High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis

被引:141
|
作者
Verhoof, Olaf J. [1 ]
Bron, Johannes L. [1 ]
Wapstra, Frits H. [1 ]
van Royen, Barend J. [1 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Dept Orthopaed Surg, Amsterdam, Netherlands
关键词
lumbar spinal stenosis; X-Stop; degenerative spondylolisthesis;
D O I
10.1007/s00586-007-0492-x
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The X-Stop interspinous distraction device has shown to be an attractive alternative to conventional surgical procedures in the treatment of symptomatic degenerative lumbar spinal stenosis. However, the effectiveness of the X-Stop in symptomatic degenerative lumbar spinal stenosis caused by degenerative spondylolisthesis is not known. A cohort of 12 consecutive patients with symptomatic lumbar spinal stenosis caused by degenerative spondylolisthesis were treated with the X-Stop interspinous distraction device. All patients had low back pain, neurogenic claudication and radiculopathy. Pre-operative radiographs revealed an average slip of 19.6%. MRI of the lumbosacral spine showed a severe stenosis. In ten patients, the X-Stop was placed at the L4 - 5 level, whereas two patients were treated at both, L3 - 4 and L4 - 5 level. The mean follow-up was 30.3 months. In eight patients a complete relief of symptoms was observed post-operatively, whereas the remaining 4 patients experienced no relief of symptoms. Recurrence of pain, neurogenic claudication, and worsening of neurological symptoms was observed in three patients within 24 months. Post-operative radiographs and MRI did not show any changes in the percentage of slip or spinal dimensions. Finally, secondary surgical treatment by decompression with posterolateral fusion was performed in seven patients (58%) within 24 months. In conclusion, the X-Stop interspinous distraction device showed an extremely high failure rate, defined as surgical re-intervention, after short term follow-up in patients with spinal stenosis caused by degenerative spondylolisthesis. We do not recommend the X-Stop for the treatment of spinal stenosis complicating degenerative spondylolisthesis.
引用
收藏
页码:188 / 192
页数:5
相关论文
共 50 条
  • [1] High failure rate of the interspinous distraction device (X-Stop) for the treatment of lumbar spinal stenosis caused by degenerative spondylolisthesis
    Olaf J. Verhoof
    Johannes L. Bron
    Frits H. Wapstra
    Barend J. van Royen
    European Spine Journal, 2008, 17 : 188 - 192
  • [2] Implant Survival Analysis and Failure Modes of the X-Stop Interspinous Distraction Device
    Tuschel, Alexander
    Chavanne, Albert
    Eder, Claudia
    Meissl, Michael
    Becker, Philipp
    Ogon, Michael
    SPINE, 2013, 38 (21) : 1826 - 1831
  • [3] Change of radiological parameters after interspinous implantation (X-stop®) in degenerative spinal stenosis
    Kim H.-Y.
    Choi B.-W.
    European Journal of Orthopaedic Surgery & Traumatology, 2013, 23 (3) : 281 - 285
  • [4] Biomechanical Evaluation of the X-Stop Device for Surgical Treatment of Lumbar Spinal Stenosis
    Wan, Zongmiao
    Wang, Shaobai
    Kozanek, Michal
    Passias, Peter G.
    Mansfield, Frederick L.
    Wood, Kirkham B.
    Li, Guoan
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2012, 25 (07): : 374 - 378
  • [5] Treatment of neurogenic claudication by interspinous decompression: application of the X STOP device in patients with lumbar degenerative spondylolisthesis
    Anderson, Paul A.
    Tribus, Cliff B.
    Kitchel, Scott H.
    JOURNAL OF NEUROSURGERY-SPINE, 2006, 4 (06) : 463 - 471
  • [6] Two-Year Results of X-Stop Interspinous Implant for the Treatment of Lumbar Spinal Stenosis A Prospective Study
    Nandakumar, Anjali
    Clark, Natasha A.
    Smith, Francis W.
    Wardlaw, Douglas
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2013, 26 (01): : 1 - 7
  • [7] Interspinous process decompression with the X-STOP device for lumbar spinal stenosis - A 4-year follow-up study
    Kondrashov, Dimitriy G.
    Hannibal, Matthew
    Hsu, Ken Y.
    Zucherman, James F.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (05): : 323 - 327
  • [8] Effects of X-Stop device on sagittal lumbar spine kinematics in spinal stenosis
    Siddiqui, Manal
    Karadimas, Efthimios
    Nicol, Malcolm
    Smith, Francis W.
    Wardlaw, Douglas
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (05): : 328 - 333
  • [9] Aperius interspinous device for degenerative lumbar spinal stenosis: a review
    Ramesh, Ashwanth
    Lyons, Frank
    Kelleher, Michael
    NEUROSURGICAL REVIEW, 2016, 39 (02) : 197 - 205
  • [10] Aperius interspinous device for degenerative lumbar spinal stenosis: a review
    Ashwanth Ramesh
    Frank Lyons
    Michael Kelleher
    Neurosurgical Review, 2016, 39 : 197 - 205