The Importance of Correlation between CBCT Analysis of Bone Density and Primary Stability When Choosing the Design of Dental Implants-Ex Vivo Study

被引:8
|
作者
Mikic, Mirko [1 ]
Vlahovic, Zoran [2 ]
Stevanovic, Momir [3 ]
Arsic, Zoran [2 ]
Mladenovic, Rasa [3 ]
机构
[1] Univ Montenegro, Fac Med, Dept Dent, Podgorica 81000, Montenegro
[2] Univ Pristina, Fac Med, Dept Dent, Kosovska Mitrovica 38220, Serbia
[3] Univ Kragujevac, Fac Med Sci, Dept Dent, Kragujevac 34000, Serbia
关键词
bone density; CBCT; primary stability; osseointegration; dental implants; BEAM COMPUTED-TOMOGRAPHY; RESONANCE FREQUENCY-ANALYSIS; INSERTION TORQUE; FOLLOW-UP; EXTRACTION; SURVIVAL; PLACEMENT; IMMEDIATE; REASONS; SURFACE;
D O I
10.3390/tomography8030107
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study aims to determine the correlation between the mean value of bone density measured on the CBCT device and the primary stability of dental implants determined by resonant frequency analysis. An experimental study was conducted on a material of animal origin: bovine femur and pig ribs. Two types of implants were used in this study: self-tapping and non-self-tapping of the same dimensions. Results of the experimental study showed a statistically significant correlation between bone density expressed in HU units and the primary stability of self-tapping and non-self- tapping dental implants expressed in ISQ units in bovine femur bones and self-tapping implants and pig rib bones. There was no statistically significant correlation between non-self-tapping dental implants in pig rib bones. Self-tapping and non-self-tapping implants did not show statistical significance in the primary stability in bones of different qualities. The analysis of bone density from CBCT images in the software of the apparatus expressed in HU units can be used to predict the degree of primary stability of self-tapping and non-self-tapping dental implants in bones of densities D1 and D2, and self-tapping dental implants in bones of the lower quality D4.
引用
收藏
页码:1293 / 1306
页数:14
相关论文
共 24 条
  • [1] Correlation between Primary Stability of Dental Implants and Bone Density: A Retrospective Analysis
    Padmaja S.
    Rajasekar A.
    Journal of Long-Term Effects of Medical Implants, 2024, 34 (04) : 65 - 70
  • [2] Evaluation of the Correlation Between the Structural Parameters of Trabecular Bone in CBCT and the Primary Stability of Dental Implants
    Shokri, Abbas
    Armand, Nafiseh
    Shahabi, Shiva
    Torkzaban, Parviz
    Tapak, Leili
    Poormoradi, Banafsheh
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2024, 39 (02) : 271 - 277
  • [3] The correlation of CBCT analysis derived bone density parameters with primary implant stability - a clinical study
    Mikic, Mirko
    Vlahovic, Zoran
    Nenezic, Dragoslav
    Videnovic, Goran
    Ilic, Danijela
    Mladenovic, Rasa
    VOJNOSANITETSKI PREGLED, 2021, 78 (08) : 811 - 817
  • [4] Can the Bone Density Estimated by CBCT Predict the Primary Stability of Dental Implants? A New Measurement Protocol
    Al-Jamal, Mustafa Faisal J.
    Al-Jumaily, Hassanien A.
    JOURNAL OF CRANIOFACIAL SURGERY, 2021, 32 (02) : E171 - E174
  • [5] Ex vivo comparison of drilling techniques for optimizing primary stability of zirconia dental implants in different bone densities
    Kawe Sagheb
    Senem Yildirimturk
    Sebahat Kaya
    Shengchi Fan
    Marius Morlock
    Keyvan Sagheb
    International Journal of Implant Dentistry, 11 (1)
  • [6] Is clinical experience important for obtaining the primary stability of dental implants with aggressive threads? An ex vivo study
    Geckili, Onur
    Bilhan, Hakan
    Geckili, Esma
    Barca-Dayan, Ece
    Dayan, Cagatay
    Bural, Canan
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2019, 24 (02): : E254 - E259
  • [7] Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test
    Bayarchimeg, Dorjpalam
    Namgoong, Hee
    Kim, Byung Kook
    Kim, Myung Duk
    Kim, Sungtae
    Kim, Tae-Il
    Seol, Yang Jo
    Lee, Yong Moo
    Ku, Young
    Rhyu, In-Chul
    Lee, Eun Hee
    Koo, Ki-Tae
    JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, 2013, 43 (01): : 30 - 36
  • [8] The Influence of Implant Shape on Primary Stability of Implants With a Thread Cutting and Forming Design: An Ex Vivo Study
    Dayan, Cagatay
    Geckili, Onur
    Bural, Canan
    JOURNAL OF ORAL IMPLANTOLOGY, 2019, 45 (03) : 181 - 185
  • [9] Relationship between the bone density estimated by cone-beam computed tomography and the primary stability of dental implants
    Isoda, Kei
    Ayukawa, Yasunori
    Tsukiyama, Yoshihiro
    Sogo, Motofumi
    Matsushita, Yasuyuki
    Koyano, Kiyoshi
    CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (07) : 832 - 836
  • [10] Ex Vivo Analysis of Ability of Osseodensification to Improve Dental Implant Primary Stability Using Xenograft Bone Walls
    Ferreira, Henrique
    Mourao, Carlos Fernando
    Mello-Machado, Rafael Coutinho
    Javid, Kayvon
    Pereira, Rodrigo dos Santos
    Senna, Plinio Mendes
    APPLIED SCIENCES-BASEL, 2023, 13 (22):