Dynamic contrast enhanced ultrasound imaging; the effect of imaging modes and parameter settings for a microvascular phantom

被引:0
|
作者
Moghimirad, Elahe [1 ,2 ,3 ]
Bamber, Jeffrey [1 ,2 ,3 ]
Harris, Emma [1 ,2 ,3 ]
机构
[1] Inst Canc Res, Joint Dept Phys, London, England
[2] Inst Canc Res, CRUK Canc Imaging Ctr, London, England
[3] Royal Marsden NHS Fdn Trust, London, England
关键词
Dynamic contrast enhanced ultrasound imaging; time intensity curve (TIC); time amplitude curve (TAC); imaging mode; parameter setting; microvascular phantom;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Dynamic contrast enhanced ultrasound (DCE-US) imaging has the potential to provide quantitative information which is sensitive to tumour perfusion, an indicator for tumour response to radiotherapy. To increase the reproducibility of time intensity curve (TIC) characteristics, we are developing a 3D DCE-US imaging system. There are, however, many choices to be made in system design, such as whether to use plane wave (PWI) or focused imaging (FI), and the values to use for parameters such as focal depth (FD), F-number (F#), mechanical index (MI) and number of angles (NA) (for PWI). We evaluated the effect of such choices on TICs (we refer to time-amplitude curve (TAC) here), using a microvascular flow phantom containing similar to 100,000 parallel microtubes, each 200 mu m in diameter. DCE-US 2D images were obtained using a Vantage (Verasonics Inc.) and a pulse-inversion algorithm. 800 frames were recorded at 10 Hz for PWI and FI. All measurements were repeated 3 times, injecting 0.4 ml of contrast agent (Sonozoid) and changing one parameter at a time, using the values: FD = 20, 40 mm; F# = 2, 4; MI = 0.11, 0.15, 0.25; NA = 3, 7, 11. For a large region of interest which included the periphery of the phantom, TACs were sharp and single-peaked for F2 but broader and double-peaked for F4, consistent with F4 averaging over a greater focal volume than F2. Choosing a smaller more central ROI reduced the effect but did not eliminate it completely. Placing the focus deeper than the center reduced the TAC amplitude due to attenuation but also resulted in a flatter TAC. PWI amplitude was greatest for 3 angles with evidence that this may be due to side lobe artefacts added to the contrast signal. TAC characteristics are thus expected to be highly sensitive to imaging parameters. This should be considered in longitudinal studies.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Dynamic-contrast-enhanced MR imaging: a phantom for offline parameter optimization and quality assurance
    Verhaart, R. F.
    Pellicer, R.
    Van der Lugt, A.
    Paulides, M. M.
    Kotek, G.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S32 - S33
  • [2] Development and Evaluation of a Phantom for Dynamic Contrast-Enhanced Imaging
    Brauweiler, Robert
    Eisa, Fabian
    Hupfer, Martin
    Nowak, Tristan
    Kolditz, Daniel
    Kalender, Willi A.
    INVESTIGATIVE RADIOLOGY, 2012, 47 (08) : 462 - 467
  • [3] Development of a dynamic flow imaging phantom for dynamic contrast-enhanced CT
    Driscoll, B.
    Keller, H.
    Coolens, C.
    MEDICAL PHYSICS, 2011, 38 (08) : 4866 - 4880
  • [4] Contrast enhanced ultrasound imaging
    Steven B. Feinstein
    Blai Coll
    Daniel Staub
    Dan Adam
    Arend F. L. Schinkel
    Folkert J. ten Cate
    Kai Thomenius
    Journal of Nuclear Cardiology, 2010, 17 : 106 - 115
  • [5] Contrast enhanced ultrasound imaging
    Feinstein, Steven B.
    Coll, Blai
    Staub, Daniel
    Adam, Dan
    Schinkel, Arend F. L.
    ten Cate, Folkert J.
    Thomenius, Kai
    JOURNAL OF NUCLEAR CARDIOLOGY, 2010, 17 (01) : 106 - 115
  • [6] High Frame Rate Contrast Enhanced Ultrasound Imaging of Lymphatic Vessel Phantom
    Zhu, Jiaqi
    Lin, Shengtao
    Harput, Sevan
    Toulemonde, Matthieu
    Leow, Chee Hau
    Tang, Meng-Xing
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [7] An Integrated System for Superharmonic Contrast-Enhanced Ultrasound Imaging: Design and Intravascular Phantom Imaging Study
    Li, Yang
    Ma, Jianguo
    Martin, K. Heath
    Yu, Mingyue
    Ma, Teng
    Dayton, Paul A.
    Jiang, Xiaoning
    Shung, K. Kirk
    Zhou, Qifa
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2016, 63 (09) : 1933 - 1943
  • [8] Imaging of Carotid Plaque Neovascularization by Contrast-Enhanced Ultrasound and Dynamic Contrast-Enhanced Magnetic Resonance Imaging
    Cattaneo, Mattia
    Sun, Jie
    Staub, Daniel
    Xu, Dongxiang
    Gallino, Jeanne Marie
    Santini, Paolo
    Porretta, Alessandra P.
    Yuan, Chun
    Balu, Niranjan
    Arnold, Marcel
    Froio, Alberto
    Limoni, Costanzo
    Wyttenbach, Rolf
    Gallino, Augusto
    CEREBROVASCULAR DISEASES, 2019, 48 (3-6) : 140 - 148
  • [9] Plane wave versus focused transmissions for contrast enhanced ultrasound imaging: the role of parameter settings and the effects of flow rate on contrast measurements
    Moghimirad, Elahe
    Bamber, Jeffrey
    Harris, Emma
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (09):
  • [10] Contrast enhanced ultrasound in liver imaging
    Nielsen, MB
    Bang, N
    EUROPEAN JOURNAL OF RADIOLOGY, 2004, 51 : S3 - S8