Blend uniformity end-point determination using near-infrared spectroscopy and multivariate calibration

被引:52
|
作者
Sulub, Yusuf [1 ]
Konigsberger, Michele [1 ]
Cheney, James [2 ]
机构
[1] Novartis Pharmaceut, Tech Res & Dev, E Hanover, NJ 07936 USA
[2] Celgene Corp, Global Pharmaceut Dev & Technol, Summit, NJ 07901 USA
关键词
Pharmaceutical; Blend uniformity; Near-infrared; Partial least-squares; Off-line calibration; ANALYTICAL TECHNOLOGY APPROACH; PART II; HOMOGENEITY; PREDICTION;
D O I
10.1016/j.jpba.2011.02.017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A multivariate calibration approach using near-infrared (NIR) spectroscopy for determining blend uniformity end-point of a pharmaceutical solid dosage form containing 29.4% (w/w) drug load with three major excipients (crospovidone, lactose, and microcrystalline cellulose) is presented. A set of 21 off-line, static calibration samples were used to develop a multivariate partial least-squares (PLS) calibration model for on-line predictions of the API content during the blending process. The concentrations of the API and the three major excipients were varied randomly to minimize correlations between the components. A micro-electrical-mechanical-system (MEMS) based NIR spectrometer was used for this study. To minimize spectral differences between the static and dynamic measurement modes, the acquired NIR spectra were preprocessed using standard normal variate (SNV) followed by second derivative Savitsky-Golay using 21 points. The performance of the off-line PLS calibration model were evaluated in real-time on 67 production scale (750L bin size) blend experiments conducted over 3 years. The real-time API-NIR (%) predictions of all batches ranged from 93.7% to 104.8% with standard deviation ranging from 0.5% to 1.8%. These results showed the attainment of blend homogeneity and were confirmed with content uniformity by HPLC of respective manufactured tablets values ranging from 95.4% to 101.3% with standard deviation ranging from 0.5% to 2.1%. Furthermore, the performance of the PLS calibration model was evaluated against off-target batches manufactured with high and low amounts of water during the granulation phase of production. This approach affects the particle size and hence blending. All the off-target batches exhibited API-NIR (%) predictions of 94.6% to 103.5% with standard deviation ranging from 0.7% to 1.9%. Using off-target data, a systematic approach was developed to determine blend uniformity end-point. This was confirmed with 3 production scale batches whereby the blend uniformity end-point was determined using the PIS calibration model. Subsequently, the uniformity was also ascertained with conventional thief sampling followed by HPLC analysis and content uniformity by HPLC of the manufactured tablets. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:429 / 434
页数:6
相关论文
共 50 条
  • [1] Modeling strategies for pharmaceutical blend monitoring and end-point determination by near-infrared spectroscopy
    Igne, Benoit
    de Juan, Anna
    Jaumot, Joaquirn
    Lallemand, Jordane
    Preys, Sebastien
    Drennen, James K.
    Anderson, Carl A.
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2014, 473 (1-2) : 219 - 231
  • [2] Determination of the immunoglobulin G precipitation end-point by an intelligent near-infrared spectroscopy system
    Yu, Chen
    Quan, Shuang
    Yang, Cui
    Zhang, Chengliang
    Fan, Jiajin
    Li, Lian
    Zang, Hengchang
    JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2021, 14 (03)
  • [3] End-point determination of the extraction processes for Stevia rebaudiana Bertoni leaves by near-infrared spectroscopy
    Gao, Lele
    Zhong, Liang
    Wei, Yongheng
    Nie, Lei
    Li, Lian
    Dong, Hailing
    Zhang, Hui
    Dong, Qin
    Zang, Hengchang
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2022, 230
  • [4] Validation of model of multivariate calibration: an application to the determination of biodiesel blend levels in diesel by near-infrared spectroscopy
    de Carvalho Rocha, Werickson Fortunato
    Nogueira, Raquel
    Vaz, Boniek Gontigo
    JOURNAL OF CHEMOMETRICS, 2012, 26 (8-9) : 456 - 461
  • [5] Near-infrared multivariate calibration updating using placebo: A content uniformity determination of pharmaceutical tablets
    Sulub, Yusuf
    LoBrutto, Rosario
    Vivilecchia, Richard
    Wabuyele, Busolo
    VIBRATIONAL SPECTROSCOPY, 2008, 46 (02) : 128 - 134
  • [6] Determination of triglycerides in human plasma using near-infrared spectroscopy and multivariate calibration methods
    da Costa, PA
    Poppi, RJ
    ANALYTICA CHIMICA ACTA, 2001, 446 (1-2) : 39 - 47
  • [7] Quantitative determination of epoxidized soybean oil using near-infrared spectroscopy and multivariate calibration
    Parreira, TF
    Ferreira, MMC
    Sales, HJS
    de Almeida, WB
    APPLIED SPECTROSCOPY, 2002, 56 (12) : 1607 - 1614
  • [8] CLASSIFICATION OF BRAZILIAN COFFEE USING NEAR-INFRARED SPECTROSCOPY AND MULTIVARIATE CALIBRATION
    Santos, Klecia M.
    Moura, Maria F. V.
    Azevedo, Francisco G.
    Lima, Kassio M. G.
    Raimundo, Ivo M., Jr.
    Pasquini, C.
    ANALYTICAL LETTERS, 2012, 45 (07) : 774 - 781
  • [9] Detecting Blending End-Point Using Mean Squares Successive Difference Test and Near-Infrared Spectroscopy
    Khorasani, Milad
    Amigo, Jose M.
    Bertelsen, Poul
    Van den Berg, Frans
    Rantanen, Jukka
    JOURNAL OF PHARMACEUTICAL SCIENCES, 2015, 104 (08) : 2541 - 2549
  • [10] A near-infrared spectroscopy-based end-point determination method for the blending process of Dahuang soda tablets
    Wu, Si-jun
    Qiu, Ping
    Li, Pian
    Li, Zheng
    Li, Wen-long
    JOURNAL OF ZHEJIANG UNIVERSITY-SCIENCE B, 2020, 21 (11): : 897 - 910