Stimulation Stability and Selectivity of Chronically Implanted Multicontact Nerve Cuff Electrodes in the Human Upper Extremity

被引:89
|
作者
Polasek, Katharine H. [1 ]
Hoyen, Harry A. [2 ,3 ]
Keith, Michael W. [2 ,3 ]
Kirsch, Robert F. [1 ,4 ]
Tyler, Dustin J. [1 ,4 ]
机构
[1] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
[2] Case Western Reserve Univ, Dept Orthopaed & Biomed Engn, Cleveland, OH 44106 USA
[3] Metrohlth Med Ctr, Cleveland, OH 44106 USA
[4] Vet Affairs Med Ctr, Cleveland Louis Stokes Dept, Cleveland, OH 44106 USA
基金
美国国家卫生研究院;
关键词
Functional electrical stimulation (FES); nerve electrodes; implantable electrodes; peripheral nerve; selective stimulation; upper extremity neuroprostheses; RECRUITMENT PROPERTIES; MUSCULOSKELETAL MODEL; SKELETAL-MUSCLE; TIME COURSE; NEUROPROSTHESIS; ACTIVATION; POSITION; MOTOR;
D O I
10.1109/TNSRE.2009.2032603
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Nine spiral nerve cuff electrodes were implanted in two human subjects for up to three years with no adverse functional effects. The objective of this study was to look at the long term nerve and muscle response to stimulation through nerve cuff electrodes. The nerve conduction velocity remained within the clinically accepted range for the entire testing period. The stimulation thresholds stabilized after approximately 20 weeks. The variability in the activation over time was not different from muscle-based electrodes used in implanted functional electrical stimulation systems. Three electrodes had multiple, independent contacts to evaluate selective recruitment of muscles. A single muscle could be selectively activated from each electrode using single-contact stimulation and the selectivity was increased with the use of field steering techniques. The selectivity after three years was consistent with selectivity measured during the implant surgery. Nerve cuff electrodes are effective for chronic muscle activation and multichannel functional electrical stimulation in humans.
引用
收藏
页码:428 / 437
页数:10
相关论文
共 50 条
  • [1] STIMULATION OF FACIAL-NERVE WITH CHRONICALLY IMPLANTED CUFF ELECTRODES AS UNCONDITIONED STIMULUS FOR CONDITIONED BLINK
    BLACKCLEWORTH, P
    FEDERATION PROCEEDINGS, 1975, 34 (03) : 397 - 397
  • [2] Spiral nerve cuff electrodes for an upper extremity neuroprosthesis
    Polasek, Katharine H.
    Hoyen, Harry A.
    Keith, Michael W.
    Kirsch, Robert F.
    Tyler, Dustin J.
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 5217 - +
  • [3] Selective Neural Electrical Stimulation of an Injured Facial Nerve Using Chronically Implanted Dual Cuff Electrodes
    Abiri, Arash
    Chau, Steven
    James, Nathan R.
    Goshtasbi, Khodayar
    Birkenbeuel, Jack L.
    Sahyouni, Ronald
    Edwards, Robert
    Djalilian, Hamid R.
    Lin, Harrison W.
    BRAIN SCIENCES, 2022, 12 (11)
  • [4] Stability of the input-output properties of chronically implanted multiple contact nerve cuff stimulating electrodes
    Grill, W.M.
    Mortimer, J.T.
    IEEE Transactions on Rehabilitation Engineering, 1998, 6 (04): : 364 - 373
  • [5] Remote Stimulation of Sciatic Nerve Using Cuff Electrodes and Implanted Diodes
    Sridharan, Arati
    Chirania, Sanchit
    Towe, Bruce C.
    Muthuswamy, Jit
    MICROMACHINES, 2018, 9 (11):
  • [6] REFLEXES INDUCED BY NERVE-STIMULATION IN WALKING CATS WITH IMPLANTED CUFF ELECTRODES
    DUYSENS, J
    STEIN, RB
    EXPERIMENTAL BRAIN RESEARCH, 1978, 32 (02) : 213 - 224
  • [7] CONSIDERATIONS FOR SAFETY WITH CHRONICALLY IMPLANTED NERVE ELECTRODES
    AGNEW, WF
    MCCREERY, DB
    EPILEPSIA, 1990, 31 : S27 - S32
  • [8] Chronic stability and selectivity of four-contact spiral nerve-cuff electrodes in stimulating the human femoral nerve
    Fisher, L. E.
    Tyler, D. J.
    Anderson, J. S.
    Triolo, R. J.
    JOURNAL OF NEURAL ENGINEERING, 2009, 6 (04)
  • [9] UNIPOLAR AND BIPOLAR STIMULATION THRESHOLDS OF HUMAN MYOCARDIUM WITH CHRONICALLY IMPLANTED PACEMAKER ELECTRODES
    DEKKER, E
    BULLER, J
    VANERVEN, FA
    AMERICAN HEART JOURNAL, 1966, 71 (05) : 671 - +
  • [10] Intraoperative testing of selectivity of spiral nerve cuff electrodes
    Polasek, KH
    Hoyen, HA
    Keith, MW
    Kirsch, RF
    Tyler, DJ
    2005 2ND INTERNATINOAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING, 2005, : 486 - 489