Application of Molecularly Imprinted Polymer-modified Potentiometric Sensor for Quantitative Determination of Histamine in Serum

被引:3
|
作者
Konishi, Atsuko [1 ]
Takegami, Shigehiko [1 ]
Idosaki, Yoko [1 ]
Kitade, Tatsuya [1 ]
机构
[1] Kyoto Pharmaceut Univ, Dept Analyt Chem, Yamashina Ku, 5 Misasaginakauchi Cho, Kyoto 6078414, Japan
关键词
Potentiometric sensor; molecularly imprinted polymer; histamine; serum; quantitative determination; selectivity; H-4; RECEPTOR;
D O I
10.2116/analsci.20N018
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A molecularly imprinted polymer-modified potentiometric histamine (HIS) sensor was prepared and used for quantitative determination of HIS in bovine serum. The calibration curve using the potential responses measured in 1 x 10(-3) mol L-1 phosphate buffer (pH 7.4) showed good linearity in the HIS concentration range of 3 x 10(-4) to 1 x 10(-2) mol L-1 (r = 0.92), with a detection limit of 1.6 x 10(-4) mol L-1. In bovine serum samples, the HIS sensor showed good recovery values of 91 - 104%. Therefore, this HIS sensor successfully determined the HIS concentration in bovine serum samples.
引用
收藏
页码:1561 / 1563
页数:3
相关论文
共 50 条
  • [1] Application of Molecularly Imprinted Polymer-modified Potentiometric Sensor for Quantitative Determination of Histamine in Serum
    Atsuko Konishi
    Shigehiko Takegami
    Yoko Idosaki
    Tatsuya Kitade
    Analytical Sciences, 2020, 36 : 1561 - 1563
  • [2] A Molecularly Imprinted Polymer-modified Potentiometric Sensor for the Detection of Glutathione
    Konishi, Atsuko
    Takegami, Shigehiko
    Kitade, Tatsuya
    ANALYTICAL SCIENCES, 2019, 35 (10) : 1111 - 1115
  • [3] A Molecularly Imprinted Polymer-modified Potentiometric Sensor for the Detection of Glutathione
    Atsuko Konishi
    Shigehiko Takegami
    Tatsuya Kitade
    Analytical Sciences, 2019, 35 : 1111 - 1115
  • [4] Atrazine sensor based on molecularly imprinted polymer-modified gold electrode
    Shoji, R
    Takeuchi, T
    Kubo, I
    ANALYTICAL CHEMISTRY, 2003, 75 (18) : 4882 - 4886
  • [5] A New Potentiometric Sensor for Determination and Screening Phenylalanine in Blood Serum Based on Molecularly Imprinted Polymer
    Bangaleh, Zahra
    Sadeghi, Hayedeh Bagheri
    Ebrahimi, Soltan Ahmad
    Najafizadeh, Parvaneh
    IRANIAN JOURNAL OF PHARMACEUTICAL RESEARCH, 2019, 18 (01): : 61 - 71
  • [6] Potentiometric sensor based on molecularly imprinted polymer for determination of melamine in milk
    Liang, Rongning
    Zhang, Ruiming
    Qin, Wei
    SENSORS AND ACTUATORS B-CHEMICAL, 2009, 141 (02): : 544 - 550
  • [7] Voltammetric sensor for vanillylmandelic acid based on molecularly imprinted polymer-modified electrodes
    Blanco-López, MC
    Lobo-Castañón, MJ
    Miranda-Ordieres, AJ
    Tuñón-Blanco, P
    BIOSENSORS & BIOELECTRONICS, 2003, 18 (04): : 353 - 362
  • [8] Determination of glycine in aqueous solutions using a molecularly imprinted polymer-modified piezosensor
    A. N. Zyablov
    A. V. Kalach
    Yu. A. Zhibrova
    V. F. Selemenev
    O. V. D’yakonova
    Journal of Analytical Chemistry, 2010, 65 : 91 - 93
  • [9] Determination of Glycine in Aqueous Solutions Using a Molecularly Imprinted Polymer-Modified Piezosensor
    Zyablov, A. N.
    Kalach, A. V.
    Zhibrova, Yu. A.
    Selemenev, V. F.
    D'yakonova, O. V.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2010, 65 (01) : 91 - 93
  • [10] Designing of molecularly imprinted polymer-based potentiometric sensor for the determination of heparin
    Li, Lifeng
    Liang, Yizeng
    Liu, Youcai
    ANALYTICAL BIOCHEMISTRY, 2013, 434 (02) : 242 - 246