Tissue-equivalent proportional counters in a high-energy neutron field at CERN

被引:2
|
作者
Lim, T
BottollierDepois, JF
Festag, JG
Golnik, H
Grillmaier, RE
Hofert, M
Lindborg, L
机构
关键词
D O I
10.1117/12.267920
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Within the framework of a research program of the Commission of European Communities (CEC) the reference radiation facility at CERN provided recently a mixed high-energy radiation field simulating the one found at civil aviation altitudes of about 10 km. In order to calibrate radiation detectors and to investigate their response to high-energy neutrons a series of measurements using five different instruments equipped with Low Pressure Tissue-Equivalent Proportional Counter (TEPC) and one Recombination Chamber (REM-2) has been carried out. The reference radiation field at CERN was produced by reactions of a primary beam mixed of 2/3 protons and 1/3 pions with a copper target and moderated by a concrete shielding. The paper summarises the results obtained with the five TEPC instruments and the REM-2, and discusses their characteristics.