A grating-based single-shot x-ray phase contrast and diffraction method for in vivo imaging

被引:101
|
作者
Bennett, Eric E. [1 ]
Kopace, Rael [1 ]
Stein, Ashley F. [1 ]
Wen, Han [1 ]
机构
[1] NHLBI, NIH, Imaging Phys Sect, Translat Med Branch, Bethesda, MD 20892 USA
关键词
differential phase contrast; spatial harmonics; dark field; scattering; Fourier analysis; CT; SHEARING INTERFEROMETER; RADIOGRAPHY; TOMOGRAPHY; RADIATION; SPECTRA;
D O I
10.1118/1.3501311
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study is to develop a single-shot version of the grating-based phase contrast x-ray imaging method and demonstrate its capability of in vivo animal imaging. Here, the authors describe the principle and experimental results. They show the source of artifacts in the phase contrast signal and optimal designs that minimize them. They also discuss its current limitations and ways to overcome them. Methods: A single lead grid was inserted midway between an x-ray tube and an x-ray camera in the planar radiography setting. The grid acted as a transmission grating and cast periodic dark fringes on the camera. The camera had sufficient spatial resolution to resolve the fringes. Refraction and diffraction in the imaged object manifested as position shifts and amplitude attenuation of the fringes, respectively. In order to quantify these changes precisely without imposing a fixed geometric relationship between the camera pixel array and the fringes, a spatial harmonic method in the Fourier domain was developed. The level of the differential phase (refraction) contrast as a function of hardware specifications and device geometry was derived and used to guide the optimal placement of the grid and object. Both ex vivo and in vivo images of rodent extremities were collected to demonstrate the capability of the method. The exposure time using a 50 W tube was 28 s. Results: Differential phase contrast images of glass beads acquired at various grid and object positions confirmed theoretical predictions of how phase contrast and extraneous artifacts vary with the device geometry. In anesthetized rats, a single exposure yielded artifact-free images of absorption, differential phase contrast, and diffraction. Differential phase contrast was strongest at bone-soft tissue interfaces, while diffraction was strongest in bone. Conclusions: The spatial harmonic method allowed us to obtain absorption, differential phase contrast, and diffraction images, all from a single raw image and is feasible in live animals. Because the sensitivity of the method scales with the density of the gratings, custom microfabricated gratings should be superior to off-the-shelf lead grids. [DOI: 10.1118/1.3501311]
引用
收藏
页码:6047 / 6054
页数:8
相关论文
共 50 条
  • [1] Single-shot grating-based x-ray differential phase contrast imaging with a modified analyzer grating
    卫晨希
    吴朝
    Faiz Wali
    魏文彬
    鲍园
    骆荣辉
    王磊
    刘刚
    田扬超
    Chinese Physics B, 2017, (10) : 505 - 511
  • [2] Single-shot grating-based x-ray differential phase contrast imaging with a modified analyzer grating
    Wei, Chen-Xi
    Wu, Zhao
    Wali, Faiz
    Wei, Wen-Bin
    Bao, Yuan
    Luo, Rong-Hui
    Wang, Lei
    Liu, Gang
    Tian, Yang-Chao
    CHINESE PHYSICS B, 2017, 26 (10)
  • [3] Single-shot grating-based X-ray phase contrast imaging via generative adversarial network
    Xu, Yueshu
    Tao, Siwei
    Bian, Yinxu
    Bai, Ling
    Tian, Zonghan
    Hao, Xiang
    Kuang, Cuifang
    Liu, Xu
    OPTICS AND LASERS IN ENGINEERING, 2022, 152
  • [4] Noise Reduction for Single-Shot Grating-Based Phase-Contrast Imaging at an X-ray Backlighter
    Schreiner, Stephan
    Akstaller, Bernhard
    Dietrich, Lisa
    Meyer, Pascal
    Neumayer, Paul
    Schuster, Max
    Wolf, Andreas
    Zielbauer, Bernhard
    Ludwig, Veronika
    Michel, Thilo
    Anton, Gisela
    Funk, Stefan
    JOURNAL OF IMAGING, 2021, 7 (09)
  • [5] Single-shot grating-based phase-contrast imaging of a micrometer sample at a laser-driven x-ray backlighter source
    Akstaller, B.
    Schreiner, S.
    Hofmann, F.
    Meyer, P.
    Neumayer, P.
    Schuster, M.
    Wolf, A.
    Zielbauer, B.
    Ludwig, V.
    Michel, T.
    Anton, G.
    Funk, S.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (06)
  • [6] Noise in x-ray grating-based phase-contrast imaging
    Weber, Thomas
    Bartl, Peter
    Bayer, Florian
    Durst, Juergen
    Haas, Wilhelm
    Michel, Thilo
    Ritter, Andre
    Anton, Gisela
    MEDICAL PHYSICS, 2011, 38 (07) : 4133 - 4140
  • [7] High sensitivity phase retrieval method in grating-based x-ray phase contrast imaging
    Wu, Zhao
    Gao, Kun
    Chen, Jian
    Wang, Dajiang
    Wang, Shenghao
    Chen, Heng
    Bao, Yuan
    Shao, Qigang
    Wang, Zhili
    Zhang, Kai
    Zhu, Peiping
    Wu, Ziyu
    MEDICAL PHYSICS, 2015, 42 (02) : 741 - 749
  • [8] Grating-based X-ray phase contrast for biomedical imaging applications
    Pfeiffer, Franz
    Herzen, Julia
    Wiliner, Marian
    Chabior, Michael
    Auweter, Sigrid
    Reiser, Maximilian
    Bamberg, Fabian
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2013, 23 (03): : 176 - 185
  • [9] Grating-Based X-Ray Phase Contrast Imaging With a Polycapillary Semilens
    Tao, Siwei
    Tian, Zonghan
    Bai, Ling
    Xu, Yueshu
    Hao, Xiang
    Kuang, Cuifang
    Yang, Yang Michael
    Liu, Xu
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2022, 34 (20) : 1104 - 1107
  • [10] A single-shot method for X-ray grating interferometry
    Lian, S.
    Kudo, H.
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948