Measurements and Monte Carlo simulations of the response of the RADOS personal dosemeters with MTS-N (LiF:Mg,Ti) and MCP-N (LiF:Mg,Cu,P) thermoluminescent detectors to X- and gamma-rays

被引:11
|
作者
Obryk, B. [1 ]
Cywicka-Jakiel, T. [1 ]
Budzanowski, M. [1 ]
Bilski, P. [1 ]
Hranitzky, C. [2 ]
Olko, P. [1 ]
Stadtmann, H. [2 ]
机构
[1] Inst Nucl Phys PAN IFJ, PL-31342 Krakow, Poland
[2] Austrian Res Ctr GmbH ARC, A-2444 Seibersdorf, Austria
关键词
thermoluminescence; LiF; detectors; TL dosemeter; Monte Carlo simulation;
D O I
10.1016/j.radmeas.2007.12.041
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The energy and angular response of the MTS-N (LiF:Mg,Ti) and MCP-N (LiF:Mg,Cu,P) TL detectors inside the RADOS personal dosemeters to X-rays photons in the energy range 16-250keV and incidence angles 0 degrees, 30 degrees, 40 degrees, 50 degrees, 60 degrees and (137)CS and (60)Co sources was measured. The irradiations were performed at the Dosimetry Laboratory of the Austrian Research Centers Seibersdorf. The dosemeters were centered on the front surface of the ISO water slab phantom (30 x 30 x 15cm(3)) and it-radiated with reference photon radiation qualities N (the narrow spectrum series), and S-Cs and S-Co nuclide radiation qualities, specified in ISO 4037 (1999). In addition, the energy response to photons of the dosemeter with MCP-N detectors has been calculated using the Monte Carlo transport code MCNP5. The simulation model of the personal dosemeter badge was set up according to the technical drawings. The results of simulations are in good agreement with measurements, except at the lowest (16 keV) photon energy. The results have confirmed that the RADOS personal dosemeter with MTS-N detectors (used by LADIS dosimetry service of the IFJ, Krakow) fulfils the requirements setup in the IEC 61066 (2006) technical recommendations for TL systems for personal dosimetry. These requirements are also fulfilled by the RADOS dosemeter equipped with highly sensitive MCP-N detectors. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:616 / 620
页数:5
相关论文
共 31 条
  • [1] ENERGY RESPONSE OF DIFFERENT TYPES OF RADOS PERSONAL DOSEMETERS WITH MTS-N (LiF:Mg,Ti) AND MCP-N (LiF:Mg,Cu,P) TL DETECTORS
    Obryk, B.
    Hranitzky, C.
    Stadtmann, H.
    Budzanowski, M.
    Olko, P.
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 211 - 214
  • [2] Long-term environment monitoring based on MTS-N (LiF:Mg, Ti) and MCP-N (LiF:Mg, Cu, P) thermoluminescent detectors
    Budzanowski, M
    Olko, P
    Obryk, B
    Ryba, E
    Nowak, A
    RADIATION MEASUREMENTS, 2004, 38 (4-6) : 821 - 824
  • [3] Photon energy response of LiF:Mg,Ti (MTS) and LiF:Mg,Cu,P (MCP) thermoluminescent detectors: Experimental measurements and microdosimetric modeling
    Parisi, Alessio
    Dabin, Jeremie
    Schoonjans, Werner
    Van Hoey, Olivier
    Megret, Patrice
    Vanhavere, Filip
    RADIATION PHYSICS AND CHEMISTRY, 2019, 163 : 67 - 73
  • [4] A kinetic model for the thermoluminescent high dose response of LiF:Mg,Cu,P (MCP-N)
    Benavente, J. F.
    Gomez-Ros, J. M.
    Correcher, V
    APPLIED RADIATION AND ISOTOPES, 2021, 170
  • [5] Dose reassessment by using PTTL method in MTS-N (LiF:Mg, Ti) thermoluminescent detectors
    Budzanowski, M.
    Sas-Bieniarz, A.
    Bilski, P.
    Bubak, A.
    Kopec, R.
    RADIATION MEASUREMENTS, 2013, 56 : 389 - 392
  • [6] Comparison of thermoluminescent readers exploring different reading protocols for LiF:Mg,Cu,P (MCP-N) detectors
    De Saint-Hubert, Marijke
    Knezevic, Zeljka
    Nascimento, Luana De Freitas
    Parisi, Alessio
    Van Hoey, Olivier
    Vanhavere, Filip
    RADIATION MEASUREMENTS, 2019, 121 : 61 - 68
  • [7] Characterisation of thermoluminescent LiF:Mg,Ti (MTS-N) detectors in the 85-550 K temperature range
    Briand, D
    Vincellér, S
    Iacconi, P
    Berkane-Krachai, A
    Bilski, P
    Benabdesselam, M
    Lapraz, D
    RADIATION PROTECTION DOSIMETRY, 2002, 100 (1-4) : 297 - 300
  • [8] Characterisation of thermoluminescent LiF:Mg,Ti (MTS-N) detectors in the 85-550 K temperature range
    Briand, D
    Vincellér, S
    Lacconi, P
    Berkane-Krachai, A
    Bilski, P
    Benabdesselam, M
    Lapraz, D
    SOLID STATE DOSIMETRY, PTS 1 AND 2, PROCEEDINGS, 2002, : 297 - 300
  • [9] Supralinearity of peak 4 and 5 in thermoluminescent lithium fluoride MTS-N (LiF : Mg, Ti) detectors at different Mg and Ti concentration
    Olko, P
    Bilski, P
    Budzanowski, M
    Molokanov, A
    Ochab, E
    Waligorski, MPR
    RADIATION MEASUREMENTS, 2001, 33 (05) : 807 - 812
  • [10] MCP-N (LiF:Mg,Cu,P) TLDs for radon measurements with charcoal canisters
    Bogacz, J
    Mazur, J
    Swakon, J
    Budzanowski, M
    Olko, P
    SOLID STATE DOSIMETRY, PTS 1 AND 2, PROCEEDINGS, 2002, : A267 - A270