Supersonic helium beam diagnostic for fluctuation measurements of electron temperature and density at the Tokamak TEXTOR

被引:14
|
作者
Kruezi, U. [1 ]
Stoschus, H. [1 ]
Schweer, B. [1 ]
Sergienko, G. [1 ]
Samm, U. [1 ]
机构
[1] Forschungszentrum Julich, Inst Energy & Climate Res, Assoc EURATOM FZJ, Trilateral Euregio Cluster, D-52425 Julich, Germany
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2012年 / 83卷 / 06期
关键词
PLASMA EDGE DIAGNOSTICS; ANOMALOUS TRANSPORT; MOLECULAR-BEAMS; FUSION PLASMAS; ATOMIC-BEAMS;
D O I
10.1063/1.4707150
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A supersonic helium beam diagnostic, based on the line-ratio technique for high resolution electron density and temperature measurements in the plasma edge (r/a > 0.9) was designed, built, and optimised at TEXTOR (Torus Experiment for Technology Oriented Research). The supersonic injection system, based on the Campargue skimmer-nozzle concept, was developed and optimised in order to provide both a high neutral helium beam density of n(0) = 1.5 x 10(18) m(-3) and a low beam divergence of +/- 1 degrees. simultaneously, achieving a poloidal resolution of Delta(poloidal) = 9 mm. The setup utilises a newly developed dead volume free piezo valve for operation in a high magnetic field environment of up to 2 T with a maximum repetition rate of 80 Hz. Gas injections are realised for a duration of 120 ms at a repetition rate of 2 Hz (duty cycle 1/3). In combination with a high sensitivity detection system, consisting of three 32 multi-channel photomultipliers (PMTs), measurements of edge electron temperature and density with a radial resolution of Delta(radial) = 2 mm and a maximum temporal resolution of Delta t similar or equal to 2 mu s (470 kHz) are possible for the first time. The diagnostic setup at TEXTOR is presented. The newly developed injection system and its theoretical bases are discussed. The applicability of the stationary collisional-radiative model as basis of the line-ratio technique is shown. Finally, an example of a fluctuation analysis demonstrating the unique high temporal and spatial resolution capabilities of this new diagnostic is presented. [http://dx.doi.org/10.1063/1.4707150]
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Millimeter wave Imaging diagnostic for simultaneous density and temperature fluctuation measurements on the TEXTOR Tokamak
    Donné, AJH
    Classen, IGJ
    Domier, CW
    Johnson, M
    Luhmann, NC
    Mazzucato, E
    Munsat, T
    Park, H
    van de Pol, MJ
    Wang, J
    Xia, Z
    CONFERENCE DIGEST OF THE 2004 JOINT 29TH INTERNATIONAL CONFERENCE ON INFRARED AND MILLIMETER WAVES AND 12TH INTERNATIONAL CONFERENCE ON TERAHERTZ ELECTRONICS, 2004, : 257 - 258
  • [2] Thermal helium beam for diagnostic of electron density and temperature
    Kornejew, P
    Bohmeyer, W
    Fussmann, G
    DIAGNOSTICS FOR EXPERIMENTAL THERMONUCLEAR FUSION REACTORS, 1996, : 599 - 602
  • [3] Status of electron temperature and density measurement with beam emission spectroscopy on thermal helium at TEXTOR
    Schmitz, O.
    Beigman, I. L.
    Vainshtein, L. A.
    Schweer, B.
    Kantor, M.
    Pospieszczyk, A.
    Xu, Y.
    Krychowiak, M.
    Lehnen, M.
    Samm, U.
    Unterberg, B.
    PLASMA PHYSICS AND CONTROLLED FUSION, 2008, 50 (11)
  • [4] Simultaneous microwave imaging system for density and temperature fluctuation measurements on TEXTOR (invited)
    Park, H
    Mazzucato, E
    Munsat, T
    Domier, CW
    Johnson, M
    Luhmann, NC
    Wang, J
    Xia, Z
    Classen, IGJ
    Donné, AJH
    van de Pol, MJ
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3787 - 3792
  • [5] DENSITY FLUCTUATION MEASUREMENTS IN TOKAMAK PLASMAS
    SURKO, CM
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1978, 23 (07): : 832 - 832
  • [6] ELECTRON-TEMPERATURE FLUCTUATION MEASUREMENTS WITH HIGH TEMPORAL RESOLUTION IN THE TEXTOR BOUNDARY PLASMA
    IVANOV, RS
    CHERNILEVSKY, AN
    SAMM, U
    VANNIEUWENHOVE, R
    VANOOST, G
    PHYSICS LETTERS A, 1993, 173 (03) : 288 - 292
  • [7] Electron temperature fluctuation measurements and techniques in the DIII-D tokamak
    Rettig, CL
    Peebles, WA
    Lohr, J
    Austin, ME
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (01): : 484 - 487
  • [8] Electron density and q profile measurements with the far-IR interferometer polarimeter on the TEXTOR tokamak
    Koslowski, HR
    Soltwisch, H
    FUSION ENGINEERING AND DESIGN, 1997, 34-35 : 143 - 150
  • [9] EDGE FLUCTUATION MEASUREMENTS IN TEXTOR WITH A THERMAL NEUTRAL BEAM PROBE
    KOMORI, A
    MASAMUNE, S
    MITARAI, O
    YAMAGIWA, K
    HONDA, C
    KADOTA, K
    FUJITA, J
    LIE, YT
    SAMM, U
    JOURNAL OF NUCLEAR MATERIALS, 1989, 162 : 231 - 235
  • [10] Electron density and temperature measurements in TEXTOR ion cyclotron wall conditioning plasmas by thermal Li beam spectroscopy
    Laengner, R.
    Unterberg, B.
    Lyssoivan, A.
    Paul, M.
    Philipps, V.
    Pospieszczyk, A.
    Samm, U.
    Schmitz, O.
    Schweer, B.
    Sergienko, G.
    Vervier, M.
    JOURNAL OF NUCLEAR MATERIALS, 2011, 415 (01) : S1166 - S1169