Quantification of HIV-RNA from dried blood spots using the Siemens VERSANT® HIV-1 RNA (kPCR) assay

被引:13
|
作者
Pirillo, Maria Franca [1 ]
Recordon-Pinson, Patricia [2 ]
Andreotti, Mauro [1 ]
Mancini, Maria Grazia [1 ]
Amici, Roberta [1 ]
Giuliano, Marina [1 ]
机构
[1] Ist Super Sanita, Dept Therapeut Res & Med Evaluat, I-00161 Rome, Italy
[2] Victor Segalen Univ, Virol Lab, Bordeaux, France
关键词
resource-limited settings; real-time PCR; automated assays; VIRAL LOAD; SPECIMENS;
D O I
10.1093/jac/dkr383
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Objectives: Simplified methods for virological monitoring in resource-limited settings are increasingly needed. We evaluated the performance of the VERSANT (R) HIV-1 RNA (kPCR) assay for the determination of HIV-1 viral load from dried blood spots (DBS). Assay sensitivity and correlation with plasma quantification values were assessed. Methods: A total of 98 DBS were prepared from fresh blood samples of HIV-infected patients. DBS were kept at room temperature for 6 weeks or 7 months before processing while the corresponding plasma samples were stored at -80 degrees C. DBS were first pre-treated in a special DBS buffer. The DBS extracts and the plasma samples were then purified and amplified using the VERSANT assay reagents. Results: In the first series of tests, performed after 6 weeks of storage, there was good correlation between quantification of viral load in plasma and in DBS (r=0.95, P<0.001). The detection rate in DBS was 100% when plasma levels were >1000 copies/mL. The sensitivity and specificity of the DBS assay were 88.2% [95% confidence interval (CI) 79.4-93.6] and 69.2% (95% CI 42.0-87.4), respectively. Using the 5000 copies/mL threshold (defining virological failure in resource-limited settings), both positive and negative predictive values were high (95.2% and 87.5%, respectively). After 7 months of storage there was a modest decrease in the detection rate and less significant correlations for samples with HIV-RNA <5000 copies/mL. Conclusions: Quantification of HIV-RNA from DBS by the VERSANT automated sample preparation and detection method can be used to diagnose virological failure in HIV-positive patients.
引用
收藏
页码:2823 / 2826
页数:4
相关论文
共 50 条
  • [1] Method Comparison between the Siemens VERSANT® HIV-1 RNA 1.5 Assay (kPCR)* and the Abbott REALTIME HIV-1 Assay
    Monga, D.
    Wagner, C.
    Gorrin, G.
    Baublitz, Y.
    Kritikos, G.
    Mosner, J.
    JOURNAL OF CLINICAL VIROLOGY, 2015, 70 : S101 - S101
  • [2] Quantification of HIV-1 RNA on dried blood spots
    Johannessen, Asgeir
    AIDS, 2010, 24 (03) : 475 - 476
  • [3] Evaluation of the new Versant HIV-1 RNA 1.0 ASSAY (KPCR) for quantification of human immunodeficiency virus type 1 RNA
    Amendola, A.
    Zaccaro, P.
    Sabatini, R.
    Bibbo, A.
    Bloisi, M.
    Patella, E.
    Natale, A.
    Rossi, G.
    Capobianchi, M. R.
    INFECTION, 2009, 37 : 94 - 94
  • [4] Multicenter Evaluation of the Versant® HIV 1 RNA 1 0 Assay (kPCR) with the Versant® kPCR Molecular System
    Poshusta, T. L.
    Schmidt, D. M.
    Yao, J. D.
    Germer, J. J.
    Mitchell, P. S.
    Taylor, P.
    Mistry, H.
    Nutter, P.
    Turczyn, J.
    Fisher, L.
    del Rosario, L.
    Hennig, M.
    Martinez, A.
    Li, H.
    Surtihadi, J.
    Molino, G.
    Nguyen, V.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2010, 12 (06): : 889 - 889
  • [5] Use of dried whole blood spots for quantification of HIV-1 RNA
    O'Shea, S
    Mullen, J
    Corbett, K
    Chrystie, I
    Newell, ML
    Banatvala, JE
    AIDS, 1999, 13 (05) : 630 - 631
  • [6] Method Comparison Between the VERSANT® HIV-1 RNA 1.0 Assay (kPCR) and the Abbott RealTime™ HIV-1 Assay
    Gonin, G. M.
    Yao, J. D.
    Wagner, C.
    Lin, C.
    Swiatkowski, P.
    Tam, S. M.
    Canchola, J.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2008, 10 (06): : 587 - 587
  • [7] Fresh Sample Study and Method Comparison for VERSANT HIV-1 RNA 1.0 Assay (kPCR)
    Wagner, C.
    Gorrin, G. M.
    Canchola, J.
    Lin, C.
    Huang, H.
    Nicol, S.
    Gipson, G.
    Magaya, L.
    Surtihadi, J.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2009, 11 (06): : 647 - 647
  • [8] Quantitation of HIV-1 RNA in dried blood and plasma spots
    Reigadas, S.
    Schrive, M. H.
    Aurillac-Lavignolle, V.
    Fleury, H. J.
    JOURNAL OF VIROLOGICAL METHODS, 2009, 161 (01) : 177 - 180
  • [9] Use of dried blood spots with Abbott RealTime HIV-1 quantitative RNA assay
    Huang, S.
    Erickson, B.
    Abravaya, K.
    Robinson, J.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2007, 9 (05): : 700 - 700
  • [10] Evaluation of the new VERSANT HIV-1 RNA 1.0 Assay (kPCR) for quantitative detection of human immunodeficiency virus type 1 RNA
    Troppan, Katharina T.
    Stelzl, Evelyn
    Violan, Deborah
    Winkler, Michaela
    Kessler, Harald H.
    JOURNAL OF CLINICAL VIROLOGY, 2009, 46 (01) : 69 - 74