Evaporative light scattering detection for microcolumn liquid chromatography

被引:0
|
作者
Alexander, JN [1 ]
机构
[1] Rohm & Haas Co, Spring House, PA 19477 USA
关键词
microcolumn liquid chromatography; packed capillary column; evaporative light scattering detector; micronebulizer; nebulization;
D O I
10.1002/(SICI)1520-667X(1998)10:6<491::AID-MCS4>3.0.CO;2-B
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A concentric micronebulizer for an evaporative light scattering detector (ELSD) in microcolumn liquid chromatography was developed. The micronebulizer uses a bellows regulating valve to control the effluent capillary tip position relative to the nebulization gas. The modified assembly also allows for control of the aerosol flight distance. Both parameters can be optimized without having to stop effluent or nebulization gas flow. The micro-ELSD was evaluated for effluent flow rates ranging from 0.5 to 10 mu L/min under reversed-phase conditions. The detector drift tube can be operated at ambient temperatures under aqueous conditions without a significant decrease in signal-to-noise ratio. In the flow injection mode the limit of detection for glucose is 50 pg. The response of the detector was linear over two different ranges, from 1 to 25 ng and from 35 to 461 ng. In this article, the construction of the micronebulizer is described and the parameters (drift tube temperature, nebulizer gas flow rate, effluent capillary position, aerosol flight distance, effluent flow rate, and mobile phase composition) that influence detector response are studied. (C) 1998 John Wiley & Sons, Inc.