A Membrane-Based ELISA Assay and Electrochemical Immunosensor for Microcystin-LR in Water Samples

被引:55
|
作者
Lotierzo, M. [1 ]
Abuknesha, R. [2 ]
Davis, F. [1 ]
Tothill, I. E. [1 ]
机构
[1] Cranfield Univ, Cranfield Hlth, Cranfield MK43 0AL, Beds, England
[2] Kings Coll London, Strand, London WC2R 2LS, England
关键词
PHOSPHATASE INHIBITION ASSAY; LINKED-IMMUNOSORBENT-ASSAY; ENVIRONMENTAL WATER; CYANOBACTERIAL TOXINS; HERBICIDE ISOPROTURON; ENZYME-IMMUNOASSAY; ANTISERUM; BINDING; BLOOMS;
D O I
10.1021/es2041042
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
We describe within this paper the development of an affinity sensor for the detection of the cyanobacterial toxin microcystin-LR The first stage of the work included acquiring and testing of the antibodies to this target. Following the investigation, a heterogeneous direct competitive enzyme-linked immunosorbent assay (ELISA) format for rnicrocystin-LR detection was developed, achieving a detection limit, LLD80 = 0.022 mu g L-1. The system was then transferred to an affinity membrane sorbent-based ELISA. This was an amenable format for immunoassay incorporation into a disposable amperometric immunosensor device. This membrane-based ELISA achieved a detection limit, LLD80 = 0.06 mu g L-1. A three-electrode immunosensor system was fabricated using thick-film screen-printing technology. Amperometric horseradish peroxidase transduction of hydrogen peroxide catalysis, at low reducing potentials, versus Ag/AgCl reference and carbon counter electrodes, was facilitated by hydroquinone-mediated electron transfer. A detection limit of 0.5 mu g L-1 for microcystin-LR was achieved. Similar levels of detection could be obtained using direct electrochemical sensing of the dye produced using the membrane-based ELISA. These techniques proved to be simple, cost-effective, and suitable for the detection of microcystin-LR in buffer and spiked tap and river water samples.
引用
收藏
页码:5504 / 5510
页数:7
相关论文
共 50 条
  • [1] Portable optical immunosensor for highly sensitive detection of microcystin-LR in water samples
    Long, F.
    He, M.
    Zhu, A. N.
    Shi, H. C.
    BIOSENSORS & BIOELECTRONICS, 2009, 24 (08): : 2346 - 2351
  • [2] Direct Electrochemical Nano-immunosensor for Microcystin-LR in Seawater
    Talamini, Lucas
    Zanato, Nicole
    Zapp, Eduardo
    Brondani, Daniela
    Vieira, Iolanda Cruz
    ELECTROANALYSIS, 2018, 30 (05) : 819 - 827
  • [3] Nanobiochar paper based electrochemical immunosensor for fast and ultrasensitive detection of microcystin-LR
    Yao, Lan
    He, Lingzhi
    Yang, Yuesuo
    Zhang, Yu
    Liu, Zairan
    Liang, Lina
    Piao, Yunxian
    SCIENCE OF THE TOTAL ENVIRONMENT, 2021, 750 (750)
  • [4] Microcystin-LR and chemically degraded microcystin-LR electrochemical oxidation
    Lopes, Ilanna C.
    Santos, Paulina V. F.
    Diculescu, Victor C.
    Peixoto, Francisco M. P.
    Araujo, Mario C. U.
    Tanaka, Auro A.
    Oliveira-Brett, Ana Maria
    ANALYST, 2012, 137 (08) : 1904 - 1912
  • [5] Phosphorescent immunosensor for simple and sensitive detection of microcystin-LR in water
    Qin, Jin
    Sun, Xiaojie
    Li, Dongxia
    Yan, Guiqin
    RSC ADVANCES, 2019, 9 (22): : 12747 - 12754
  • [6] Label-Free Immunosensor for the Determination of Microcystin-LR in Water
    Almeida de Oliveira, Ravila
    Zanato, Nicole
    Cruz Vieira, Iolanda
    ELECTROANALYSIS, 2020, 32 (10) : 2166 - 2173
  • [7] Carbon Nanohorn Sensitized Electrochemical Immunosensor for Rapid Detection of Microcystin-LR
    Zhang, Jing
    Lei, Jianping
    Xu, Chuanlai
    Ding, Lin
    Ju, Huangxian
    ANALYTICAL CHEMISTRY, 2010, 82 (03) : 1117 - 1122
  • [8] Electrochemical degradation of microcystin-LR
    Feng, CP
    Sugiura, N
    Masaoka, Y
    Maekawa, T
    JOURNAL OF ENVIRONMENTAL SCIENCE AND HEALTH PART A-TOXIC/HAZARDOUS SUBSTANCES & ENVIRONMENTAL ENGINEERING, 2005, 40 (02): : 453 - 465
  • [9] A graphene and multienzyme functionalized carbon nanosphere-based electrochemical immunosensor for microcystin-LR detection
    Zhao, Huimin
    Tian, Junping
    Quan, Xie
    COLLOIDS AND SURFACES B-BIOINTERFACES, 2013, 103 : 38 - 44
  • [10] Label-Free Electrical Immunosensor for Highly Sensitive and Specific Detection of Microcystin-LR in Water Samples
    Tan, Feng
    Saucedo, Nuvia Maria
    Ramnani, Pankaj
    Mulchandani, Ashok
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2015, 49 (15) : 9256 - 9263