A Comparison of the Immunochemical Methods, PETINIA and EMIT, With That of HPLC-UV for the Routine Monitoring of Mycophenolic Acid in Heart Transplant Patients

被引:0
|
作者
Kunicki, Pawel K. [1 ]
Pawinski, Tomasz [2 ]
Boczek, Aleksandra [1 ]
Was, Joanna [1 ]
Bodnar-Broniarczyk, Magdalena [2 ]
机构
[1] Inst Cardiol, Dept Clin Biochem, Clin Pharmacol Unit, PL-04628 Warsaw, Poland
[2] Med Univ Warsaw, Fac Pharm, Dept Drug Chem, Warsaw, Poland
关键词
mycophenolic acid; PETINIA; EMIT; HPLC-UV; therapeutic drug monitoring; PERFORMANCE LIQUID-CHROMATOGRAPHY; ENZYME-MULTIPLIED IMMUNOASSAY; TANDEM MASS-SPECTROMETRY; SOLID-ORGAN TRANSPLANTATION; GLUCURONIDE METABOLITES; HUMAN PLASMA; ASSAY; RECIPIENTS; MOFETIL; MPA;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The aim of this study was to evaluate particle enhanced turbidimetric inhibition immunoassay (PETINIA) recently developed for mycophenolic acid (MPA) determination in plasma and to compare it with a reference high-performance liquid chromatography (HPLC) method, using samples from heart transplant recipients. The results are presented in the context of PETINIA being compared with enzyme multiplied immunoassay technique (EMIT). Methods: PETINIA evaluation was performed using 194 routine trough plasma samples at steady state. EMIT was evaluated using 677 samples from 61 steady-state 12-hour profiles obtained from 35 heart transplant patients. Evaluation was undertaken on a Dimension EXL 200 analyzer (PETINIA) and on a Viva-E analyzer (EMIT). Results: The mean MPA concentration measured by PETINIA was significantly higher than that measured by high-performance liquid chromatography combined with UV detector (2.36 +/- 1.30 mcg/mL versus 1.82 +/- 1.23 mcg/mL, respectively, P < 0.0001). Bland-Altman analysis revealed a mean bias of 0.54 mcg/mL [95% confidence interval (CI), 0.49-0.59] comprising 33.48% (95% CI, 30.34-36.61). Passing-Bablok regression was: y = 1.100x + 0.38 (95% CI for slope: 1.044-1.154 and for intercept: 0.30-0.47). Regardless of a significant observed correlation (r(2) = 0.9230, P, 0.0001), the statistical analyses showed a significant difference between PETINIA and the reference chromatographic method. The mean MPA concentration measured by EMIT was significantly higher than that measured by HPLC (7.48 +/- 8.34 mcg/mL versus 5.57 +/- 6.61 mcg/mL, respectively, P < 0.0001) with a mean bias of 1.91 mcg/mL (95% CI, 1.75-2.07) comprising 35.91% (95% CI, 34.37-37.45). The significant difference between EMIT and HPLC was confirmed by Passing-Bablok regression: y = 1.300x + 0.24 (95% CI for slope: 1.279-1.324 and for intercept: 0.18-0.29). The analysis of the determinations, grouped by sampling time, revealed positive bias between EMIT and HPLC ranging from 24.54% to 42.77% and inversely proportional to MPA concentrations with r(2) = 0.9122 (P < 0.001). Conclusions: The new immunochemical PETINIA method was associated with significantly higher MPA concentrations in routine therapeutic drug monitoring samples from heart transplant patients. The magnitude of the MPA overestimation was similar to that observed by use of the EMIT method.
引用
收藏
页码:311 / 318
页数:8
相关论文
共 25 条
  • [1] Comparison of petinia and emit immunoassay methods to measure mycophenolic acid concentration in transplanted patients
    Bartoli, A. N.
    Pitta, L.
    Villani, A.
    Burroni, D.
    Molinaro, M.
    THERAPEUTIC DRUG MONITORING, 2011, 33 (04) : 537 - 537
  • [2] Mycophenolic acid concentrations determined by EMIT and HPLC in heart transplant patients
    Kunicki, PK
    Pawinski, T
    THERAPEUTIC DRUG MONITORING, 2005, 27 (02) : 232 - 232
  • [3] Comparison of mycophenolic acid concentrations determined by a new PETINIA assay on the Dimension EXL analyzer and a HPLC-UV method
    Dasgupta, Amitava
    Tso, Gertie
    Chow, Loretta
    CLINICAL BIOCHEMISTRY, 2013, 46 (7-8) : 685 - 687
  • [4] Simple HPLC-UV Method for Mycophenolic Acid (MPA) Monitoring in Plasma
    Kunicki, P. K.
    Boczek, A.
    Was, J.
    THERAPEUTIC DRUG MONITORING, 2013, 35 (05) : 731 - 731
  • [5] HPLC-UV assay for monitoring total and unbound mycophenolic acid concentrations in children
    Zeng, L.
    Nath, C. E.
    Shaw, P. J.
    Earl, J. W.
    McLachlan, A. J.
    BIOMEDICAL CHROMATOGRAPHY, 2009, 23 (01) : 92 - 100
  • [6] Emit and FPIA for cyclosporine A routine monitoring in heart transplant patients
    Kunicki, PK
    Sobieszczanska-Malek, M
    THERAPEUTIC DRUG MONITORING, 2003, 25 (04) : 512 - 512
  • [7] CEDIA® mycophenolic acid assay compared with HPLC-UV in specimens from transplant recipients
    Westley, Ian S.
    Ray, John E.
    Morris, Raymond G.
    THERAPEUTIC DRUG MONITORING, 2006, 28 (05) : 632 - 636
  • [8] Mycophenolic acid AUC estimation: comparison of HPLC and EMIT assay data in pediatric renal transplant patients with tacrolimus comedication
    Seger, C.
    Jungraithmayr, T. C.
    Steimer, W.
    Zimmerhackl, L. B.
    Griesmacher, A.
    THERAPEUTIC DRUG MONITORING, 2011, 33 (04) : 539 - 539
  • [9] Comparative Routine Therapeutic Drug Monitoring of Mycophenolic Acid in human Plasma with HPLC-UV and Isotope Dilution LC-MS/MS
    Paal, Michael
    Habler, Katharina
    Northoff, Bernd
    Vogeser, Michael
    CLINICAL LABORATORY, 2020, 66 (04) : 613 - 620
  • [10] Comparison of the Emit immunoassay with HPLC for therapeutic drug monitoring of mycophenolic acid in pediatric renal-transplant recipients on mycophenolate mofetil therapy
    Weber, LT
    Shipkova, M
    Armstrong, VW
    Wagner, N
    Schütz, E
    Mehls, O
    Zimmerhackl, LB
    Oellerich, M
    Tönshoff, B
    CLINICAL CHEMISTRY, 2002, 48 (03) : 517 - 525