A Dual-Label Time-Resolved Fluorescence Immunoassay for the Simultaneous Determination of Cardiac Troponin T and Myoglobin

被引:18
|
作者
Wang, Ling [1 ]
Xu, Mujuan [1 ]
Huang, Ruolan [1 ]
Chang, Xiao [1 ]
Chen, Cuicui [2 ]
Li, Laiqing [2 ]
Zhang, Zhong [1 ,3 ]
Han, Yue [1 ]
机构
[1] Shenzhen Chinese Tradit Med Hosp, Dept ICU, Shenzhen, Peoples R China
[2] Guangzhou Youdi Biotechnol Co, Guangzhou, Guangdong, Peoples R China
[3] Shenzhen Chinese Tradit Med Hosp, Dept Cardiovasc, Shenzhen 518033, Guangdong, Peoples R China
来源
SLAS TECHNOLOGY | 2017年 / 22卷 / 02期
关键词
cardiac troponin; myoglobin; acute myocardial infarction; dual-label time-resolved fluorescence immunoassay; ACUTE MYOCARDIAL-INFARCTION; SERUM MYOGLOBIN; EARLY-DIAGNOSIS; FLUOROIMMUNOASSAY; ASSAY;
D O I
10.1177/2211068216680317
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The aim of this study was to establish a dual-label time-resolved fluorescence immunoassay (TRFIA) for the simultaneous determination of cardiac troponin T (cTnT) and myoglobin (MYO) for the early diagnosis of acute myocardial infarction. The sandwich immunoassay was used to detect the concentration of cTnT and MYO in serum. cTnT and MYO in serum were captured by anti-cTnT and anti-MYO antibodies immobilized on microtiter wells and then banded together with another anti-cTnT and anti-MYO labeled with europium(III) Sm3+ and samarium(III) Eu3+ chelates, followed by fluorescence measurement using time-resolved fluorometry. The performance of this TRFIA was evaluated using the clinical serum and compared with the commercial assays. The linear correlation coefficients (R-2) of the cTnT and MYO standard curves were 0.9993 and 0.9995, respectively. The sensitivity for cTnT detection was 2.21 pg/mL (linear dynamic range was 3.24-963.71 pg/mL), and the average recovery was 100.57%. The sensitivity for MYO detection was 3.24 ng/mL (linear dynamic range was 4.95-976.85 ng/mL), and the average recovery was 99.79%. High correlation coefficients (R-2) were obtained between the commercial assays and dual-label TRFIA (R-2 = 0.999). The present dual-label TRFIA has high sensitivity, specificity, and accuracy in clinical sample analysis. It is a good alternative to the single-label diagnostic methods.
引用
收藏
页码:130 / 135
页数:6
相关论文
共 50 条
  • [1] A DUAL-LABEL TIME-RESOLVED FLUORESCENCE IMMUNOASSAY FOR THE SIMULTANEOUS DETERMINATION OF β-HUMAN CHORIONIC GONADOTROPIN AND PROGESTERONE
    Sun, Wenqiao
    Ren, Shoulei
    Yuan, Guangwen
    Wang, Yan
    Li, Laiqing
    Zhang, Licheng
    ANALYTICAL LETTERS, 2014, 47 (18) : 2900 - 2908
  • [2] A dual-label time-resolved fluorescence immunoassay for the simultaneous determination of ferritin and β2-microglobulin
    Liu, Zhi
    Huang, Jing
    Ou, Rui-ming
    Yao, Meng-dong
    She, Yan-lin
    Chen, Rui
    Li, Cheng
    Xu, Li
    Abudureyimu, Aikebaier
    Zhang, Qing
    Liu, Shuang
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (06)
  • [3] Simultaneous detection of deoxynivalenol and zearalenone by dual-label time-resolved fluorescence immunoassay
    Zhang, Jue
    Gao, Lei
    Zhou, Bin
    Zhu, Lan
    Zhang, Yi
    Huang, Biao
    JOURNAL OF THE SCIENCE OF FOOD AND AGRICULTURE, 2011, 91 (02) : 193 - 197
  • [4] Simultaneous detection of sulfamethazine and sulfaquinoxaline using a dual-label time-resolved fluorescence immunoassay
    Le, Tao
    Yan, Peifeng
    Liu, Jin
    Wei, Shu
    FOOD ADDITIVES AND CONTAMINANTS PART A-CHEMISTRY ANALYSIS CONTROL EXPOSURE & RISK ASSESSMENT, 2013, 30 (07): : 1264 - 1269
  • [5] A dual-label time-resolved fluorescence immunoassay for the simultaneous determination of cystatin C and β2-microglobulin in urine
    Liu, Y.
    Liu, Y. H.
    Bei, W. J.
    Wang, K.
    Cui, T. T.
    Li, H. L.
    Wu, D. X.
    Chen, S. Q.
    Tan, N.
    Chen, J. Y.
    BRITISH JOURNAL OF BIOMEDICAL SCIENCE, 2017, 74 (04) : 193 - 197
  • [6] A dual-label time-resolved fluorescence immunoassay for the simultaneous determination of carcinoembryonic antigen and squamous cell carcinoma antigen
    Mao, Qian
    Liu, Xiaohong
    Chen, CuiCui
    Ye, Jingwen
    Liang, Huankun
    Li, Binxian
    Sun, Xinyi
    Sun, Lin
    Sun, Yuman
    Mu, Xiaomei
    Li, Laiqing
    BIOTECHNOLOGY AND APPLIED BIOCHEMISTRY, 2018, 65 (06) : 816 - 821
  • [7] Simultaneous Detection of Prolactin and Growth Hormone Using a Dual-label Time-resolved Fluorescence Immunoassay
    Zhong, Shuhai
    Liang, Huankun
    Peng, Fenglan
    Lu, Yanhong
    Liu, Tiancai
    Kulchytski, Uladzimir
    Dong, Wenqi
    JOURNAL OF FLUORESCENCE, 2024, 35 (3) : 1417 - 1422
  • [8] Simultaneous detection of two verotoxin genes using dual-label time-resolved fluorescence immunoassay with duplex PCR
    Watanabe, K
    Arakawa, H
    Maeda, M
    LUMINESCENCE, 2002, 17 (02) : 123 - 129
  • [9] DUAL-LABEL TIME-RESOLVED FLUOROIMMUNOASSAY FOR SIMULTANEOUS DETECTION OF MYOGLOBIN AND CARBONIC ANHYDRASE-III IN SERUM
    VUORI, J
    RASI, S
    TAKALA, T
    VAANANEN, K
    CLINICAL CHEMISTRY, 1991, 37 (12) : 2087 - 2092
  • [10] A Dual-Label Time-Resolved Fluorescence Immunoassay for the Simultaneous Qualitative/Quantitative Detection of Hepatitis A and Hepatitis E Virus IgM Antibodies
    Sun, Wenqiao
    Chen, Cuicui
    Zhong, Shuhai
    Liang, Huankun
    Guo, Guiling
    Li, Laiqing
    Zhang, Licheng