THE INFLUENCE OF BONE-DENSITY ON IN VIVO K X-RAY FLUORESCENCE BONE-LEAD MEASUREMENTS

被引:2
|
作者
Lodwick, Camille J. [1 ]
Lodwick, Jeffrey C. [2 ]
Spitz, Henry B. [3 ]
机构
[1] Oregon State Univ, Dept Nucl Engn & Radiat Hlth Phys, Corvallis, OR 97331 USA
[2] Occupat Safety & Hlth Adm, US Dept Labor, Salt Lake City, UT 84070 USA
[3] Univ Cincinnati, Dept Mech Ind & Nucl Engn, Cincinnati, OH 45221 USA
来源
HEALTH PHYSICS | 2011年 / 100卷 / 05期
关键词
bones; human; Monte Carlo; Pb-210; x rays; CALIBRATION;
D O I
10.1097/HP.0b013e3181f725af
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Mathematical simulations and benchmark measurements were performed to assess the impact that normal variations in human calcium content have on in vivo K x-ray fluorescence measurements of lead in bone. Four sets of cortical bone tissue simulants were fabricated containing from 20.8% to 23.8% calcium ( by weight) for measurement in a surrogate ( phantom) of the human leg. The net counts detected in the coherent backscatter peak at 88.034 keV using a Cd-109 source indicate a positive trend, with a variability of up to 17% over the range of assessed calcium content. Mathematical simulations confirm this trend and also demonstrate that the contribution of 87.3 keV Pb K-beta 2 counts, which are unresolved in measurements, do not contribute significantly to the coherent peak at low levels of bone-lead content. Both measurements and simulations confirm that calcium is a statistically significant parameter in predicting the K-XRF response and suggest that lead levels may be over-predicted for individuals having low bone density compared to the calibration matrix. Simulations identify a 4.5% negative bias in measured lead values for each 1% increase in calcium weight percent in the bone matrix as compared to the calibration matrix. It is therefore important to accommodate this uncertainty when performing epidemiological studies of populations having a wide range of bone densities. Health Phys. 100(5):502-507; 2011
引用
收藏
页码:502 / 507
页数:6
相关论文
共 50 条
  • [1] In vivo K-shell X-ray fluorescence bone lead measurements in young adults
    Ahmed, N
    Osika, NA
    Wilson, AM
    Fleming, DEB
    JOURNAL OF ENVIRONMENTAL MONITORING, 2005, 7 (05): : 457 - 462
  • [2] Contamination of in vivo bone-lead measurements
    Todd, AC
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (01): : 229 - 240
  • [3] THE ACCURACY OF VOLUMETRIC BONE-DENSITY MEASUREMENTS IN DUAL X-RAY ABSORPTIOMETRY
    SABIN, MA
    BLAKE, GM
    MACLAUGHLINBLACK, SM
    FOGELMAN, I
    CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (03) : 210 - 214
  • [4] Three decades of in vivo x-ray fluorescence of lead in bone
    Chettle, DR
    X-RAY SPECTROMETRY, 2005, 34 (05) : 446 - 450
  • [5] Bone lead concentrations assessed by in vivo X-ray fluorescence
    Ambrose, TM
    Al-Lozi, M
    Scott, MG
    CLINICAL CHEMISTRY, 2000, 46 (08) : 1171 - 1178
  • [6] Calculating the uncertainty in lead concentration for in vivo bone lead X-ray fluorescence
    Todd, AC
    Chettle, DR
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (13): : 2033 - 2039
  • [7] BONE DENSITY MEASUREMENTS IN VIVO - IMPROVEMENT OF X-RAY DENSITOMETRY
    MASON, RL
    RUTHVEN, C
    SCIENCE, 1965, 150 (3693) : 221 - +
  • [8] Uncertainty calculations for the measurement of in vivo bone lead by x-ray fluorescence
    O'Meara, J. M.
    Fleming, D. E. B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (08): : 2449 - 2461
  • [9] Portable X-ray fluorescence for bone lead measurements of Australian eagles
    Hampton, Jordan O.
    Specht, Aaron J.
    Pay, James M.
    Pokras, Mark A.
    Bengsen, Andrew J.
    SCIENCE OF THE TOTAL ENVIRONMENT, 2021, 789
  • [10] An interlaboratory comparison of bone lead measurements via K-shell x-ray fluorescence
    Parsons, P. J.
    Bellis, D. J.
    Hetter, K. M.
    Geraghty, C.
    Berglind, N. A.
    Ginde, N. R.
    Mata, P.
    Todd, A. C.
    X-RAY SPECTROMETRY, 2008, 37 (01) : 76 - 83