Noise characteristics of the gas ionization cascade used in low vacuum scanning electron microscopy

被引:5
|
作者
Tileli, Vasiliki [1 ]
Knowles, W. Ralph [2 ]
Toth, Milos [2 ]
Thiel, Bradley L. [1 ]
机构
[1] SUNY Albany, Coll Nanoscale Sci & Engn, Albany, NY 12203 USA
[2] FEI Co, Hillsboro, OR 97124 USA
关键词
noise; optimisation; scanning electron microscopy; MULTIPLICATION NOISE; IMAGE-ENHANCEMENT; AMPLIFICATION; SCATTERING; PHOTOMULTIPLIER; CONTRAST; SEM;
D O I
10.1063/1.3159883
中图分类号
O59 [应用物理学];
学科分类号
摘要
The noise characteristics of gas cascade amplified electron signals in low vacuum scanning electron microscopy (LVSEM) are described and analyzed. We derive expressions for each component contributing to the total noise culminating in a predictive, quantitative model that can be used for optimization of LVSEM operating parameters. Signal and noise behavior is characterized experimentally and used to validate the model. Under most operating conditions, the noise is dominated by the excess noise generated in the gas amplification cascade. At high gains, the excess noise increases proportionally with gain such that the signal-to-noise ratio is constant. The effects of several instrument operating parameters, including working distance, gas pressure, beam current, and detector bias, are condensed and presented in the form of a master curve.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Master curves for gas amplification in low vacuum and environmental scanning electron microscopy
    Thiel, BL
    ULTRAMICROSCOPY, 2004, 99 (01) : 35 - 47
  • [2] Salivary glands: Scanning and low vacuum scanning electron microscopy
    Konig, B
    Kaneda, K
    Watanabe, I
    JOURNAL OF DENTAL RESEARCH, 1996, 75 (05) : 1085 - 1085
  • [3] Charging processes in low vacuum scanning electron microscopy
    Thiel, BL
    Toth, M
    Craven, JP
    MICROSCOPY AND MICROANALYSIS, 2004, 10 (06) : 711 - 720
  • [4] Characteristics and reduction of noise in scanning hot electron microscopy
    Kikegawa, Nobutaka
    Fuguya, Kazuhito
    Vazquez, Francisco
    Ikeda, Yoshihiro
    Japanese Journal of Applied Physics, Part 1: Regular Papers & Short Notes & Review Papers, 1998, 37 (12 A): : 6580 - 6584
  • [5] Characteristics and reduction of noise in scanning hot electron microscopy
    Kikegawa, N
    Furuya, K
    Vazquez, F
    Ikeda, Y
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1998, 37 (12A): : 6580 - 6584
  • [6] Two-stage gas amplifier for ultrahigh resolution low vacuum scanning electron microscopy
    Thiel, BL
    Toth, M
    Schroemges, RPM
    Scholtz, JJ
    van Veen, G
    Knowles, WR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (03):
  • [7] On the role of electron-ion recombination in low vacuum scanning electron microscopy
    Toth, M
    Thiel, BL
    Donald, AM
    JOURNAL OF MICROSCOPY, 2002, 205 : 86 - 95
  • [8] Preliminary studies of asphaltene aggregates by low vacuum scanning electron microscopy
    Bragado, GAC
    Guzmán, ETR
    Yacamán, MJ
    PETROLEUM SCIENCE AND TECHNOLOGY, 2001, 19 (1-2) : 45 - 53
  • [9] Observation of the otolithic membrane by low-vacuum scanning electron microscopy
    Nakai, Y
    Masutani, H
    Kato, A
    Sugiyama, T
    ORL-JOURNAL FOR OTO-RHINO-LARYNGOLOGY AND ITS RELATED SPECIALTIES, 1996, 58 (01): : 9 - 12
  • [10] Imaging deep trap distributions by low vacuum scanning electron microscopy
    Toth, Milos
    Knowles, W. Ralph
    Phillips, Matthew R.
    APPLIED PHYSICS LETTERS, 2007, 90 (07)