Brain MR Imaging at Ultra-low Radiofrequency Power

被引:21
|
作者
Sarkar, Subhendra N. [1 ]
Alsop, David C. [1 ]
Madhuranthakam, Ananth J. [2 ]
Busse, Reed F. [3 ]
Robson, Philip M. [1 ]
Rofsky, Neil M. [1 ]
Hackney, David B. [1 ]
机构
[1] Harvard Univ, Beth Israel Deaconess Med Ctr, Sch Med, Dept Radiol, Boston, MA 02215 USA
[2] GE Healthcare, Appl Sci Lab, Boston, MA USA
[3] GE Healthcare, Appl Sci Lab, Madison, WI USA
关键词
FAST SPIN-ECHO; FLIP ANGLES; SEQUENCES; STIMULATION; HYPERECHO; IMPLANT; SYSTEMS;
D O I
10.1148/radiol.11092445
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To explore the lower limits for radiofrequency (RF) power-induced specific absorption rate (SAR) achievable at 1.5 T for brain magnetic resonance (MR) imaging without loss of tissue signal or contrast present in high-SAR clinical imaging in order to create a potentially viable MR method at ultra-low RF power to image tissues containing implanted devices. Materials and Methods: An institutional review board-approved HIPAA-compliant prospective MR study design was used, with written informed consent from all subjects prior to MR sessions. Seven healthy subjects were imaged prospectively at 1.5 T with ultra-low-SAR optimized three-dimensional (3D) fast spin-echo (FSE) and fluid-attenuated inversion-recovery (FLAIR) T2-weighted sequences and an ultra-low-SAR 3D spoiled gradient-recalled acquisition in the steady state T1-weighted sequence. Corresponding high-SAR two-dimensional (2D) clinical sequences were also performed. In addition to qualitative comparisons, absolute signal-to-noise ratios (SNRs) and contrast-to-noise ratios (CNRs) for multicoil, parallel imaging acquisitions were generated by using a Monte Carlo method for quantitative comparison between ultra-low-SAR and high-SAR results. Results: There were minor to moderate differences in the absolute tissue SNR and CNR values and in qualitative appearance of brain images obtained by using ultra-low-SAR and high-SAR techniques. High-SAR 2D T2-weighted imaging produced slightly higher SNR, while ultra-low-SAR 3D technique not only produced higher SNR for T1-weighted and FLAIR images but also higher CNRs for all three sequences for most of the brain tissues. Conclusion: The 3D techniques adopted here led to a decrease in the absorbed RF power by two orders of magnitude at 1.5 T, and still the image quality was preserved within clinically acceptable imaging times. (C)RSNA, 2011
引用
收藏
页码:550 / 557
页数:8
相关论文
共 50 条
  • [1] Three-dimensional brain MRI for DBS patients within ultra-low radiofrequency power limits
    Sarkar, Subhendra N.
    Papavassiliou, Efstathios
    Hackney, David B.
    Alsop, David C.
    Shih, Ludy C.
    Madhuranthakam, Ananth J.
    Busse, Reed F.
    La Ruche, Susan
    Bhadelia, Rafeeque A.
    MOVEMENT DISORDERS, 2014, 29 (04) : 546 - 549
  • [2] Ultra-Low Power Transmitter
    Ghasempour, Mohsen
    Shang, Delong
    Xia, Fei
    Yakovlev, Alex
    2012 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS 2012), 2012, : 1807 - 1810
  • [3] Ultra-low power wearables
    Parameshachari, B.D.
    Rocha, Álvaro
    Fung, Chun Che Lance
    Personal and Ubiquitous Computing, 2023, 27 (03) : 1257 - 1259
  • [4] ULTRA-LOW FIELD MR IMAGING OF HEPATIC HEMANGIOMAS (AT 0.02 AND 0.04 T)
    EKELUND, L
    BJORNEBRINK, J
    ATHLIN, L
    EUROPEAN JOURNAL OF RADIOLOGY, 1992, 14 (01) : 41 - 45
  • [5] Statistical analysis of radiofrequency energy harvester with bandpass filter for ultra-low power applications
    Inbaraj, Divya
    Kailasam, Muthumeenakshi
    Sankararajan, Radha
    INTERNATIONAL JOURNAL OF NUMERICAL MODELLING-ELECTRONIC NETWORKS DEVICES AND FIELDS, 2023, 36 (04)
  • [6] Ultra-Low Power and Ultra-Low Voltage Devices and Circuits for IoT Applications
    Hiramoto, T.
    Takeuchi, K.
    Mizutani, T.
    Ueda, A.
    Saraya, T.
    Kobayashi, M.
    Yamamoto, Y.
    Makiyama, H.
    Yamashita, T.
    Oda, H.
    Kamohara, S.
    Sugii, N.
    Yamaguchi, Y.
    2016 IEEE SILICON NANOELECTRONICS WORKSHOP (SNW), 2016, : 146 - 147
  • [7] Ultra-low energy imaging for metrology
    Joy, DC
    METROLOGY, INSPECTION, AND PROCESS CONTROL FOR MICROLITHOGRAPHY XII, 1998, 3332 : 42 - 50
  • [8] Ultra-low voltage ultra-low power NMOS bulk-biased mixer
    School of Electronics and Information Engineering, Beijing University of Aeronautics and Astronautics, Beijing 100191, China
    Beijing Hangkong Hangtian Daxue Xuebao, 2009, 4 (480-484):
  • [9] An ultra-low voltage, ultra-low power fully recycling folded cascode amplifier
    Akbari, Meysam
    Hashemipour, Omid
    Javid, Ardavan
    2014 22nd Iranian Conference on Electrical Engineering (ICEE), 2014, : 514 - 518
  • [10] Ultra-low Power Current Comparator*
    Perumalla, Subrahmanyam
    Mandal, Pradip
    2023 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, ISCAS, 2023,