The effect of crosslinking UHMWPE on in vitro wear rates of fixed and mobile bearing knees

被引:1
|
作者
McNulty, DE [1 ]
Swope, SW [1 ]
Auger, DD [1 ]
Smith, T [1 ]
机构
[1] DePuy Orthopaed Inc, Dept Mat Res, Warsaw, IN 46581 USA
关键词
in vitro wear; mobile bearing knee; fixed bearing knee; crosslinked UHMWPE;
D O I
10.1520/STP11923S
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study evaluated the wear performance of conventional (gamma irradiation at 4Mrads in low oxygen) and moderately crosslinked (gamma irradiation at 5Mrads and melt-annealed) ultra high molecular weight polyethylene (UHMWPE) materials in fixed and mobile knee bearing designs. Kinematic and load inputs corresponding to the gait cycle outlined in the ISO displacement controlled knee simulation draft standard, ISO/CD 14243-3, were maintained in all simulations. Wear rates were determined gravimetrically. Surface finish changes and worn surface characteristics were also noted. Moderately crosslinked fixed bearing knee component wear rate was 74% lower than conventional fixed bearing components. There was no discernable difference in wear rate for the mobile bearing knee for the two materials. However, the mobile bearing knee wear rates were 94% and 79% lower than fixed bearing conventional and fixed bearing moderately crosslinked materials, respectively. In the fixed bearing knee, the majority of the wear reduction of the moderately crosslinked material compared to the conventional material was attributed to enhanced crosslinking due to the melt-annealing process rather than the dose difference. The reduction in wear rate of the mobile bearing knee compared to the fixed bearing was attributed to the reduced cross shear motion at the tibio-femoral articulation since mobile bearing wear was independent of material.
引用
收藏
页码:73 / 85
页数:13
相关论文
共 50 条
  • [1] Particle size and morphology of UHMWPE wear debris in failed total knee arthroplasties - a comparison between mobile bearing and fixed bearing knees
    Huang, CH
    Ho, FY
    Ma, HM
    Yang, CT
    Liau, JJ
    Kao, HC
    Young, TH
    Cheng, CK
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (05) : 1038 - 1041
  • [2] Comparison of wear of mobile and fixed bearing knees tested in a knee simulator
    Johnson, TS
    Laurent, MP
    Yao, JQ
    Blanchard, CR
    WEAR, 2003, 255 (7-12) : 1107 - 1112
  • [3] In vitro wear assessments of fixed and mobile UHMWPE total knee replacement
    Affatato, Saverio
    Bracco, Pierangiola
    Sudanese, Alessandra
    MATERIALS & DESIGN, 2013, 48 : 44 - 51
  • [4] Rotating Platform versus Fixed-bearing Total Knees An In Vitro Study of Wear
    Haider, Hani
    Garvin, Kevin
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (11) : 2677 - 2685
  • [5] Mobile-bearing versus fixed-bearing knees
    Bercovy, M
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (07): : 1113 - 1114
  • [6] Mobile-bearing versus fixed-bearing knees - Reply
    Callaghan, JJ
    Greenwald, AS
    Bourne, RB
    Rorabeck, CH
    Dorr, LD
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (07): : 1114 - 1114
  • [7] Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation
    McEwen, HMJ
    Fisher, J
    Goldsmith, AAJ
    Auger, DD
    Hardaker, C
    Stone, MH
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2001, 12 (10-12) : 1049 - 1052
  • [8] Wear of fixed bearing and rotating platform mobile bearing knees subjected to high levels of internal and external tibial rotation
    H. M. J. McEwen
    J. Fisher
    A. A. J. Goldsmith
    D. D. Auger
    C. Hardaker
    M. H. Stone
    Journal of Materials Science: Materials in Medicine, 2001, 12 : 1049 - 1052
  • [9] Mobile-bearing knees reduce rotational asymmetric wear
    Ho, Fang-Yuan
    Ma, Hon-Ming
    Liau, Jiann-Jong
    Yeh, Chuan-Ren
    Huang, Chun-Hsiung
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (462) : 143 - 149
  • [10] The effect of contact area on wear in relation to fixed bearing and mobile bearing knee replacements
    Sathasivam, S
    Walker, PS
    Campbell, PA
    Rayner, K
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2001, 58 (03): : 282 - 290