Speech understanding in background noise with the two-microphone adaptive beamformer BEAM™ in the nucleus Freedom™ cochlear implant system

被引:141
|
作者
Spriet, Ann
Van Deun, Lieselot
Eftaxiadis, Kyriaky
Laneau, Johan
Moonen, Marc
van Dijk, Bas
van Wieringen, Astrid
Wouters, Jan
机构
[1] Katholieke Univ Leuven, SISTA, SCD, ESAT, B-3001 Louvain, Belgium
[2] Katholieke Univ Leuven, Dept Neurosci, ExpORL, B-3001 Louvain, Belgium
[3] Cochlear Ltd, Lane Cove, Australia
[4] Cochlear CTCE, Mechelen, Belgium
来源
EAR AND HEARING | 2007年 / 28卷 / 01期
关键词
D O I
10.1097/01.aud.0000252470.54246.54
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: This paper evaluates the benefit of the two-microphone adaptive beamformer BEAM (TM) in the Nucleus Freedom (TM) cochlear implant (CI) system for speech understanding in background noise by Cl users. Design: A double-blind evaluation of the two-microphone adaptive beamformer BEAM and a hardware directional microphone was carried out with five adult Nucleus CI users. The test procedure consisted of a pre- and post-test in the lab and a 2-wk trial period at home. In the pre- and post-test, the speech reception threshold (SRT) with sentences and the percentage correct phoneme scores for CVC words were measured in quiet and background noise at different signal-to-noise ratios. Performance was assessed for two different noise configurations (with a single noise source and with three noise sources) and two different noise materials (stationary speech-weighted noise and multitalker babble). During the 2-wk trial period at home, the CI users evaluated the noise reduction performance in different listening conditions by means of the SSQ questionnaire. In addition to the perceptual evaluation, the noise reduction performance of the beamformer was measured physically as a function of the direction of the noise source. Results: Significant improvements of both the SRT in noise (average improvement of 5-16 dB) and the percentage correct phoneme scores (average improvement of 10-41%) were observed with BEAM compared to the standard hardware directional microphone. In addition, the SSQ questionnaire and subjective evaluation in controlled and real-life scenarios suggested a possible preference for the beamformer in noisy environments. Conclusions: The evaluation demonstrates that the adaptive noise reduction algorithm BEAM in the Nucleus Freedom CI-system. may significantly increase the speech perception by cochlear implantees in noisy listening conditions. This is the first mono-lateral (adaptive) noise reduction strategy actually implemented in a mainstream commercial CI.
引用
收藏
页码:62 / 72
页数:11
相关论文
共 16 条
  • [1] Speech recognition in noise for cochlear implantees with a two-microphone monaural adaptive noise reduction system
    Wouters, J
    Vanden Berghe, J
    EAR AND HEARING, 2001, 22 (05): : 420 - 430
  • [2] A two-microphone noise reduction system for cochlear implant users with nearby microphones -: Part II:: Performance evaluation
    Kompis, Martin
    Bertram, Matthias
    Senn, Pascal
    Mueller, Joachim
    Pelizzone, Marco
    Haeusler, Rudolf
    EURASIP JOURNAL ON ADVANCES IN SIGNAL PROCESSING, 2008, 2008 (1)
  • [3] A Two-Microphone Noise Reduction System for Cochlear Implant Users with Nearby Microphones—Part II: Performance Evaluation
    Martin Kompis
    Matthias Bertram
    Pascal Senn
    Joachim Müller
    Marco Pelizzone
    Rudolf Häusler
    EURASIP Journal on Advances in Signal Processing, 2008
  • [4] Two-Microphone Spatial Filtering Improves Speech Reception for Cochlear-Implant Users in Reverberant Conditions With Multiple Noise Sources
    Goldsworthy, Raymond L.
    TRENDS IN HEARING, 2014, 18
  • [5] Speech Understanding and Sound Source Localization by Cochlear Implant Listeners Using a Pinna-Effect Imitating Microphone and an Adaptive Beamformer
    Dorman, Michael F.
    Natale, Sarah
    Loiselle, Louise
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2018, 29 (03) : 197 - 205
  • [6] Speech Understanding in Noise by Patients With Cochlear Implants Using a Monaural Adaptive Beamformer
    Dorman, Michael F.
    Natale, Sarah
    Spahr, Anthony
    Castioni, Erin
    JOURNAL OF SPEECH LANGUAGE AND HEARING RESEARCH, 2017, 60 (08): : 2360 - 2363
  • [7] A noise cross PSD estimator based on improved minimum statistics method for two-microphone speech enhancement dedicated to a bilateral cochlear implant
    Kallel, Fathi
    Ghorbel, Mohamed
    Frikha, Mondher
    Berger-Vachon, Christian
    Ben Hamida, Ahmed
    APPLIED ACOUSTICS, 2012, 73 (03) : 256 - 264
  • [8] Two-microphone spatial filtering provides speech reception benefits for cochlear implant users in difficult acoustic environments
    Goldsworthy, Raymond L.
    Delhorne, Lorraine A.
    Desloge, Joseph G.
    Braida, Louis D.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2014, 136 (02): : 867 - 876
  • [9] A Two-Microphone Noise Reduction System for Cochlear Implant Users with Nearby Microphones—Part I: Signal Processing Algorithm Design and Development
    Martin Kompis
    Matthias Bertram
    Jacques François
    Marco Pelizzone
    EURASIP Journal on Advances in Signal Processing, 2008
  • [10] A two-microphone noise reduction system for cochlear implant users with nearby microphones - Part I: Signal processing algorithm design and development
    Kompis, Martin
    Bertram, Matthias
    Francois, Jacques
    Pelizzone, Marco
    EURASIP JOURNAL ON ADVANCES IN SIGNAL PROCESSING, 2008, 2008 (1)