia
    Rummeny, Ernst
    Pfeiffer, Franz
    Pfeiffer, Daniela
    PLOS ONE, 2018, 13 (12):
  • [4] K-edge subtraction imaging for angiography at a compact synchrotron source
    Kulpe, Stephanie
    Dierolf, Martin
    Braig, Eva-Maria
    Guenther, Benedikt
    Achterhold, Klaus
    Gleich, Bernhard
    Herzen, Julia
    Rummeny, Ernst
    Pfeiffer, Franz
    Pfeiffer, Daniela
    ADVANCES IN LABORATORY-BASED X-RAY SOURCES, OPTICS, AND APPLICATIONS VII, 2019, 11110
  • [5] Dual energy CT in acute appendicitis: value of low mono-energy
    Lev-Cohain, Naama
    Sosna, Jacob
    Meir, Yuval
    Dar, Gili
    Shussman, Noam
    Leichter, Isaac
    Caplan, Nadia
    Goldberg, S. Nahum
    CLINICAL IMAGING, 2021, 77 : 213 - 218
  • [6] Mono-energy neutron testing of Single Event Effects
    Blomgren, J.
    Pomp, S.
    Bourselier, J. -C.
    Osterlund, M.
    Prokofiev, A.
    Koning, A.
    INTERNATIONAL CONFERENCE ON NUCLEAR DATA FOR SCIENCE AND TECHNOLOGY, VOL 2, PROCEEDINGS, 2008, : 1287 - +
  • [7] Compact Undulator for Cornell High Energy Synchrotron Source
    Temnykh, Alexander
    Kobela, Thomas
    Lyndaker, Aaron
    Savino, James
    Suttner, Ethan
    Li, Yulin
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2012, 22 (03)
  • [8] VENOUS SYNCHROTRON CORONARY ANGIOGRAPHY
    THOMLINSON, W
    GMUR, N
    CHAPMAN, D
    GARRETT, R
    LAZARZ, N
    MORRISON, J
    REISER, P
    PADMANABHAN, V
    ONG, L
    GREEN, S
    THOMPSON, A
    ZEMAN, H
    HOFSTADTER, R
    BROWN, G
    GIACOMINI, J
    GORDON, H
    RUBENSTEIN, E
    LANCET, 1991, 337 (8737): : 360 - 360
  • [9] Intravenous coronary angiography with synchrotron radiation
    Dix, WR
    Dill, T
    Kupper, W
    Lohmann, M
    Reime, B
    Ventura, R
    CARS 2001: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2001, 1230 : 883 - 888
  • [10] CORONARY ANGIOGRAPHY WITH SYNCHROTRON-RADIATION
    DIX, WR
    BESCH, HJ
    GRAEFF, W
    HAMM, CW
    ILLING, G
    KUPPER, W
    LOHMANN, M
    MEINERTZ, T
    MENK, RH
    REIME, B
    RUST, C
    SCHILDWACHTER, L
    TAFELMEIER, U
    WALENTA, AH
    JOURNAL DE PHYSIQUE IV, 1994, 4 (C9): : 279 - 286