Novel Polymeric Membrane and Coated Wire Electrodes for Determination of Alverine Citrate in Pharmaceutical Formulations and Biological Fluids

被引:0
|
作者
Khalil, M. M. [1 ]
Issa, Y. M. [2 ]
Zayed, S. I. M. [3 ,4 ]
Ali, F. Q. [1 ]
机构
[1] Beni Suef Univ, Fac Sci, Dept Chem, Bani Suwayf, Egypt
[2] Cairo Univ, Fac Sci, Dept Chem, Giza, Egypt
[3] Beni Suef Univ, Fac Ind Educ, Bani Suwayf, Egypt
[4] Taif Univ, Fac Sci, Dept Chem, At Taif, Saudi Arabia
来源
关键词
Alverine citrate; potentiometry; polymeric membrane electrode; coated wire electrode; biological fluids; ION-SELECTIVE ELECTRODES; POTENTIOMETRIC DETERMINATION; SENSORS; HYDROCHLORIDE; CONSTRUCTION;
D O I
暂无
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
Selective and sensitive alverine electrodes have been fabricated by constructing a polymeric membrane (PME) and coated wire (CWE) electrodes. These are based on the ion-pair between phosphotungstic acid (PME1, CWE1) and phosphomolybdic acid (PME2, CWE2) in a poly (vinyl chloride) matrix, plasticized with DBP. The influence of membrane composition and pH on the potentiometric responses of electrodes was investigated. The response characteristics of the membrane were compared with polymeric membrane electrode (PME) as well as with coated wire electrode (CWE). The electrodes exhibit Nernstian slope of 59.1 +/- 0.4, 59.3 +/- 0.6, 59.2 +/- 1.1 and 59.5 +/- 1.1 mV decade(-1) with limits of detection of 7.6x10(-6), 8.3x10(-6), 6.6x10(-6) and 8.1x10(-6) mol L-1 alverine citrate for PME1, PME2, CWE1 and CWE2, respectively. Furthermore, the electrodes generated constant potentials in the pH range of 1.44-8.05 and 1.42-8.31 for PME1 and PME2, respectively. Selectivity coefficients for alverine citrate relative to numbers of potential interfering substances were investigated. The electrodes were highly selective for alverine over a large number of similar compounds. The proposed electrodes displayed useful analytical characteristics for the determination of alverine citrate in bulk powder, pharmaceutical formulation, and biological fluids (urine and plasma). The practical utility of the proposed electrodes has also been demonstrated by thier usage as indicator electrodes in potentiometric titration of phosphotungstic acid with alverine cation solution.
引用
收藏
页码:3442 / 3457
页数:16
相关论文
共 50 条
  • [1] New Coated Wire Sensors for Potentiometric Determination of Gemifloxacin in Pure Form, Pharmaceutical Formulations and Biological Fluids
    Al-Mohaimeed, Amal M.
    Al-Tamimi, Salma A.
    Alarfaj, Nawal A.
    Aly, Fatma A.
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2012, 7 (12): : 12518 - 12530
  • [2] Ion Selective Electrodes for Determination of Cefditoren Pivoxil in Pharmaceutical Formulations and Biological Fluids
    Al-Tamimi, Salma A.
    Al-Mohaimeed, Amal M.
    Alarfaj, Nawal A.
    Aly, Fatma A.
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2013, 8 (03): : 3988 - 4001
  • [3] Membrane electrodes for determination of two antihypertensive drugs in pharmaceutical formulations of either single or binary mixtures and in biological fluids
    El-Ghobashy, Mohamed R.
    Zaazaa, Hala E.
    DRUG DISCOVERIES AND THERAPEUTICS, 2010, 4 (03): : 217 - 222
  • [4] Development of Membrane Electrodes for the Specific Determination of Tetryzoline Hydrochloride in Presence of its Degradation Product in Pharmaceutical Formulations and Biological Fluids
    Hassan, Nagiba Y.
    Lotfy, Hayam M.
    Saleh, Sarah S.
    Salem, Hesham
    ANALYTICAL & BIOANALYTICAL ELECTROCHEMISTRY, 2015, 7 (01): : 75 - 90
  • [5] A Novel Electrochemical Sensor for Determination of Sildenafil Citrate (Viagra) in Pure Form and in Biological and Pharmaceutical Formulations
    Farghali, R. A.
    Ahmed, Rasha A.
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2012, 7 (12): : 13008 - 13019
  • [6] Coated Wire PVC Membrane Sensor for Selective Determination of Copper Ion in Environmental Samples and Pharmaceutical Formulations
    Rizk, Nashwa M. H.
    CURRENT ANALYTICAL CHEMISTRY, 2011, 7 (03) : 235 - 242
  • [7] Construction and Optimization of Selective Membrane Electrodes for Determination of Doxepin Hydrochloride in Pharmaceutical Preparations and Biological Fluids
    El-Tohamy, Maha
    Razeq, Sawsan
    El-Maamly, Magda
    Shalaby, Abdalla
    JOURNAL OF THE KOREAN CHEMICAL SOCIETY-DAEHAN HWAHAK HOE JEE, 2010, 54 (02): : 198 - 207
  • [8] Selective potentiometric determination of zolpidem hemitartrate in tablets and biological fluids by using polymeric membrane electrodes
    Kelani, KM
    JOURNAL OF AOAC INTERNATIONAL, 2004, 87 (06) : 1309 - 1318
  • [9] ELECTROCHEMICAL SENSORS FOR DIRECT DETERMINATION OF SIMVASTATIN IN PHARMACEUTICAL FORMULATIONS AND BIOLOGICAL FLUIDS
    Alarfaj, Nawal A.
    Aly, Fatma A.
    El-Tohamy, Maha
    JOURNAL OF THE CHILEAN CHEMICAL SOCIETY, 2012, 57 (02): : 1140 - 1146
  • [10] NOVEL POLYMERIC MEMBRANE ELECTRODES FOR SELECTIVE DETERMINATION OF TRIMETHOPRIM
    AHMED, MA
    ELBESHLAWY, MM