Point-of-care testing of MicroRNA based on personal glucose meter and dual signal amplification to evaluate drug-induced kidney injury

被引:19
|
作者
Huang, Xitong [1 ]
Li, Jiwei [1 ]
Lu, Mi [1 ]
Zhang, Wangning [1 ]
Xu, Zhiming [1 ]
Yu, Bo-Yang [1 ]
Tian, Jiangwei [1 ]
机构
[1] China Pharmaceut Univ, Sch Tradit Chinese Pharm, Res Ctr Traceabil & Standardizat TCMs, State Key Lab Nat Med,Jiangsu Key Lab TCM Evaluat, Nanjing 211198, Peoples R China
基金
中国国家自然科学基金;
关键词
MicroRNA; DNAzyme; Point-of-care testing; Signal amplification; Drug-induced kidney injury; CISPLATIN-INDUCED NEPHROTOXICITY; ELECTROCHEMICAL DETECTION; QUANTITATIVE DETECTION; BIOMARKERS; ELECTROCHEMILUMINESCENCE; SENSORS; MIRNAS; PROBES;
D O I
10.1016/j.aca.2020.03.051
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The upregulation or downregulation of microRNA-21 (miRNA-21) is closely related with drug-induced kidney injury (DIKI). As a potential and significant biomarker, the point-of-care testing (POCT) of miRNA-21 is worthy of attention and can provide essential information for clinical diagnosis. Hence, we design a portable and sensitive POCT assay for miRNA-21 using personal glucose meters (PGM). The whole operational system is constructed on streptavidin-coated magnetic beads (MBs) modified with substrate strands linked invertase and DNAzyme molecules each silenced by a locking strand. In the presence of miRNA-21, the locking strand can hybridize to miRNA-21, which originates the activation of the DNAzyme. The DNAzyme cleaves the substrate strands and induces the release of invertase from the surface of MBs. The separated invertase hydrolyzes sucrose to glucose which can be measured by PGM. The dual enzyme mediated catalyzation by DNAzyme and invertase therefore triggers the signal amplification. We establish a linear relationship between PGM and different concentration of miRNA-21 in the range of 100 fM to 1 pM. The limit of detection is 68.08 fM, which is comparable with some of the previous reports. The biosensor also exhibits excellent sequence selectivity, well-presented reproducibility and stability. Notably, by detecting miRNA-21 in urine, this method has been successfully used to predict DIKI and evaluate the protection effect of drugs on DIKI. Therefore, a dependable and low-cost POCT strategy for the detection of miRNA-21 is established, which is promising to supply valuable information for drug screening and evaluation of DIKI. (C) 2020 Elsevier B.V. All rights reserved.
引用
收藏
页码:72 / 79
页数:8
相关论文
共 20 条
  • [1] Point-of-care testing of methotrexate using a controlled release sensor based on a personal glucose meter
    Yang, Liyu
    Yang, Siyi
    Zhang, Ting
    Zhao, Langyi
    Fa, Huanbao
    Wang, Yongzhong
    Huo, Danqun
    Hou, Changjun
    Yang, Mei
    ANALYTICAL METHODS, 2025, 17 (06) : 1345 - 1353
  • [2] MicroRNA-122 can be measured in capillary blood which facilitates point-of-care testing for drug-induced liver injury
    Vliegenthart, A. D. Bastiaan
    Berends, Cecile
    Potter, Carmelita M. J.
    Kersaudy-Kerhoas, Maiwenn
    Dear, James W.
    BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 2017, 83 (09) : 2027 - 2033
  • [3] Point-of-care testing of protein biomarkers by integrating a personal glucose meter with a concatenated DNA amplifier
    Shi, Hai
    Zheng, Ji
    Wang, Yanxia
    Zhu, Sha
    Xiang, Yang
    Zhu, Xiaoli
    Li, Genxi
    SENSORS AND ACTUATORS B-CHEMICAL, 2020, 322
  • [4] CRISPR Cas12a-based "sweet" biosensor coupled with personal glucose meter readout for the point-of-care testing of Salmonella
    Zhou, Chi
    Huang, Di
    Wang, Ziyi
    Shen, Peijie
    Wang, Pu
    Xu, Zhinan
    JOURNAL OF FOOD SCIENCE, 2022, 87 (09) : 4137 - 4147
  • [5] 3D origami electrochemical immunodevice for sensitive point-of-care testing based on dual-signal amplification strategy
    Ma, Chao
    Li, Weiping
    Kong, Qingkun
    Yang, Hongmei
    Bian, Zhaoquan
    Song, Xianrang
    Yu, Jinghua
    Yan, Mei
    Biosensors and Bioelectronics, 2015, 63 : 7 - 13
  • [6] 3D origami electrochemical immunodevice for sensitive point-of-care testing based on dual-signal amplification strategy
    Ma, Chao
    Li, Weiping
    Kong, Qingkun
    Yang, Hongmei
    Bian, Zhaoquan
    Song, Xianrang
    Yu, Jinghua
    Yan, Mei
    BIOSENSORS & BIOELECTRONICS, 2015, 63 : 7 - 13
  • [7] Detection of microRNA: A Point-of-Care Testing Method Based on a pH-Responsive and Highly Efficient Isothermal Amplification
    Feng, Chang
    Mao, Xiaoxia
    Shi, Hai
    Bo, Bing
    Chen, Xiaoxia
    Chen, Tianshu
    Zhu, Xiaoli
    Li, Genxi
    ANALYTICAL CHEMISTRY, 2017, 89 (12) : 6631 - 6636
  • [8] Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing
    Tang, ZP
    Louie, RF
    Lee, JH
    Lee, DM
    Miller, EE
    Kost, GJ
    CRITICAL CARE MEDICINE, 2001, 29 (05) : 1062 - 1070
  • [9] Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing.
    Tang, Z
    Louie, RF
    Lee, JH
    Lee, DM
    Miller, EE
    Kost, GJ
    CLINICAL CHEMISTRY, 2000, 46 (06) : A8 - A9
  • [10] A Microfluidic Paper-Based Electrochemical Immunosensor for Point-of-Care Testing of Kidney Injury Molecule-1
    Mao, Xinyue
    Cao, Wenting
    Yuan, Xinyan
    Liang, Xiao
    Zhang, Guanghui
    Wang, Yongzhong
    Jin, Xiaofeng
    IEEE SENSORS JOURNAL, 2024, 24 (22) : 36324 - 36333