The Use of Finite Element Method Analysis for Modeling Different Osteotomy Patterns and Biomechanical Analysis of Craniosynostosis Correction

被引:2
|
作者
Augello, Marcello [1 ,2 ]
Maurer, Manfred M. [3 ]
Berg-Boerner, Isabelle [2 ,4 ]
Zeilhofer, Hans-Florian [2 ,4 ]
Cattin, Philippe [5 ]
Juergens, Philipp [2 ,4 ]
机构
[1] Hosp Aarau, Dept Craniomaxillofacial Surg, Aarau, Switzerland
[2] Univ Basel, Hightech Res Ctr Craniomaxillofacial Surg, Allschwil, Switzerland
[3] CADFEM Suisse AG, Aadorf, Switzerland
[4] Univ Hosp, Dept Craniomaxillofacial Surg, Basel, Switzerland
[5] Univ Basel, Dept Biomech Engn, Allschwil, Switzerland
关键词
Craniosynostosis; FEM analysis; skull molding; FRONTOORBITAL ADVANCEMENT; SIMULATION; SURGERY; BONE; POROSITY; CHILDREN; ONLAY; SKULL;
D O I
10.1097/SCS.0000000000005579
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: Several post-processing algorithms for 3D visualization of the skull in craniosynostosis with their specific advantages and disadvantages have been already described. The Finite Element Method (FEM) described herein can also be used to evaluate the efficacy of the cutting patterns with respect to an increase in the projected surface area under assumed uniform loading of the manipulated and cut bone segments. Methods: The FEM analysis was performed. Starting with the classic cranial osteotomies for bifrontal craniotomy and orbital bandeau a virtually mirroring of the unaffected triangular shaped frontal bone was performed to achieve a cup-shaped sphere of constant thickness of 2.5 mm with a radius of 65 mm. Mechanical properties required for the analysis were Young's modulus of 340 MPa and Poisson's ratio of 0.22. Four different cutting patterns from straight to curved geometries have been projected onto the inner surface of the sphere with a cutting depth set to 2/3rds of the shell thickness. The necessary force for the deformation, the resulting tensions and the volume loss due to the osteotomy pattern were measured. Results: Better outcomes were realized with pattern D. The necessary force was 73.6% smaller than the control group with 66N. Best stress distribution was achieved. Curved cutting patterns led to the highest peak of stress and thus to a higher risk of fracture. Straight bone cuts parallel to the corners or to the thighs of the sphere provided a better distribution of stresses with a small area with high stress. Additionally, also with pattern D a surface increase of 20.7% higher than reference was registered. Conclusion: As a proof of concept for different cutting geometries for skull molding in the correction of craniosynostosis, this computational model shows that depending of the cutting pattern different biomechanical behavior is achieved.
引用
收藏
页码:1877 / 1881
页数:5
相关论文
共 50 条
  • [1] Modeling and biomechanical analysis of craniosynostosis correction with the use of finite element method
    Wolanski, Wojciech
    Larysz, Dawid
    Gzik, Marek
    Kawlewska, Edyta
    INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2013, 29 (09) : 916 - 925
  • [2] Biomechanical Comparison between Rotational Scarf Osteotomy and Translational Scarf Osteotomy: A Finite Element Analysis
    Li, Yan
    Wang, Yue
    Wang, Feng
    Tang, Kanglai
    Tao, Xu
    ORTHOPAEDIC SURGERY, 2023, 15 (12) : 3243 - 3253
  • [3] Biomechanical Analysis of Orthodontic Correction of Buccally Malposed Canine: A Finite Element Analysis
    Ali, Muthanna Ibraheem
    Al Sheakli, Iman I.
    Al Zaidee, Salah R.
    JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2021, 9 (02): : 43 - 51
  • [4] Biomechanical assessment of unilateral sagittal split ramus osteotomy: a finite element analysis
    Assari, Ahmad
    Alasseri, Nasser
    Mosadomi, Hezekiah A.
    FRONTIERS OF ORAL AND MAXILLOFACIAL MEDICINE, 2025, 7
  • [5] Optimal untwisting of the orbital bandeau in unicoronal craniosynostosis correction: A finite element analysis
    Winnand, Philipp
    Cevik, Ezgi
    Ooms, Mark
    Heitzer, Marius
    Bock, Anna
    Hoelzle, Frank
    Modabber, Ali
    Raith, Stefan
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2024, 157
  • [6] Finite element modeling and biomechanical analysis of eyeball and extraocular muscles
    Zhou Yiyi
    Wang Guangzhi
    Ding Hui
    Jiao Yonghong
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 4978 - 4981
  • [7] Finite element modeling from medical images for biomechanical analysis
    Chae, SW
    Kwon, GW
    COMPUTATIONAL FLUID AND SOLID MECHANICS 2003, VOLS 1 AND 2, PROCEEDINGS, 2003, : 1638 - 1641
  • [8] Biomechanical analysis of foot with different foot arch heights: a finite element analysis
    Sun, Pi-Chang
    Shih, Shih-Liang
    Chen, Yu-Ling
    Hsu, Yu-Chun
    Yang, Ruei-Cheng
    Chen, Chen-Sheng
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2012, 15 (06) : 563 - 569
  • [9] Optimal hinge level in opening wedge high tibial osteotomy: Biomechanical analysis using finite element method
    Kyung, Min Gyu
    Bae, Tae Soo
    Baek, Hyeong Ho
    Chang, Moon Jong
    Kim, Tae Woo
    Kang, Seung-Baik
    CLINICAL BIOMECHANICS, 2023, 107
  • [10] Biomechanical Effect of Distal Tibial Oblique Osteotomy: A Preliminary Finite-Element Analysis
    Sakai, Tatsuya
    Fujii, Masanori
    Kitamura, Kenji
    Tanaka, Hirofumi
    Mawatari, Masaaki