Producing parallel x rays with a bent-crystal monochromator and an x-ray tube

被引:8
|
作者
Zhong, Z [1 ]
Dilmanian, FA
Bacarian, T
Zhong, N
Chapman, D
Ren, B
Wu, XY
Weinman, HJ
机构
[1] Brookhaven Natl Lab, Natl Synchrotron Light Source, Upton, NY 11973 USA
[2] Brookhaven Natl Lab, Dept Med, Upton, NY 11973 USA
[3] IIT, CSRRI, Chicago, IL 60616 USA
[4] Schering AG, D-13342 Berlin, Germany
关键词
monochromatic x ray; parallel beam; monochromator; computed tomography;
D O I
10.1118/1.1395024
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A bent Laue monochromator and a conventional x-ray tube were used to produce a fan beam that was parallel in the plane perpendicular to the plane of the fan. The x-ray fan beam was tunable in energy and had about 12% energy bandwidth at a slice height of 5 mm when tuned to 50 keV. The beam's energy was slightly coupled to the vertical position on the beam's height. The slice height could be varied from I to 10 mm. The flux at 50 keV was approximately 2 X 10(6) photons/mm(2)/s with a rotating anode tungsten x-ray tube operating at 120 kVp and 100 mA. The narrow energy bandwidth of the beam produced is advantageous over a conventional divergent polychromatic beam for all radiography applications, while the parallelism of the beam enhances its intensity by about threefold and offers some advantages for computed tomography. (C) 2001 American Association of Physicists in Medicine.