Characterization of Seamless CdTe Photon Counting X-Ray Detector

被引:2
|
作者
Kauppinen, Matti [1 ,2 ]
Winkler, Alex [2 ,3 ]
Lamsa, Vili [2 ]
Matikkala, Mikko [2 ]
Zoladz, Miroslaw [4 ]
Grybos, Pawel [4 ]
Kleczek, Rafal [4 ]
Kmon, Piotr [4 ]
Szczygiel, Robert [4 ]
Fabritius, Tapio [1 ]
机构
[1] Fac Informat Technol & Elect Engn, Res Unit Optoelect & Measurement Tech, Oulu 90590, Finland
[2] Detect Technol Oyj, Oulu 90590, Finland
[3] Univ Helsinki, Helsinki Inst Phys, Fac Phys, Helsinki 00014, Finland
[4] AGH Univ Sci & Technol, Dept Measurement & Elect, Fac Elect Engn, Krakow, Poland
关键词
ASIC; cadmium telluride (CdTe); direct conversion; material discrimination; multienergy; photon counting detector; X-ray imaging; PIXEL DETECTORS; POLARIZATION; CALIBRATION;
D O I
10.1109/TIM.2021.3070615
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Spectrally selective X-ray imaging provides improved material and tissue discrimination in comparison with the state-of-the-art dual energy technologies that are commonly used in medical, industrial, and security applications. Cadmium telluride (CdTe)- and cadmium zinc telluride (CdZnTe)-based line scanners and small size two-dimensional X-ray sensors are emerging to the market, but the need for large-scale panels is axiomatic. In this study, a seamless CdTe tile was developed that enables the implementation of large-sized, energy selective X-ray detector panels. The developed tile consists of a 64 x 64 pixel array (with 150 mu m pitch) with a necessary substrate, ASIC, and CdTe crystal. The performance of the constructed seamless tile was characterized by focusing on spectral resolution and stability. In addition, a simple pixel trimming method that automates the equalization of each energy selective pixel was developed and analyzed. The obtained results suggest that the proposed concept of seamless (tileable) detector structures is a feasible approach to scale up panel sizes. The seamless tile shows comparable spectral resolution and stability performance with commercial CdTe sensors. The effect of tile to tile variation, the realization of a large-scale panel, as well as the charge sharing performance were left out of the scope and are to be studied in the next phase.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Tutorial on X-ray photon counting detector characterization
    Ren, Liqiang
    Zheng, Bin
    Liu, Hong
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2018, 26 (01) : 1 - 28
  • [2] Experimental characterization of an X-ray photon counting detector
    Wang, X.
    Wei, T.
    Deng, Z.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (07)
  • [3] Characterization of a X-ray pixellated CdTe detector with TIMEPIX photon-counting readout chip
    Ruat, Marie
    Ponchut, Cyril
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4799 - 4803
  • [4] Characteristic evaluation of a novel CdTe photon counting detector for X-ray imaging
    Cho, Hyo-Min
    Kim, Hee-Joung
    Ryu, Hyun-Ju
    Choi, Yu-Na
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 63 (01) : 120 - 127
  • [5] Characteristic evaluation of a novel CdTe photon counting detector for X-ray imaging
    Hyo-Min Cho
    Hee-Joung Kim
    Hyun-Ju Ryu
    Yu-Na Choi
    Journal of the Korean Physical Society, 2013, 63 : 120 - 127
  • [6] Characterization of a Pixelated CdTe X-Ray Detector Using the Timepix Photon-Counting Readout Chip
    Ruat, Marie
    Ponchut, Cyril
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (05) : 2392 - 2401
  • [7] Spectral X-ray phase contrast imaging with a CdTe photon-counting detector
    Navarrete, Carlos
    Procz, Simon
    Schuetz, Michael Kilian
    Roque, Gerardo
    Fey, Julian
    Avila, Carlos
    Olivo, Alessandro
    Fiederle, Michael
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 971 (971):
  • [8] A counting CdTe pixel detector for hard X-ray and γ-ray imaging
    Fischer, P
    Kouda, M
    Krüger, H
    Lindner, M
    Sato, G
    Takahashi, T
    Watanabe, S
    Wermes, N
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (06) : 2401 - 2404
  • [9] Analysis of artifact with X-ray CT using energy band by photon counting CdTe detector
    Matsumoto, G.
    Imura, Y.
    Morii, H.
    Miyake, A.
    Aoki, T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 292 - 294
  • [10] Characterization of Pixirad-1 photon counting detector for X-ray imaging
    Delogu, P.
    Oliva, P.
    Bellazzini, R.
    Brez, A.
    de Ruvo, P. L.
    Minuti, M.
    Pinchera, M.
    Spandre, G.
    Vincenzi, A.
    JOURNAL OF INSTRUMENTATION, 2016, 11