Mechatronic system for in-bore MRI-guided insertion of needles to the prostate: An in vivo needle guidance accuracy study

被引:13
|
作者
Cepek, Jeremy [1 ,2 ]
Lindner, Uri [1 ]
Ghai, Sangeet [3 ]
Louis, Alyssa S. [1 ]
Davidson, Sean R. H. [4 ]
Gertner, Mark [5 ]
Hlasny, Eugen [3 ]
Sussman, Marshall S. [3 ]
Fenster, Aaron [2 ,6 ]
Trachtenberg, John [1 ]
机构
[1] Univ Hlth Network, Div Urol, Dept Surg Oncol, Toronto, ON, Canada
[2] Univ Western Ontario, Robarts Res Inst, London, ON N6A 5B7, Canada
[3] Univ Hlth Network, Joint Dept Med Imaging, Toronto, ON, Canada
[4] Univ Hlth Network, Ontario Canc Inst, Toronto, ON, Canada
[5] Univ Hlth Network, Ontario Canc Inst, Div Biophys & Bioimaging, Toronto, ON, Canada
[6] Univ Western Ontario, Biomed Engn, London, ON N6A 5B7, Canada
基金
加拿大健康研究院;
关键词
needle guidance; prostate cancer; focal therapy; focal laser ablation therapy; targeted biopsy; FOCAL THERAPY; THERMAL THERAPY; CANCER; BRACHYTHERAPY; BIOPSY; SCANNER;
D O I
10.1002/jmri.24742
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
BackgroundTo present our experiences in initial clinical evaluation of a novel mechatronic system for in-bore guidance of needles to the prostate for MRI-guided prostate interventions in 10 patients. We report accuracy of this device in the context of focal laser ablation therapy for localized prostate cancer. MethodsAn MRI-compatible needle guidance device was developed for transperineal prostate interventions. Ten patients underwent MRI-guided focal laser ablation therapy with device-mediated laser fiber delivery. We recorded needle guidance error and needle delivery time. ResultsA total of 37 needle insertions were evaluated. Median needle guidance error was 3.5 mm (interquartile range, 2.1-5.4 mm), and median needle delivery time was 9 min (interquartile range, 6.5-12 min). ConclusionThis system provides a reliable method of accurately aligning needle guides for in-bore transperineal needle delivery to the prostate. J. Magn. Reson. Imaging 2015;42:48-55. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:48 / 55
页数:8
相关论文
共 50 条
  • [1] Design of an MRI-Compatible Modularized Needle Driver for In-Bore MRI-Guided Prostate Interventions
    Kim, Ki-Young
    Li, Meng
    Gonenc, Berk
    Shang, Weijian
    Eslami, Sohrab
    Iordachita, Iulian
    2015 15TH INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS), 2015, : 1520 - 1525
  • [2] MRI-guided prostate focal laser ablation therapy using a mechatronic needle guidance system
    Cepek, Jeremy
    Lindner, Uri
    Ghai, Sangeet
    Davidson, Sean R. H.
    Trachtenberg, John
    Fenster, Aaron
    MEDICAL IMAGING 2014: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2014, 9036
  • [3] Comparison of analgesic techniques in MRI-guided in-bore prostate biopsy
    Quentin, M.
    Arsov, C.
    Ullrich, T.
    Valentin, B.
    Hiester, A.
    Blondin, D.
    Albers, P.
    Antoch, G.
    Schimmoeller, L.
    EUROPEAN RADIOLOGY, 2019, 29 (12) : 6965 - 6970
  • [4] Comparison of analgesic techniques in MRI-guided in-bore prostate biopsy
    M. Quentin
    C. Arsov
    T. Ullrich
    B. Valentin
    A. Hiester
    D. Blondin
    P. Albers
    G. Antoch
    L. Schimmöller
    European Radiology, 2019, 29 : 6965 - 6970
  • [5] Accuracy study of a robotic system for MRI-guided prostate needle placement
    Seifabadi, Reza
    Cho, Nathan B. J.
    Song, Sang-Eun
    Tokuda, Junichi
    Hata, Nobuhiko
    Tempany, Clare M.
    Fichtinger, Gabor
    Iordachita, Iulian
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2013, 9 (03): : 305 - 316
  • [6] In-bore transrectal MRI-guided prostate biopsies: Are there risk factors for complications?
    Meier-Schroers, Michael
    Homsi, Rami
    Kukuk, Guido
    Wolter, Karsten
    Decker, Georges
    Fischer, Stefan
    Marx, Christian
    Schmeel, Frederic Carsten
    Block, Wolfgang
    Sprinkart, Alois Martin
    Traeber, Frank
    Schild, Hans Heinz
    Willinek, Winfried
    EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (12) : 2169 - 2173
  • [7] Robotic system for MRI-guided prostate biopsy: feasibility of teleoperated needle insertion and ex vivo phantom study
    Seifabadi, Reza
    Song, Sang-Eun
    Krieger, Axel
    Cho, Nathan Bongjoon
    Tokuda, Junichi
    Fichtinger, Gabor
    Iordachita, Iulian
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2012, 7 (02) : 181 - 190
  • [8] Robotic system for MRI-guided prostate biopsy: feasibility of teleoperated needle insertion and ex vivo phantom study
    Reza Seifabadi
    Sang-Eun Song
    Axel Krieger
    Nathan Bongjoon Cho
    Junichi Tokuda
    Gabor Fichtinger
    Iulian Iordachita
    International Journal of Computer Assisted Radiology and Surgery, 2012, 7 : 181 - 190
  • [9] In-bore prostate transperineal interventions with an MRI-guided parallel manipulator: system development and preliminary evaluation
    Eslami, Sohrab
    Shang, Weijian
    Li, Gang
    Patel, Nirav
    Fischer, Gregory S.
    Tokuda, Junichi
    Hata, Nobuhiko
    Tempany, Clare M.
    Iordachita, Iulian
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2016, 12 (02): : 199 - 213
  • [10] MRI-guided in-bore biopsy of the prostate - defining the optimal number of cores needed
    Gross, Moritz
    Eisenhuber, Edith
    Assinger, Petra
    Schima, Raphael
    Susani, Martin
    Doblhammer, Stefan
    Schima, Wolfgang
    CANCER IMAGING, 2024, 24 (01)