The Stability and Measuring Technology of the Maximum Optical Path Difference of Photo-Elastic Modulator Interferograms

被引:2
|
作者
Zhang Min-juan [3 ]
Wang Zhi-bin [1 ,2 ,3 ]
Li Xiao [3 ]
Li Jin-hua [3 ]
Wang Yan-chao [3 ]
机构
[1] North Univ China, Minist Educ, Key Lab Instrument Sci & Dynam Measurement, Taiyuan 030051, Peoples R China
[2] North Univ China, Natl Key Lab Elect Measurement Technol, Taiyuan 030051, Peoples R China
[3] North Univ China, Engn & Technol Res Ctr Shanxi Prov Opt Elect Info, Taiyuan 030051, Peoples R China
关键词
Photo-elastic modulator Fourier transform spectrometers (PEM-FTS); Interferograms; The maximum optical path difference; Zero-crossing counting; SPECTROMETRY;
D O I
10.3964/j.issn.1000-0593(2015)05-1436-04
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In order to improve the accuracy and stability of the rebuilt spectrums, it is necessary that stability analysis and nicety measuring of the maximum optical path difference of interferograms in the photo-elastic modulator Fourier transform spectrometers(PEM-FTS). The maximum optical difference of interferograms is uncertain parameter, and it is relate to the resonant state, characteristic of frequency-thermal drift and driving voltage of PEM. Therefore, based on the principle of photo-elastic modulator Fourier transform interferometer, the model of the freguency-thermal drift is built, and the variety of the maximum optical path difference is analyzed; A measuring method of the maximum optical path difference is put forward, which is zero-crossing counting of laser's interference signal when the driving signal of PEM is as the standard. In the method the dual channel highspeed comparator and FPGA are used to transform sine wave to square wave, to realize zero-crossing trigger counting and errors compensation. On the condition that the 670.8 nm laser is as the power source to produce the reference interferograms by the PEM interferometer, the 77.471 tan maximum optical path difference could be measured by the zero-crossing counting. the measuring errors is less than 0.167 nm, the rebuilt spectral peak wavelength errors of the infrared blackbody is less than 2 nm. the result is content with PEM-FTS.
引用
收藏
页码:1436 / 1439
页数:4
相关论文
共 10 条
  • [1] Buican T N, 2004, 24 ARM SCI C P OR FL
  • [2] Buican T N, 2010, US Patent, Patent No. 02624121
  • [3] Birefringence interferometers for ultra-high-speed FT spectrometry and hyperspectral imaging - I. Dynamic model of the resonant photoelastic modulator
    Buican, Tudor N.
    [J]. VIBRATIONAL SPECTROSCOPY, 2006, 42 (01) : 51 - 58
  • [4] Ultra-rapid-scanning Fourier transform infrared spectrometry
    Griffiths, PR
    Hirsche, BL
    Manning, CJ
    [J]. VIBRATIONAL SPECTROSCOPY, 1999, 19 (01) : 165 - 176
  • [5] Retardation correction for photoelastic modulator-based multichannel reflectance difference spectroscopy
    Hu, C. G.
    Sun, L. D.
    Li, Y. N.
    Hohage, M.
    Flores-Camacho, J. M.
    Hu, X. T.
    Zeppenfeld, P.
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2008, 25 (06) : 1240 - 1245
  • [6] Li C S, 2011, APPL OPTICS, V50, P5315
  • [7] Rugged high speed rotary imaging Fourier Transform Spectrometer for industrial use
    Wadsworth, W
    Dybwad, JP
    [J]. VIBRATIONAL SPECTROSCOPY-BASED SENSOR SYSTEMS, 2002, 4577 : 83 - 88
  • [8] WANG Ming-yan, 2012, ACTA OPT SINICA, V32
  • [9] WANG Yan-chao, 2013, SPECTROSC SPECT ANAL, V33, P1439
  • [10] ZHANG Min-juan, 2013, CHINESE J LASER, V40, P1