Evaluation of CEDIA and DRI Drugs of Abuse Immunoassays for Urine Screening on a Thermo Indiko Plus Analyzer

被引:12
|
作者
Koehler, Katharina M. [1 ]
Hammer, Raija [1 ]
Riedy, Kathrin [1 ]
Auwaerter, Volker [1 ]
Neukamm, Merja A. [1 ]
机构
[1] Univ Freiburg, Inst Forens Med, Med Ctr, Fac Med, Freiburg, Germany
关键词
abstinence control; diagnostic tests; ethyl glucuronide; forensic toxicology; VALIDATION; KITS;
D O I
10.1002/jcla.22021
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: For analysis of urine samples during abstinence control for driving ability assessment (medical and psychological assessment, MPA), a reliable screening method for ethyl glucuronide and drugs of abuse (cannabinoids, opiates, cocaine, amphetamines, methadone, and benzodiazepines) is needed. Methods: In this study CEDIA and DRI immunoassays were applied on a Thermo Fisher Scientific Indiko Plus analyzer. Precision and accuracy as well as sensitivity and specificity at the required cut-offs for the MPA were evaluated. Results: The specificity was satisfactory and ranged from 91% for methamphetamine to 100% for opiates, cocaine metabolite, amphetamine, EDDP, and benzodiazepines. Moreover, sensitivity was 100% for all assays except for cannabinoids (91%). Conclusion: The presented method can therefore be recommended for abstinence control. (C) 2016 Wiley Periodicals, Inc.