Design of a Centrifugal Blood Pump: Heart Turcica Centrifugal

被引:10
|
作者
Demir, Onur [1 ]
Biyikli, Emre [1 ]
Lazoglu, Ismail [1 ]
Kucukaksu, Suha [2 ]
机构
[1] Koc Univ, Dept Mech Engn, Mfg & Automat Res Ctr, TR-34450 Istanbul, Turkey
[2] Florence Nightingale Hosp, Dept Cardiovasc Surg, Istanbul, Turkey
关键词
Blade height profile; Centrifugal blood pump; Design; Left ventricular assist device; Volute tongue profile; COMPUTATIONAL FLUID-DYNAMICS; FLOW VISUALIZATION; NUMERICAL-ANALYSIS; IMPELLER DESIGN; HEMOLYSIS TESTS; ESTABLISH;
D O I
10.1111/j.1525-1594.2010.01167.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A prototype of a new implantable centrifugal blood pump system named Heart Turcica Centrifugal (HTC) was developed as a left ventricular assist device (LVAD) for the treatment of end-stage cardiac failure. In the development of HTC, effects of blade height and volute tongue profiles on the hydraulic and hemolytic performances of the pump were investigated. As a result, the prototype was manufactured using the best blade height and volute tongue profiles. Performance of the prototype model was experimentally evaluated in a closed-loop flow system using water as the medium. The hydraulic performance requirement of an LVAD (5 L/min flow rate against a pressure difference of 100 mm Hg) was attained at 2800 rpm rotational speed.
引用
收藏
页码:720 / U138
页数:6
相关论文
共 50 条
  • [1] DEVELOPMENT OF A NEW CENTRIFUGAL LVAD SYSTEM: HEART TURCICA CENTRIFUGAL
    Demir, O.
    Biyikli, E.
    Lazoglu, I.
    Kucukaksu, S.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2010, 33 (07): : 470 - 470
  • [2] Impeller design for a miniaturized centrifugal blood pump
    Takano, T
    Schulte-Eistrup, S
    Yoshikawa, M
    Nakata, K
    Kawahito, S
    Maeda, T
    Nonaka, K
    Linneweber, J
    Glueck, J
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    ARTIFICIAL ORGANS, 2000, 24 (10) : 821 - 825
  • [3] DESIGN OF A CENTRIFUGAL BLOOD PUMP WITH MAGNETIC SUSPENSION
    YAMANE, T
    IKEDA, T
    ORITA, T
    TSUTSUI, T
    JIKUYA, T
    ARTIFICIAL ORGANS, 1995, 19 (07) : 625 - 630
  • [4] EVALUATION OF THE MAGLEV MOTOR AND DESIGN OF THE CENTRIFUGAL BLOOD PUMP FOR PEDIATRIC ARTIFICIAL HEART
    Osa, M.
    Masuzawa, T.
    Tatsumi, E.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (08): : 609 - 609
  • [5] A STRAIGHT PATH CENTRIFUGAL BLOOD PUMP CONCEPT IN THE CAPIOX CENTRIFUGAL PUMP
    KIJIMA, T
    OSHIYAMA, H
    HORIUCHI, K
    NOGAWA, A
    HAMASAKI, H
    AMANO, N
    NOJIRI, C
    FUKASAWA, H
    AKUTSU, T
    ARTIFICIAL ORGANS, 1993, 17 (07) : 593 - 598
  • [6] Design, Manufacturing and Tests of an Implantable Centrifugal Blood Pump
    Bock, Eduardo
    Antunes, Pedro
    Uebelhart, Beatriz
    Leao, Tarcisio
    Fonseca, Jeison
    Cavalheiro, Andre
    Santos Filho, Diolino
    Cardoso, Jose Roberto
    Utiyama, Bruno
    Leme, Juliana
    Silva, Cibele
    Andrade, Aron
    Arruda, Celso
    TECHNOLOGICAL INNOVATION FOR SUSTAINABILITY, 2011, 349 : 410 - +
  • [7] The Design and Evaluation of a Portable Extracorporeal Centrifugal Blood Pump
    Wu, Peng
    Xiang, Wenjing
    Yin, Chengke
    Li, Shu
    FRONTIERS IN PHYSIOLOGY, 2021, 12
  • [8] Design considerations of volute geometry of a centrifugal blood pump
    Chan, WK
    Wong, YW
    Hu, W
    ARTIFICIAL ORGANS, 2005, 29 (12) : 937 - 948
  • [9] Design of a Small Centrifugal Blood Pump With Magnetic Bearings
    Jahanmir, Said
    Hunsberger, Andrew Z.
    Ren, Zhaohui
    Heshmat, Hooshang
    Heshmat, Crystal
    Tomaszewski, Michael J.
    Walton, James F.
    ARTIFICIAL ORGANS, 2009, 33 (09) : 714 - 726
  • [10] Centrifugal blood pump scaling
    Lorenz, M.
    Smith, W.
    ASAIO Journal, 2000, 46 (02)