Changing pulsatility by delaying the rotational speed phasing of a rotary left ventricular assist device

被引:12
|
作者
Date, Kazuma [1 ,2 ]
Nishimura, Takashi [3 ]
Arakawa, Mamoru [1 ]
Takewa, Yoshiaki [1 ]
Kishimoto, Satoru [1 ]
Umeki, Akihide [2 ]
Ando, Masahiko [2 ]
Mizuno, Toshihide [1 ]
Tsukiya, Tomonori [1 ]
Ono, Minoru [2 ]
Tatsumi, Eisuke [1 ]
机构
[1] Natl Cerebral & Cardiovasc Ctr Res Inst, Dept Artificial Organs, 5-7-1 Fujishiro Dai, Suita, Osaka 5658565, Japan
[2] Univ Tokyo, Dept Cardiothorac Surg, Tokyo, Japan
[3] Tokyo Metropolitan Geriatr Hosp, Dept Cardiac Surg, Itabashi Ku, 35-2 Sakaecho, Tokyo 1730015, Japan
关键词
Continuous-flow LVAD; Synchronization with cardiac beat; Rotary pump; LEFT-HEART BYPASS; BEAT SYNCHRONIZING SYSTEM; FLOW; SUPPORT; PERFUSION; PRESSURE; RECOVERY; MODEL;
D O I
10.1007/s10047-016-0920-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Continuous-flow left ventricular assist devices (LVADs) have improved the prognosis of end-stage heart failure. However, continuous-flow LVADs diminish pulsatility, which possibly result in bleeding, aortic insufficiency, and other adverse effects. We previously developed a novel control system for a continuous-flow LVAD (EVAHEART(A (R)); Sun Medical), and demonstrated that we could create sufficient pulsatility by increasing its rotational speed (RS) in the systolic phase (Pulsatile Mode) in the normal heart model. Here, we aimed to evaluate differences between systolic assist with advanced and delayed loads by shifting the timing of increased RS. We implanted EVAHEART in six goats (55.3 +/- 4.3 kg) with normal hearts. We reduced their heart rates to < 60 bpm using propranolol and controlled the heart rates at 80 and 120 bpm using ventricular pacing. We shifted the timing of increasing RS from -60 to +60 ms in the systolic phase. We found significant increases in all the following parameters when assessments of delayed timing (+60 ms) were compared with assessments of advanced timing (-60 ms): pulse pressure, mean dP/dt max of aortic pressure, and energy-equivalent pulse pressure. During continuous-flow LVAD support, pulsatility can be controlled using a rotary pump. In particular, pulsatility can be shifted by delaying increased RS.
引用
收藏
页码:18 / 25
页数:8
相关论文
共 50 条
  • [1] Changing pulsatility by delaying the rotational speed phasing of a rotary left ventricular assist device
    Kazuma Date
    Takashi Nishimura
    Mamoru Arakawa
    Yoshiaki Takewa
    Satoru Kishimoto
    Akihide Umeki
    Masahiko Ando
    Toshihide Mizuno
    Tomonori Tsukiya
    Minoru Ono
    Eisuke Tatsumi
    Journal of Artificial Organs, 2017, 20 : 18 - 25
  • [2] Rotational speed modulation used with continuous-flow left ventricular assist device provides good pulsatility
    Naito, Noritsugu
    Nishimura, Takashi
    Iizuka, Kei
    Takewa, Yoshiaki
    Umeki, Akihide
    Ando, Masahiko
    Ono, Minoru
    Tatsumi, Eisuke
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2018, 26 (01) : 119 - 123
  • [3] Pulsatility During Left Ventricular Assist Device Treatment
    Kawamoto, Shunsuke
    Saiki, Yoshikatsu
    CIRCULATION JOURNAL, 2015, 79 (09) : 1895 - 1896
  • [4] In Vitro Investigation of the Effect of Left Ventricular Assist Device Speed and Pulsatility Mode on Intraventricular Hemodynamics
    Chassagne, Fanette
    Miramontes, Marissa
    Chivukula, Venkat Keshav
    Li, Song
    Beckman, Jennifer A.
    Mahr, Claudius
    Aliseda, Alberto
    ANNALS OF BIOMEDICAL ENGINEERING, 2021, 49 (05) : 1318 - 1332
  • [5] In Vitro Investigation of the Effect of Left Ventricular Assist Device Speed and Pulsatility Mode on Intraventricular Hemodynamics
    Fanette Chassagne
    Marissa Miramontes
    Venkat Keshav Chivukula
    Song Li
    Jennifer A. Beckman
    Claudius Mahr
    Alberto Aliseda
    Annals of Biomedical Engineering, 2021, 49 : 1318 - 1332
  • [6] Physiological control of rotary left ventricular assist device
    Yi, Wu
    PROCEEDINGS OF THE 26TH CHINESE CONTROL CONFERENCE, VOL 6, 2007, : 469 - 474
  • [7] Pulsatility and Aortic Insufficiency during Left Ventricular Assist Device Treatment
    Imamura, Teruhiko
    Kinugawa, Koichiro
    Nitta, Daisuke
    Hatano, Masaru
    Kinoshita, Osamu
    Nawata, Kan
    Ono, Minoru
    JOURNAL OF CARDIAC FAILURE, 2015, 21 (10) : S190 - S191
  • [8] Hemodynamic controller for left ventricular assist device based on pulsatility ratio
    Choi, Seongjin
    Boston, J. Robert
    Antaki, James F.
    ARTIFICIAL ORGANS, 2007, 31 (02) : 114 - 125
  • [9] Arterial pulsatility under phasic left ventricular assist device support
    Bozkurt, Selim
    van Tuijl, Sjoerd
    van de Vosse, Frans N.
    Rutten, Marcel C. M.
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2016, 27 (05) : 451 - 460
  • [10] Impact of the Rotational Pump Speed of the Implantable Left Ventricular Assist Device on Right Ventricular Pump Function
    Sakamoto, Takafumi
    Fujino, Takeo
    Higo, Taiki
    Ide, Tomomi
    Tanoue, Yoshihisa
    Shiose, Akira
    Tsutsui, Hiroyuki
    JOURNAL OF CARDIAC FAILURE, 2016, 22 (09) : S164 - S164