Fracture of a ceramic liner in a total hip arthroplasty with a sandwich cup

被引:17
|
作者
Popescu, Dragos [1 ]
Gallart, X. [1 ]
Garcia, S. [1 ]
Bori, G. [1 ]
Tomas, X. [2 ]
Riba, J. [1 ]
机构
[1] Univ Barcelona, Dept Orthoped & Trauma Surg, Hip Unit, Hosp Clin, E-08036 Barcelona, Spain
[2] Univ Barcelona, Dept Radiol, Hosp Clin, E-08036 Barcelona, Spain
关键词
hip arthroplasty; liner fracture; complication;
D O I
10.1007/s00402-007-0536-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Alumina ceramic-on-ceramic total hip arthroplasty (THA) has been widely used due to its advantages such as low wear, scratch resistance, wettable surface and relatively low biological reactivity of the wear particles. Nevertheless, this material in THA still persists to be one of the major concerns about the risk of fracture, due to its brittleness. Many authors have reported a fracture of the ceramic head but few reported a fracture of the ceramic acetabular insert. In order to reduce the rigidity of the ceramic-on-ceramic coupling and prevent an impingement between the rim of the ceramic liner and the metal neck of the femoral stem, a modular acetabular component with a sandwich insertion (alumina/polyethylene/titanium) was proposed. We report the fracture of the ceramic acetabular liner of such a ceramic sandwich cup due to a slightly retroverted position of the cup that causes an impingement between the femoral stem and the rim of the insert. The fracture occurred 3 years after the operation without trauma. At revision the entire cup was replaced using a polyethylene liner without inner ceramic liner.
引用
收藏
页码:783 / 785
页数:3
相关论文
共 50 条
  • [1] Fracture of a ceramic liner in a total hip arthroplasty with a sandwich cup
    Dragos Popescu
    X. Gallart
    S. Garcia
    G. Bori
    X. Tomas
    J. Riba
    Archives of Orthopaedic and Trauma Surgery, 2008, 128 : 783 - 785
  • [2] Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup
    Hasegawa, M
    Sudo, A
    Hirata, H
    Uchida, A
    JOURNAL OF ARTHROPLASTY, 2003, 18 (05): : 658 - 661
  • [3] Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup
    Hwang, Deuk-Soo
    Kim, Young-Mo
    Lee, Chang-Hwan
    JOURNAL OF ARTHROPLASTY, 2007, 22 (03): : 468 - 471
  • [4] Revision total hip arthroplasty with sandwich-type ceramic on ceramic liner
    Lee, S. -H.
    Hwang, J. -H.
    Kim, B. -K.
    Hong, S. -H.
    Bioceramics and Alternative Bearings in Joint Arthroplasty, 2007, : 289 - 293
  • [5] Ceramic Liner Fracture in Total Hip Arthroplasty: A Case Report
    Tashtanov, Baikozho R.
    Korytkin, Andrey A.
    V. Pavlov, Vitalii
    Shubnyakov, Igor I.
    TRAVMATOLOGIYA I ORTOPEDIYA ROSSII, 2022, 28 (03): : 63 - 73
  • [6] Risk factors for ceramic liner fracture after total hip arthroplasty
    Traina, Francesco
    De Fine, Marcello
    Bordini, Barbara
    Toni, Aldo
    HIP INTERNATIONAL, 2012, 22 (06) : 607 - 614
  • [7] Bilateral fractures of alumina ceramic heads in total hip arthroplasty and subsequent ceramic re-fracture after revisional arthroplasty in sandwich ceramic cup
    Kee-Haeng Lee
    Yong-Sik Kim
    Suk-Ku Han
    European Journal of Orthopaedic Surgery & Traumatology, 2010, 20 : 415 - 418
  • [8] Bilateral fractures of alumina ceramic heads in total hip arthroplasty and subsequent ceramic re-fracture after revisional arthroplasty in sandwich ceramic cup
    Lee, Kee-Haeng
    Kim, Yong-Sik
    Han, Suk-Ku
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2010, 20 (05): : 415 - 418
  • [9] Preventing ceramic liner fracture after Delta ceramic-on-ceramic total hip arthroplasty
    Lee, Young-Kyun
    Lim, Jae-Young
    Ha, Yong-Chan
    Kim, Tae-Young
    Jung, Woon-Hwa
    Koo, Kyung-Hoi
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (07) : 1155 - 1162
  • [10] Preventing ceramic liner fracture after Delta ceramic-on-ceramic total hip arthroplasty
    Young-Kyun Lee
    Jae-Young Lim
    Yong-Chan Ha
    Tae-Young Kim
    Woon-Hwa Jung
    Kyung-Hoi Koo
    Archives of Orthopaedic and Trauma Surgery, 2021, 141 : 1155 - 1162