Transurethral ultrasound applicators with directional heating patterns for prostate thermal therapy:: In vivo evaluation using magnetic resonance thermometry

被引:104
|
作者
Diederich, CJ [1 ]
Stafford, RJ
Nau, WH
Burdette, EC
Price, RE
Hazle, JD
机构
[1] Univ Calif San Francisco, Thermal Therapy Res Grp, Radiat Oncol Dept, San Francisco, CA 94143 USA
[2] MD Anderson Canc Ctr, Dept Radiol, Houston, TX 77030 USA
[3] Acoust MedSyst Inc, Champaign, IL 61821 USA
关键词
intraluminal; thermal therapy; thermal coagulation; prostate; MR thermal monitoring; ultrasound; hyperthermia;
D O I
10.1118/1.1639959
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A catheter-based transurethral ultrasound applicator with angularly directional heating patterns has been designed for prostate thermal therapy and evaluated in canine prostate in vivo using MRI to monitor and assess performance. The ultrasound transducer array (3.5 mm diameter tubular transducers, 180degrees active sectors, similar to7.5 MHz) was integrated to a flexible delivery catheter (4 mm OD), and encapsulated within an expandable balloon (35 mm X 10 mm OD, 80 ml min(-1) ambient water) for coupling and cooling of the prostatic urethra. These devices were used to thermally coagulate targeted portions of the canine prostate (n = 2) while using MR thermal imaging (MRTI) to monitor the therapy. MRI was also used for target definition, positioning of the applicator, and evaluation of target viability post-therapy. MRTI was based upon the complex phase-difference mapping technique using an interleaved gradient echo-planar imaging sequence with lipid suppression. MRTI derived temperature distributions, thermal dose exposures, T1-contrast enhanced MR images, and histology of sectioned prostates were used to define destroyed tissue zones and characterize the three-dimensional heating patterns. The ultrasound applicators produced similar to180degrees directed zones of thermal coagulation within targeted tissue which extended 15-20 mm radially to the outer boundary of the prostate within 15 min. Transducer activation lengths of 17 mm and 24 mm produced contiguous zones of coagulation extending axially similar to18 rum and similar to25 mm from base to apex, respectively. Peak temperatures around 90 degreesC were measured, with similar to50 degreesC - 52 degreesC corresponding to outer boundary t(43) = 240 min at - 15 min treatment time. These devices are MRI compatible, and when coupled with multiplanar MRTI provide a means for selectively controlling the length and sector angle of therapeutic thermal treatment in the prostate. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:405 / 413
页数:9
相关论文
共 34 条
  • [1] Magnetic resonance guided directional transurethral ultrasound thermal therapy
    Ross, AB
    Diederich, CJ
    Nau, WH
    Tyreus, PD
    Gill, H
    Bouley, D
    Butts, RK
    Rieke, V
    Daniel, B
    Sommer, G
    THERMAL TREATMENT OF TISSUE: ENERGY DELIVERY AND ASSESSMENT II, 2003, 4954 : 192 - 199
  • [2] Combination of implantable and transurethral ultrasound applicators for prostate thermal therapy
    Diederich, CJ
    Nau, WH
    Deardorff, D
    Khalil, IS
    Stauffer, PR
    Burdette, EC
    1997 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 & 2, 1997, : 1337 - 1340
  • [3] A system for MRI-guided thermal therapy of prostate disease with transurethral ultrasound heating applicators
    Chopra, R
    Choy, V
    Baker, S
    Tang, K
    Boyes, A
    Bronskill, M
    MEDICAL PHYSICS, 2005, 32 (07) : 2422 - 2422
  • [4] Development of a system for MRI-guided thermal therapy of prostate disease with transurethral ultrasound heating applicators
    Chopra, R
    Baker, N
    Choy, V
    Bronskill, M
    Thermal Treatment of Tissue: Energy Delivery and Assessment III, 2005, 5698 : 23 - 31
  • [5] Prostate thermal therapy with interstitial and transurethral ultrasound applicators: A feasibility study
    Diederich, CJ
    Nau, WH
    Deardorff, DL
    Bustany, ISK
    Burdette, EC
    Stauffer, PR
    Wu, M
    SURGICAL APPLICATIONS OF ENERGY, PROCEEDINGS OF, 1998, 3249 : 2 - 12
  • [6] Method for MRI-guided conformal thermal therapy of prostate with planar transurethral ultrasound heating applicators
    Chopra, R
    Burtnyk, M
    Haider, MA
    Bronskill, MJ
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21): : 4957 - 4975
  • [7] Transurethral ultrasound applicators with dynamic multi-sector control for prostate thermal therapy:: In vivo evaluation under MR guidance
    Kinsey, Adam M.
    Diederich, Chris J.
    Rieke, Viola
    Nau, William H.
    Pauly, Kim Butts
    Bouley, Donna
    Sommer, Graham
    MEDICAL PHYSICS, 2008, 35 (05) : 2081 - 2093
  • [8] The feasibility of conformal thermal therapy with transurethral ultrasound heating applicators and MR temperature feedback
    Choy, Vanessa
    Tang, Kee
    Wachsmuth, Jeff
    Chopra, Rajiv
    Bronskill, Michael
    THERAPEUTIC ULTRASOUND, 2006, 829 : 146 - +
  • [9] Combination of transurethral and interstitial ultrasound applicators for high-temperature prostate thermal therapy
    Diederich, CJ
    Nau, WH
    Burdette, EC
    Bustany, ISK
    Deardorff, DL
    Stauffer, PR
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2000, 16 (05) : 385 - 403
  • [10] Biothermal modeling of transurethral ultrasound applicators for MR-guided prostate thermal therapy
    Ross, AB
    Diederich, CJ
    Nau, WH
    Tyreus, D
    Gill, H
    Bouley, D
    Butts, RK
    Rieke, V
    Daniel, B
    Sommer, G
    Thermal Treatment of Tissue: Energy Delivery and Assessment III, 2005, 5698 : 220 - 227