Monte carlo study on dose distributions around 192Ir, 169Yb, and 125I brachytherapy sources using EGSnrc-based egs_brachy user-code

被引:1
|
作者
Mishra, Subhalaxmi [1 ]
Mishra, Bibekananda [2 ]
Selvam, T. Palani [1 ,3 ]
Deshpande, Sudesh [4 ]
Pathan, Munir Shabbir [1 ]
Kumar, Rajesh [1 ]
机构
[1] Bhabha Atom Res Ctr, Div Radiol Phys & Advisory, Hlth Safety & Environm Grp, Mumbai, Maharashtra, India
[2] Atom Energy Regulatory Board, Div Radiol Safety, Mumbai, Maharashtra, India
[3] Homi Bhabha Natl Inst, Mumbai, Maharashtra, India
[4] PD Hinduja Natl Hosp & MRC, Dept Radiat Oncol, Mumbai, Maharashtra, India
关键词
Brachytherapy; egs_brachy; EGSnrc code system; Monte Carlo; TG-43; dosimetry; DOSIMETRY; IR-192; I-125; AAPM; RECOMMENDATIONS; CO-60; EYE;
D O I
10.4103/jmp.jmp_16_22
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction/: As per the recommendations of the American Association of Physicists in Medicine Task Group 43, Monte Carlo (MC) investigators should reproduce previously published dose distributions whenever new features of the code are explored. The purpose of the present study is to benchmark the TG-43 dosimetric parameters calculated using the new MC user-code egs_brachy of EGSnrc code system for three different radionuclides 192Ir, 169Yb, and 125I which represent high-, intermediate-, and low-energy sources, respectively. Materials and Methods: Brachytherapy sources investigated in this study are high-dose rate (HDR) 192Ir VariSource (Model VS2000), 169Yb HDR (Model 4140), and 125I -low-dose-rate (LDR) (Model OcuProsta). The TG-43 dosimetric parameters such as air-kerma strength, Sk, dose rate constant, Lambda, radial dose function, g(r) and anisotropy function, F(r,theta) and two-dimensional (2D) absorbed dose rate data (along-away table) are calculated in a cylindrical water phantom of mass density 0.998 g/cm3 using the MC code egs_brachy. Dimensions of phantom considered for 192Ir VS2000 and 169Yb sources are 80 cm diameter x80 cm height, whereas for 125I OcuProsta source, 30 cm diameter x30 cm height cylindrical water phantom is considered for MC calculations. Results: The dosimetric parameters calculated using egs_brachy are compared against the values published in the literature. The calculated values of dose rate constants from this study agree with the published values within statistical uncertainties for all investigated sources. Good agreement is found between the egs_brachy calculated radial dose functions, g(r), anisotropy functions, and 2D dose rate data with the published values (within 2%) for the same phantom dimensions. For 192Ir VS2000 source, difference of about 28% is observed in g(r) value at 18 cm from the source which is due to differences in the phantom dimensions. Conclusion: The study validates TG-43 dose parameters calculated using egs_brachy for 192Ir, 169Yb, and 125I brachytherapy sources with the values published in the literature.
引用
收藏
页码:270 / 278
页数:9
相关论文
共 13 条
  • [1] Radial dose functions for 103Pd, 125I, 169Yb and 192Ir brachytherapy sources:: an EGS4 Monte Carlo study
    Mainegra, E
    Capote, R
    López, E
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03): : 703 - 717
  • [2] EGSnrc Monte Carlo calculated dosimetry parameters for 192Ir and 169Yb brachytherapy sources
    Taylor, R. E. P.
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2008, 35 (11) : 4933 - 4944
  • [3] Dose rate constants for 125I, 103Pd, 192Ir and 169Yb brachytherapy sources:: an EGS4 Monte Carlo study
    Mainegra, E
    Capote, R
    Lopez, E
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06): : 1557 - 1566
  • [4] Dosimetric Study of HDR 192Ir and LDR 125I Brachytherapy Sources Using the Penelope Monte Carlo Code
    Benhabib, S. M.
    Zhu, D.
    Austerlitz, C.
    Mota, H.
    Sempau, J.
    Allison, R.
    Sibata, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [5] Maximum RBE change in 192Ir, 125I, and 169Yb brachytherapy and the corresponding effect on treatment planning
    Nusrat, Humza
    Karim-Picco, Salesha
    Pang, Geordi
    Paudel, Moti
    Sarfehnia, Arman
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (01)
  • [6] Anisotropy function for 192Ir low-dose-rate brachytherapy sources:: an EGS4 Monte Carlo study
    Capote, R
    Mainegra, E
    López, E
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (05): : 1487 - 1499
  • [7] Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, 192Ir, 103Pd, and 169Yb sources
    Melhus, Christopher S.
    Rivard, Mark J.
    MEDICAL PHYSICS, 2006, 33 (06) : 1729 - 1737
  • [8] Experimental determination of radial dose function and anisotropy function of GammaMed Plus 192Ir high-dose-rate brachytherapy source in a bounded water phantom and its comparison with egs_brachy Monte Carlo simulation
    Buchapudi, Rekha Reddy
    Manickam, Ravikumar
    Chandaraj, Varatharaj
    JOURNAL OF MEDICAL PHYSICS, 2019, 44 (04) : 246 - 253
  • [9] Investigation the effect of a magnetic field on the dose distribution of I-125, Ir-192, Yb-169, and Co-60 brachytherapy sources by Monte Carlo simulation
    Jafarzadeh, Naser
    Hejazi, Payman
    Mansoury, Mohammad Ali Tajik
    Khodabakhshi, Reza
    Riazi, Zafar
    Gholami, Somayeh
    APPLIED RADIATION AND ISOTOPES, 2022, 187
  • [10] Evaluation of 192Ir, 60Co and 169Yb Sources for High Dose Rate Prostate Brachytherapy Inverse Planning Using An Interior Point Constraint Generation Algorithm
    Chung, E. Mok Tsze
    Safigholi, H.
    Nicolae, A.
    Davidson, M.
    Ravi, A.
    Aleman, D.
    Song, W.
    MEDICAL PHYSICS, 2016, 43 (06) : 3464 - 3465