Anesthetic Consideration for Patients with Micra Leadless Pacemaker

被引:14
|
作者
Karuppiah, Sathappan [1 ]
Prielipp, Richard [1 ]
Banik, Ratan K. [1 ]
机构
[1] Univ Minnesota, Dept Anesthesiol, Twin Cities Campus,B515 Mayo Mem Bldg, Minneapolis, MN 55455 USA
关键词
EMI; MAGNET; MICRA; PERIOPERATIVE MANAGEMENT; DEVICES;
D O I
10.4103/aca.ACA_191_19
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
MICRA, miniaturized leadless single chamber pacemaker, is inserted directly into the right ventricular myocardium via transcatheter approach. We present a case of a 66-year-old patient with a Micra pacemaker scheduled for kidney-pancreas transplant. The patient is pacemaker dependent. The preoperative cardiology consult did not comment on the need of reprogramming. One hour prior to the surgery, the anesthesia team was unable to locate the pacemaker on the chest wall. The Medtronic hotline was called, and the caregivers learned that the particular pacemaker is buried within the ventricular wall and is not responsive to an external magnet. Thus, the case was delayed and a cardiac electrophysiology team was contacted to reprogram the pacemaker to VOO (fixed ventricular pacing) mode. We suggest that the pacemaker can pose perioperative challenges due to its novelty, paucity of report, and guidelines.
引用
收藏
页码:494 / 496
页数:3
相关论文
共 50 条
  • [1] Leadless Pacemaker Implantation in Hemodialysis Patients Experience With the Micra Transcatheter Pacemaker
    El-Chami, Mikhael F.
    Clementy, Nicolas
    Garweg, Christophe
    Omar, Razali
    Duray, Gabor Z.
    Gornick, Charles C.
    Leyva, Francisco
    Sagi, Venkata
    Piccini, Jonathan P.
    Soejima, Kyoko
    Stromberg, Kurt
    Roberts, Paul R.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2019, 5 (02) : 162 - 170
  • [2] Mortality After Micra Leadless Pacemaker Implantation
    Kamath, Akshay R.
    Heard, Brittney Z.
    Darby, Andrew E.
    Malhotra, Rohit
    Mangrum, James M.
    Bilchick, Kenneth C.
    Mason, Pamela K.
    Ferguson, John D.
    Mehta, Nishaki K.
    CIRCULATION, 2018, 138
  • [3] Strategies to improve atrioventricular synchrony in patients with a Micra AV leadless pacemaker
    Garweg, Christophe
    Breitenstein, Alexander
    Clementy, Nicolas
    De Asmundis, Carlo
    Iacopino, Saverio
    Johansen, Jens Brock
    Sharman, David
    Theis, Cathrin
    Prat, Xavier Vinolas
    Winter, Stefan
    Reichlin, Tobias
    EUROPACE, 2024, 26 (03):
  • [4] Periprocedural anticoagulation therapy in patients undergoing micra leadless pacemaker implantation
    Kadado, Anis John
    Chalhoub, Fadi
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2023, 371 : 221 - 225
  • [5] Leadless pacemaker implant in patients with a history of open heart surgery: experience with the Micra transcatheter pacemaker
    Garweg, C.
    Iacopino, S.
    El-Chami, M. F.
    Veltmann, C.
    Clementy, N.
    Grubman, E.
    Johansen, J. B.
    Knops, R.
    Schalij, M. J.
    Piccini, J. P.
    Soejima, K.
    Stromberg, K.
    Fagan, D. H.
    Roberts, P. R.
    EUROPEAN HEART JOURNAL, 2020, 41 : 773 - 773
  • [6] Is the new Micra-leadless pacemaker entirely safe?
    Da Costa, Antoine
    Romeyer-Bouchard, Cecile
    Guichard, Jean Baptiste
    Gerbay, Antoine
    Isaaz, Karl
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 212 : 97 - 99
  • [7] OUTCOMES OF MICRA LEADLESS PACEMAKER IMPLANTATION WITH UNINTERRUPTED ANTICOAGULATION
    Kiani, Soroosh
    Black, George
    Rao, Birju
    Thakkar, Nancy
    Massad, Christopher
    Akshar, Patel
    Merchant, Faisal
    Hoskins, Michael
    Delurgio, David
    Patel, Anshul
    Shah, Anand
    Leon, Angel
    Westerman, Stacy
    Lloyd, Michael
    El Chami, Mikhael
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (09) : 420 - 420
  • [8] Implantation of Micra AV: A leadless atrioventricular synchronous pacemaker
    Li, Ke
    Liu, Jinfeng
    ASIAN JOURNAL OF SURGERY, 2023, 46 (12) : 5761 - 5762
  • [9] Subclinical cardiac perforation caused by a Micra leadless pacemaker
    Togashi, Ikuko
    Sato, Toshiaki
    Hoshida, Kyoko
    Soejima, Kyoko
    JOURNAL OF ARRHYTHMIA, 2018, 34 (03) : 326 - 328
  • [10] Outcomes of Micra leadless pacemaker implantation with uninterrupted anticoagulation
    Kiani, Soroosh
    Black, George B.
    Rao, Birju
    Thakkar, Nancy
    Massad, Christopher
    Patel, Akshar V.
    Merchant, Faisal M.
    Hoskins, Michael H.
    De Lurgio, David B.
    Patel, Anshul M.
    Shah, Anand D.
    Leon, Angel R.
    Westerman, Stacy B.
    Lloyd, Michael S.
    El-Chami, Mikhael F.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (08) : 1313 - 1318