Mercury-plated iridium-based microelectrode arrays for trace metal detection

被引:0
|
作者
Cao, Zhi [1 ,2 ,3 ]
Shang, Haiping [1 ,2 ]
Wang, Yinghui [1 ,2 ,3 ]
Wang, Weibing [1 ,2 ,3 ]
机构
[1] Univ Chinese Acad Sci, Sch Microelect, Beijing, Peoples R China
[2] Chinese Acad Sci, Inst Microelect, Beijing, Peoples R China
[3] Chinese Acad Sci, Kunshan Branch, Inst Microelect, Kunshan, Jiangsu, Peoples R China
关键词
microelectrode array; COMSOL Multiphysics; SEM; AFM; cyclic voltammetry; LINEAR SWEEP VOLTAMMETRY; MICRODISK ELECTRODES;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Trace metals are one of the important components of a marine ecosystem, and their biogeochemical cycles are important for understanding the structure and function of such ecosystems. In this paper, recessed mercury-plated iridium-based microelectrode array sensors are developed for the in-situ detection of trace metals in the ocean. Using COMSOL Multiphysics as a tool simulation and by changing the diffusion layer thickness, the electrode size, and the electrode spacing, we clarified the changes in different types of electrode arrays and determined the exact range of the parameters of the microelectrode array that can avoid the shielding effect. Based on the simulation results, four different parameters of the microelectrode array were successfully fabricated. A scanning electron microscope (SEM) was used to observe the geometry and surface morphology of the microelectrode array. An atomic force microscope (AFM) was used to observe the etched morphology and surface roughness of the microelectrode. The results show that the microelectrode processing technology can meet the production requirements. A cyclic voltammetry test was performed in 1 mmol/L K3Fe(CN)(6). The results show that the diffusion layer thickness is greater than the electrode radius, and the electrode can be considered a microelectrode with a radial diffusion characteristic. Thus, the proposed microelectrode array can be used to detect metals. The accuracy of the simulation is verified by comparing the simulation with the test results. The simulation results can be used to guide the design of microelectrode arrays and the selection of test conditions.
引用
收藏
页码:1102 / 1107
页数:6
相关论文
共 50 条
  • [1] Mercury-plated iridium-based microelectrode arrays for trace metals detection by voltammetry: optimum conditions and reliability
    Belmont, C.
    Tercier, M.-L.
    Buffle, J.
    Fiaccabrino, G. C.
    Analytica Chimica Acta, 329 (03):
  • [2] Mercury-plated iridium-based microelectrode arrays for trace metals detection by voltammetry: Optimum conditions and reliability
    Belmont, C
    Tercier, ML
    Buffle, J
    Fiaccabrino, GC
    KoudelkaHep, M
    ANALYTICA CHIMICA ACTA, 1996, 329 (03) : 203 - 214
  • [3] PREPARATION AND PROPERTIES OF A MERCURY-PLATED IRIDIUM-BASED MICROELECTRODE
    DEVITRE, RR
    TERCIER, ML
    TSACOPOULOS, M
    BUFFLE, J
    ANALYTICA CHIMICA ACTA, 1991, 249 (02) : 419 - 425
  • [4] Gel-Integrated mercury-plated microelectrode arrays for trace metal detection
    Cao, Zhi
    Shang, Haiping
    Wang, Yinghui
    Wang, Shengkai
    Wang, Weibing
    2019 INTERNATIONAL CONFERENCE ON ELECTRONICS PACKAGING (ICEP 2019), 2019, : 279 - 282
  • [5] Gel-Integrated mercury-plated microelectrode arrays for trace metal detection
    Cao, Zhi
    Shang, Haiping
    Wang, Yinghui
    Wang, Shengkai
    Wang, Weibing
    2019 International Conference on Electronics Packaging, ICEP 2019, 2019, : 279 - 282
  • [6] IRIDIUM-BASED MERCURY FILM MICROELECTRODE MODIFIED WITH NAFION
    GOLAS, J
    CHEMIA ANALITYCZNA, 1994, 39 (03): : 359 - 363
  • [7] Mercury-plated thin film ITO microelectrode array for analysis of heavy metals
    Rehacek, V
    Shtereva, K
    Novotny, I
    Tvarozek, V
    Breternitz, V
    Spiess, L
    Knedlik, C
    VACUUM, 2005, 80 (1-3) : 132 - 136
  • [8] Adsorptive stripping analysis of trace nickel at iridium-based ultramicroelectrode arrays
    Wang, J
    Wang, JY
    Adeniyi, WK
    Kounaves, SP
    ELECTROANALYSIS, 2000, 12 (01) : 44 - 47
  • [9] Determination of heavy metals by a mercury-plated diamond-like carbon microelectrode array
    Rehacek, V.
    Hotovy, I.
    Vojs, M.
    PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON NANOSCIENCE AND TECHNOLOGY, 2007, 61 : 982 - 986
  • [10] Shrink-Induced Microelectrode Arrays for Trace Mercury Ions Detection