Determination of Chloroanisoles and Chlorophenols in Cork and Wine by using HS-SPME and GC-ECD Detection

被引:1
|
作者
Ozhan, Didehan [1 ]
Anli, R. Ertan [1 ]
Vural, Nilufer [2 ]
Bayram, Mustafa [1 ]
机构
[1] Ankara Univ, Fac Engn, Dept Food Engn, TR-06110 Ankara, Turkey
[2] Ankara Univ, Inst Biotechnol, Sect Instrumental Anal, TR-06110 Ankara, Turkey
关键词
chloroanisoles; chlorophenols; cork; cork taint; HS-SPME (HeadSpace-Solid Phase Microextraction); GC-ECD (Gas Chromatography-Electron Capture Detection); wine; SOLID-PHASE MICROEXTRACTION; CAPILLARY GAS-CHROMATOGRAPHY; EXTRACTION METHOD; WATER SAMPLES; LIQUID-CHROMATOGRAPHY; TRACE ENRICHMENT; 2,4,6-TRICHLOROANISOLE; PHENOLS; TAINT; QUANTIFICATION;
D O I
暂无
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
Cork taint is an off-flavor problem in wine, the main reason being the presence of 2,4,6-trichloroanisole (TCA) in the cork stopper. In addition to the TCA, the presence of other chloroanisole and chlorophenol family compounds (the perception limits of which are very low) can also result in, or contribute to, corktaint problem. In this Study, the levels of 2,4-dichloroanisole (DCA), 2,4,6-trichloroanisole (TCA), 2,3,4,6-tetrachloroanisole (TeCA), pentachloroanisole (PCA), 2,4,6-trichlorophenol (TCP), 2,3,4,6-tetrachlorophenol (TeCP) and penthaclorophenol (PCP) were assayed in cork stoppers (natural, agglomerated and colmate) and in red wine samples from different wineries in Turkey using HS-SPME and GC-ECD detection. The performance parameters for all chloroanisole and chlorophenol compounds were as follows: recovery 92.48-102.53%, R 0.992-0.996. The LOQ values were DCA (8.4 ng/L), TCA (0.8 ng/L), TeCa (0.6 ng/L), PCP (0.8 ng/L), TCP (0.8 ng/L), TeCP (1.2 ng/L), and PCP (1.1 ng/L) respectively. In cork stoppers, the amounts of 2,4,6 TCA ranged between 5.4-130.6 ng/g. The 2,3,4,6 TeCA ranged between 1.12-8.2 ng/g and the PCA ranged between nd (not detected)-11.01 ng/g. In the wine samples, 2,4,6 TCA ranged between 1.42-70.2 ng/L. The 2,3,4,6 TeCA ranged between nd-15.1 ng/L and the PCA ranged from nd-5.16 ng/L. The results indicated that there was a significant correlation between the TCA in wines and the TCA in cork stoppers.
引用
收藏
页码:71 / 77
页数:7
相关论文
共 50 条
  • [1] Matrix effect during the application of a rapid method using HS-SPME followed by GC-ECD for the analysis of 2,4,6-TCA in wine and cork soaks
    Vlachos, P.
    Kampioti, A.
    Kornaros, M.
    Lyberatos, G.
    FOOD CHEMISTRY, 2007, 105 (02) : 681 - 690
  • [2] OPTIMIZATION OF THE HS-SPME METHOD COUPLED WITH GC/μECD FOR DETERMINATION OF METHYLMERCURY IN WATER
    Kralova, Zuzana
    Vavrova, Milada
    Sucman, Emanuel
    Komarkova, Petra
    FRESENIUS ENVIRONMENTAL BULLETIN, 2014, 23 (12B): : 3362 - 3365
  • [3] Method development by GC-ECD and HS-SPME-GC-MS for beer volatile analysis
    da Silva, Gisele C.
    da Silva, Abner A. S.
    da Silva, Larissa S. N.
    Godoy, Ronoel Luiz de O.
    Nogueira, Luciana C.
    Quiterio, Simone L.
    Raices, Renata S. L.
    FOOD CHEMISTRY, 2015, 167 : 71 - 77
  • [4] Use of experimental design to develop a method for analysis of 1,3-dichloropropene isomers in water by HS-SPME and GC-ECD
    Antelo, Angel
    Lasa, Maitena
    Millan, Esmeralda
    CHROMATOGRAPHIA, 2007, 66 (7-8) : 555 - 563
  • [5] DETERMINATION OF CHLORPYRIFOS IN SURFACE WATER USING SPE-DI-SPME/GC-ECD
    Miclean, Mirela
    Senila, Lacrimioara
    Cadar, Oana
    Roman, Marius
    Levei, Erika Andrea
    Tanaselia, Claudiu
    Toeroek, Annamaria
    Majdik, Cornelia
    STUDIA UNIVERSITATIS BABES-BOLYAI CHEMIA, 2014, 59 (03): : 43 - 48
  • [6] Analysis of trihalomethanes in water and air from indoor swimming pools using HS-SPME/GC/ECD
    Sa, Christopher S. A.
    Boaventura, Rui A. R.
    Pereira, Isabel B.
    JOURNAL OF ENVIRONMENTAL SCIENCE AND HEALTH PART A-TOXIC/HAZARDOUS SUBSTANCES & ENVIRONMENTAL ENGINEERING, 2011, 46 (04): : 355 - 363
  • [7] HS-SPME/GC-MS methodologies for the analysis of volatile compounds in cork material
    N. Moreira
    P. Lopes
    M. Cabral
    P. Guedes de Pinho
    European Food Research and Technology, 2016, 242 : 457 - 466
  • [8] HS-SPME/GC-MS methodologies for the analysis of volatile compounds in cork material
    Moreira, N.
    Lopes, P.
    Cabral, M.
    Guedes de Pinho, P.
    EUROPEAN FOOD RESEARCH AND TECHNOLOGY, 2016, 242 (04) : 457 - 466
  • [9] Determination of chlorophenols using SPME and GC/MS
    Guidotti, M
    Ravaioli, G
    ANNALI DI CHIMICA, 1998, 88 (9-10) : 629 - 635
  • [10] Determination of Residual Dimethyl Sulfate in Cephalosporin Using HS-SPME/GC-MS
    Lu, Rongchun