Testing methods for MR safety and compatibility of medical devices

被引:23
|
作者
Schaefers, G
Melzer, A
机构
[1] MR Comp GmbH, MR Testing Lab, D-45894 Gelsenkirchen, Germany
[2] Univ Appl Sci, INSITE Med, Gelsenkirchen, Germany
[3] Univ Appl Sci, Dept Engn Phys, Gelsenkirchen, Germany
关键词
magnetic resonance imaging; MRI safety; MR testing; MRI compatibility; implants; medical devices;
D O I
10.1080/13645700600640865
中图分类号
R61 [外科手术学];
学科分类号
摘要
Magnetic resonance (MR) safety and image compatibility are important issues when using medical devices in the environment of magnetic resonance imaging ( MRI). MR testing of medical devices is required for device approval by the FDA and the EU notified bodies. Testing methods have been established for basic issues and were published as first ASTM International standards ( formerly ASTM, the American Society of Testing and Materials). Moreover, individual safety concerns such as the safe functioning of the device and the MR system are a necessity for medical devices within an MR environment. Standardized tests increase the safety of the patient and support both the MR user and the device manufacturers. This paper describes the state of the art of MR testing methods and requirements for medical devices.
引用
收藏
页码:71 / 75
页数:5
相关论文
共 50 条
  • [1] Magnetic Resonance (MR) Safety and Compatibility - Survey of Methods and Standards for Testing and Marking Medical Devices
    Schaefers, Gregor
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 12, 2009, 25 (12): : 285 - 287
  • [2] Testing MR safety and compatibility
    Schaefers, Gregor
    IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 2008, 27 (03): : 23 - 27
  • [3] Neurological implants: current standards and testing methods for magnetic resonance safety and compatibility of medical devices
    Schaefers, G.
    JOURNAL OF NEUROLOGY, 2007, 254 : 77 - 77
  • [4] Test methods to determine magnetic resonance (MR) safety and MR image compatibility of implants/devices
    Schaefers, Gregor
    RADIOLOGE, 2019, 59 (10): : 875 - 884
  • [5] MRI safety and compatibility of implants and medical devices
    Woods, TO
    STAINLESS STEELS FOR MEDICAL AND SURGICAL APPLICATIONS, 2003, 1438 : 82 - 90
  • [6] Practical Aspects of MR Imaging Safety Test Methods for MR Conditional Active Implantable Medical Devices
    Al-Dayeh, Louai
    Rahman, Mizan
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2020, 28 (04) : 559 - +
  • [7] MR safety assessment of active implantable medical devices
    Aissani, Sarra
    Laistler, Elmar
    Felblinger, Jacques
    RADIOLOGE, 2019, 59 (SUPPL 1): : 40 - 45
  • [8] Testing of break resistance of medical safety devices
    Wicker, Sabine
    Lechner, Pia
    Rabenau, Holger F.
    Gottschalk, Rene
    Bockenheimer, Alexander
    Berger, Christina
    ZENTRALBLATT FUR ARBEITSMEDIZIN ARBEITSSCHUTZ UND ERGONOMIE, 2008, 58 (07): : 196 - 202
  • [9] Assessing MR-compatibility of somatosensory stimulation devices: A systematic review on testing methodologies
    Travassos, Carolina
    Sayal, Alexandre
    Direito, Bruno
    Pereira, Joao
    Sousa, Teresa
    Castelo-Branco, Miguel
    FRONTIERS IN NEUROSCIENCE, 2023, 17
  • [10] MR compatibility of mechatronic devices: Design criteria
    Chinzei, K
    Kikinis, R
    Jolesz, FA
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION, MICCAI'99, PROCEEDINGS, 1999, 1679 : 1020 - 1030