Application of a silicon drift detector to measuring internal conversion electrons and analysis of their spectra

被引:4
|
作者
Zhang, Haifei [1 ]
Wu, Rui [1 ]
Wan, Wenjie [2 ]
Feng, Tiancheng [1 ]
Lei, Weiguo [1 ]
Wang, Yindong [1 ]
Long, Bin [1 ]
Yin, Jingpeng [1 ]
Zhan, Jinfeng [1 ]
Zhou, Jianbin [2 ]
机构
[1] Northwest Inst Nucl Technol, Xian 710024, Shaanxi, Peoples R China
[2] Chengdu Univ Technol, Coll Nucl Technol & Automat Engn, Chengdu 610059, Sichuan, Peoples R China
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 2020年 / 950卷 / 950期
关键词
Silicon drift detector (SDD); Internal conversion electrons (ICEs); Pearson IV distribution; Levenberg-Marquardt algorithm; (CS)-C-137; Xe-131m; SPECTROMETRY; RESOLUTION; SYSTEM; 137CS; RAYS;
D O I
10.1016/j.nima.2019.162941
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Small-volume, low-capacitance semiconductor X-ray silicon drift detectors (SDDs) have an excellent ability to resolve closely spaced X-ray lines. But the application to electrons is less known. We designed a detector system and performed spectrometry using SDDs for radionuclides emitting mono-energetic internal conversion electrons (ICEs) with energies from 129 keV to 656 keV. The system is compact, maintains simple readout electronics, and provides high-resolution measurements of ICEs without cooling to liquid nitrogen temperatures and without bulky lead shielding. Preliminary measurements using Cs-137 and Xe-131m are conducted to verify that SDDs are suitable as ICE spectrometers. Unlike that of X/gamma spectrometry, the shape of the ICE spectra is a non-Gaussian distribution with apparent kurtosis. Therefore, a Pearson IV distribution function combined with a Levenberg-Marquardt fitting algorithm is utilized to analyze the ICE spectra quantitatively. The ratios of the net counts of ICE peaks for Cs-137 and Xe-131m as measured using SDDs with these fitting approaches are consistent with those reported in literature with a relative deviation of less than 3.4%.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Manufactured silicon diode used as an internal conversion electrons detector
    Corrêa, AAS
    Bueno, CC
    Gonçalves, JAC
    Mendes, PFPR
    Pinto, JKC
    de Souza, JP
    Santos, MDD
    BRAZILIAN JOURNAL OF PHYSICS, 2004, 34 (3A) : 973 - 975
  • [2] Compton electrons in silicon drift detector:: First results
    Çonka-Nurdan, I
    Nurdan, K
    Laihem, K
    Walenta, AH
    Fiorini, C
    Hörnel, N
    Strüder, L
    Venanzi, C
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 3677 - 3681
  • [3] Preliminary results on Compton electrons in silicon drift detector
    Çonka-Nurdan, T
    Nurdan, K
    Laihem, K
    Walenta, AH
    Fiorini, C
    Freisleben, B
    Hörnel, N
    Pavel, NA
    Strüder, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) : 2526 - 2532
  • [4] Compton electrons in silicon drift detector:: First results
    Çonka-Nurdan, T
    Nurdan, K
    Laihem, K
    Walenta, AH
    Fiorini, C
    Hörnel, N
    Strüder, L
    Venanzi, C
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2080 - 2084
  • [5] A crystal detector for measuring beta and internal conversion electrons in flowing air containing fission gases
    Schell, WR
    Vives-Batlle, J
    Yoon, SR
    Tobin, MJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3): : 773 - 777
  • [6] Deconvolution of 238,239,240pu conversion electron spectra measured with a silicon drift detector
    Pomme, S.
    Marouli, M.
    Paepen, J.
    Markovic, N.
    Pollanen, R.
    APPLIED RADIATION AND ISOTOPES, 2018, 134 : 233 - 239
  • [7] ANALYSIS OF BACKSCATTER MOSSBAUER-SPECTRA OBTAINED WITH INTERNAL-CONVERSION ELECTRONS
    KRAKOWSKI, RA
    MILLER, RB
    NUCLEAR INSTRUMENTS & METHODS, 1972, 100 (01): : 93 - +
  • [8] DETECTION OF INTERNAL CONVERSION ELECTRONS IN CHARGED PARTICLE SPECTRA
    MOORE, CF
    KULLECK, JG
    COKER, WR
    CLARKSON, RG
    KENT, JJ
    COURTNEY, WJ
    NUCLEAR INSTRUMENTS & METHODS, 1970, 79 (02): : 353 - &
  • [9] Conversion electron spectrometry of Pu isotopes with a silicon drift detector
    Pomme, S.
    Paepen, J.
    Perajarvi, K.
    Turunen, J.
    Pollanen, R.
    APPLIED RADIATION AND ISOTOPES, 2016, 109 : 183 - 188
  • [10] Knock-on electrons in WA98 silicon drift detector