Evaluation of performance of the GENECUBE assay for rapid molecular identification of Staphylococcus aureus and methicillin resistance in positive blood culture medium

被引:14
|
作者
Hida, Yukio [1 ]
Uemura, Keiichi [2 ]
Sugimoto, Hiroyasu [3 ]
Kawashima, Yosuke [4 ]
Koyanagi, Norito [2 ]
Notake, Shigeyuki [5 ]
Akashi, Yusaku [6 ,7 ]
Sakaguchi, Shohei [1 ]
Kimura, Hideki [1 ]
Suzuki, Hiromichi [6 ,7 ]
机构
[1] Univ Fukui Hosp, Dept Clin Lab, Eiheiji, Fukui, Japan
[2] Chutoen Gen Med Ctr, Dept Clin Lab, Shizuoka, Japan
[3] TOYOBO Co Ltd, Tsuruga Inst Biotechnol, Tsuruga, Fukui, Japan
[4] TOYOBO Co Ltd, Diagnost Syst Dept, Osaka, Osaka, Japan
[5] Tsukuba Med Ctr Hosp, Dept Clin Lab, Tsukuba, Ibaraki, Japan
[6] Tsukuba Med Ctr Hosp, Dept Clin Lab Med, Tsukuba, Ibaraki, Japan
[7] Tsukuba Med Ctr Hosp, Div Infect Dis, Dept Med, Tsukuba, Ibaraki, Japan
来源
PLOS ONE | 2019年 / 14卷 / 07期
关键词
STREAM INFECTIONS; STAPHSR ASSAY; S; AUREUS; VANCOMYCIN; EMERGENCE; OUTCOMES; SYSTEM; PCR;
D O I
10.1371/journal.pone.0219819
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Rapid identification of causative agents from positive blood culture media is a prerequisite for the timely targeted treatment of patients with sepsis. The GENECUBE (TOYOBO Co., Ltd.) is a novel, fully-automated gene analyzer that can purify DNAs and amplify target DNAs. In this study, we evaluated the ability of two newly developed GENECUBE assays to directly detect the nuc and mecA genes in blood culture medium; nuc is specific to Staphylococcus aureus, and mecA indicates methicillin resistance. We examined 263 positive blood culture samples taken at three hospitals from patients suspected of having staphylococcal bacteremia. The results were then compared with those obtained using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry, antimicrobial susceptibility testing (Microscan system or Dry-plate EIKEN), and sequencing analysis. The GENECUBE assays had sensitivity and specificity of 100% in detecting both S. aureus and methicillin resistance in positive blood culture. The turnaround time of the examination was evaluated for 36 positive blood culture samples. The time between the initiation of incubation and completion of the GENECUBE examination was 23 h (inter-quartile range: IQR 21-37 h); the time between reporting of Gram stain examination and completion of the GENECUBE examination was 52 min (IQR 48-62 min). These findings show that the GENECUBE assays significantly reduce the assay time with no loss of sensitivity or specificity.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Evaluation of the BD Max StaphSR Assay for Rapid Identification of Staphylococcus aureus and Methicillin-Resistant S-aureus in Positive Blood Culture Broths
    Dalpke, Alexander H.
    Hofko, Marjeta
    Hamilton, Fiona
    Mackenzie, Laura
    Zimmermann, Stefan
    Templeton, Kate
    JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (11) : 3630 - 3632
  • [2] Impact of Rapid Identification of Staphylococcus Species in Positive Blood Culture Using GeneXpert Methicillin-Resistant Staphylococcus aureus/ Staphylococcus aureus Blood Culture Assay Combined with Antibiotic Stewardship
    AlQahtani, Hajar
    Alqahtani, Fulwah Yahya
    Aleanizy, Fadilah Sfouq
    Baloch, Saeed
    Tabb, Deanne
    MICROBIAL DRUG RESISTANCE, 2021, 27 (08) : 1037 - 1043
  • [3] Performance of the BD GeneOhm StaphSR Assay for Simultaneous Identification of Methicillin Sensitive and Methicillin Resistance Staphylococcus aureus from Gram Positive Blood Cultures
    Gaynor, P.
    Seth, A.
    Lam, X.
    Burday, M.
    Fernandes, H.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2008, 10 (06): : 617 - 617
  • [4] Evaluation of the BinaxNOW PBP2a assay for the direct detection of methicillin resistance in Staphylococcus aureus from positive blood culture bottles
    Pilar Romero-Gomez, Maria
    Quiles-Melero, Inmaculada
    Navarro, Carolina
    Ramon Pano-Pardo, Jose
    Gomez-Gil, Rosa
    Mingorance, Jesus
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2012, 72 (03) : 282 - 284
  • [5] Clinical impact of a real-time PCR assay for rapid identification of Staphylococcus aureus and determination of methicillin resistance from positive blood cultures
    Cattoir, V.
    Merabet, L.
    Djibo, N.
    Rioux, C.
    Legrand, P.
    Girou, E.
    Lesprit, P.
    CLINICAL MICROBIOLOGY AND INFECTION, 2011, 17 (03) : 425 - 431
  • [6] Evaluation of three molecular assays for rapid identification of methicillin-resistant Staphylococcus aureus
    Francois, Patrice
    Bento, Manuela
    Renzi, Gesuele
    Harbarth, Stephan
    Pittet, Didier
    Schrenzel, Jacques
    JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (06) : 2011 - 2013
  • [7] Direct identification of Staphylococcus aureus from positive blood culture specimens with a rapid test
    Garcia, P
    Levican, J
    Quiroga, T
    Montiel, F
    Palavecino, E
    JOURNAL OF MEDICAL MICROBIOLOGY, 2002, 51 (06) : 530 - 531
  • [8] Evaluation of a novel optical assay for rapid detection of methicillin-resistant Staphylococcus aureus in liquid culture
    Knaack, Dennis
    Idelevich, Evgeny A.
    Koerber-Ingang, Barbara
    Kresken, Michael
    Becker, Karsten
    JOURNAL OF MICROBIOLOGICAL METHODS, 2018, 146 : 68 - 70
  • [9] Evaluation of colorimetric nitrate reductase assay for rapid detection of methicillin resistance in clinical isolates of Staphylococcus aureus
    Niveditha, N.
    Sistla, Sujatha
    INDIAN JOURNAL OF MEDICAL RESEARCH, 2016, 143 : 87 - 91
  • [10] Evaluation of three rapid methods for detection of methicillin resistance in Staphylococcus aureus
    Louie, L
    Matsumura, SO
    Choi, E
    Louie, M
    Simor, AE
    JOURNAL OF CLINICAL MICROBIOLOGY, 2000, 38 (06) : 2170 - 2173