StarClose® vascular closure system (VCS) is safe and effective in patients who ambulate early following successful femoral artery access closureuresults from the RISE clinical trial

被引:17
|
作者
Burke, M. Nicholas [1 ]
Hermiller, James [2 ]
Jaff, Michael R. [3 ,4 ]
机构
[1] Abbott NW Hosp, Minneapolis Heart Inst Fdn, Minneapolis, MN 55407 USA
[2] St Vincents Hosp, Indianapolis, IN USA
[3] Massachusetts Gen Hosp, Dept Med, Div Cardiovasc Med, Boston, MA 02114 USA
[4] Massachusetts Gen Hosp, Vasc Ultrasound Core Lab, Boston, MA 02114 USA
关键词
catheterization; femoral access; vessel closure; vascular complications; ambulation time; CARDIAC-CATHETERIZATION; CORONARY ANGIOPLASTY; COMPLICATIONS; REGISTRY; DEVICES;
D O I
10.1002/ccd.23176
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The RISE study was a prospective, multicenter, single-arm study evaluating the safety and efficacy of the StarClose (R) Vascular Closure System (VCS) in the femoral artery in subjects who were ambulated early following percutaneous diagnostic coronary or peripheral catheterization procedures. Background: Previous studies have demonstrated that several vascular closure devices are able to safely decrease the amount of bedrest needed following angiography. To date, no device has been shown to be safe for immediate ambulation. Methods: Between April 2006 and November 2006, 315 patients undergoing a catheterization via the femoral artery were enrolled (intent-to-treat population, ITT). Of these, 171 subjects were not excluded by angiographic or clinical criteria and had a StarClose device placed to attain hemostasis. If hemostasis was immediately achieved, patients were ambulated within 20 min for 20 feet without assistance (per-protocol population, PP). Patients were followed for 30 days after the index procedure. The primary endpoint of the study was time to ambulation (TTA). Results: Immediate hemostasis was achieved in 156 patients (91.2%). The mean TTA in the PP was 8.29 min, median 5.92. There were no deaths or major vascular complications, and the rate of minor vascular complications was 1.9% (3/156). Conclusion: The StarClose (R) VCS device is both safe and effective allowing immediate ambulation of patients who have undergone catheterization via the femoral artery. (c) 2011 Wiley Periodicals, Inc.
引用
收藏
页码:45 / 52
页数:8
相关论文
共 1 条
  • [1] Results from the prospective, multicenter AMBULATE-CAP trial: Reduced use of urinary catheters and protamine with hemostasis via the mid-bore venous vascular closure system (vascade MVP) following multi-access cardiac ablation procedures
    Al-Ahmad, Amin
    Mittal, Suneet
    DeLurgio, David
    Joseph Gallinghouse, G.
    Horton, Rodney P.
    Preminger, Mark W.
    David Burkhardt, J.
    Natale, Andrea
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (02) : 191 - 199