Response of LiF:Mg,Ti thermoluminescent dosimeters at photon energies relevant to the dosimetry of brachytherapy (<1 MeV)

被引:39
|
作者
Tedgren, Asa Carlsson [1 ,2 ]
Hedman, Angelica [1 ,3 ]
Grindborg, Jan-Erik [2 ]
Carlsson, Gudrun Alm [1 ]
机构
[1] Linkoping Univ, Fac Hlth Sci, Dept Med & Hlth Sci IMH, SE-58185 Linkoping, Sweden
[2] Swedish Radiat Safety Author, SE-17116 Stockholm, Sweden
[3] Cty Council Ostergotland, Dept Radiat Phys UHL, SE-58185 Linkoping, Sweden
关键词
LiF:Mg; Ti TLD; dosimetry; detector response; efficiency; x-ray energy spectra; brachytherapy; MONTE-CARLO SIMULATIONS; LIF-MG; X-RAYS; TI; SPECTRA; UPDATE;
D O I
10.1118/1.3633892
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: High energy photon beams are used in calibrating dosimeters for use in brachytherapy since absorbed dose to water can be determined accurately and with traceability to primary standards in such beams, using calibrated ion chambers and standard dosimetry protocols. For use in brachytherapy, beam quality correction factors are needed, which include corrections for differences in mass energy absorption properties between water and detector as well as variations in detector response (intrinsic efficiency) with radiation quality, caused by variations in the density of ionization (linear energy transfer (LET) -distributions) along the secondary electron tracks. The aim of this work was to investigate experimentally the detector response of LiF:Mg, Ti thermoluminescent dosimeters (TLD) for photon energies below 1 MeV relative to Co-60 and to address discrepancies between the results found in recent publications of detector response. Methods: LiF:Mg,Ti dosimeters of formulation MTS-N Poland were irradiated to known values of air kerma free-in-air in x-ray beams at tube voltages 25-250 kV, in Cs-137- and Co-60-beams at the Swedish Secondary Standards Dosimetry Laboratory. Conversions from air kerma free-in-air into values of mean absorbed dose in the dosimeters in the actual irradiation geometries were made using EGSnrc Monte Carlo simulations. X-ray energy spectra were measured or calculated for the actual beams. Detector response relative to that for Co-60 was determined at each beam quality. Results: An increase in relative response was seen for all beam qualities ranging from 8% at tube voltage 25 kV (effective energy 13 keV) to 3%-4% at 250 kV (122 keV effective energy) and Cs-137 with a minimum at 80 keV effective energy (tube voltage 180 kV). The variation with effective energy was similar to that reported by Davis [Radiat. Prot. Dosim. 106, 33-43 (2003)] with our values being systematically lower by 2%-4%. Compared to the results by Nunn [Med. Phys. 35, 1861-1869 (2008)], the relative detector response as a function of effective energy differed in both shape and magnitude. This could be explained by the higher maximum read-out temperature (350 degrees C) used by Nunn [Med. Phys. 35, 1861-1869 (2008)], allowing light emitted from high-temperature peaks with a strong LET dependence to be registered. Use of TLD-100 by Davis [Radiat. Prot. Dosim. 106, 33-43 (2003)] with a stronger super-linear dose response compared to MTS-N was identified as causing the lower relative detector response in this work. Conclusions: Both careful dosimetry and strict protocols for handling the TLDs are required to reach solid experimental data on relative detector response. This work confirms older findings that an over-response relative to Co-60 exists for photon energies below 200-300 keV. Comparison with the results from the literature indicates that using similar protocols for annealing and read-out, dosimeters of different makes (TLD-100, MTS-N) differ in relative detector response. Though universality of the results has not been proven and further investigation is needed, it is anticipated that with the use of strict protocols for annealing and read-out, it will be possible to determine correction factors that can be used to reduce uncertainties in dose measurements around brachytherapy sources at photon energies where primary standards for absorbed dose to water are not available. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3633892]
引用
收藏
页码:5539 / 5550
页数:12
相关论文
共 50 条
  • [1] Response of lithium formate EPR dosimeters at photon energies relevant to the dosimetry of brachytherapy
    Adolfsson, Emelie
    Carlsson, Gudrun Alm
    Grindborg, Jan-Erik
    Gustafsson, Hakan
    Lund, Eva
    Tedgren, Asa Carlsson
    MEDICAL PHYSICS, 2010, 37 (09) : 4946 - 4959
  • [2] An Investigation of the Energy Dependence of LiF:Mg,Ti Thermoluminescent Dosimeters in Photon Beams
    Massillon-Jl, G.
    Minniti, R.
    O'Brien, M.
    Soares, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Kinetic parameters of LiF:Mg,Ti thermoluminescent dosimeters
    Sohrabi, M
    Jafarizadeh, M
    Zahedifar, M
    RADIATION PHYSICS AND CHEMISTRY, 1998, 51 (4-6) : 401 - 402
  • [4] The Response of LiF:Mg, Ti Thermoluminescent Dosimeters to Low-Energy Photons
    Rasmussen, B.
    Davis, S.
    Micka, J.
    DeWerd, L.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [5] A comparison of the effectiveness of thermoluminescent crystals LiF:Mg,Ti, and LiF:Mg,Cu,P for clinical dosimetry
    Harris, CK
    Elson, HR
    Lamba, MAS
    Foster, AE
    MEDICAL PHYSICS, 1997, 24 (09) : 1527 - 1529
  • [6] Measurement of the thermoluminescent response (supralinearity and efficiency) of LiF:Mg,Ti exposed to 0.7 MeV protons
    Gamboa-deBuen, I
    Ruiz, C
    Oliver, A
    Lopez, K
    Brandan, ME
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1998, 134 (01): : 136 - 148
  • [7] LIF DOSIMETRY .2. EFFECTS OF MG AND TI ON THERMOLUMINESCENT EMISSION-SPECTRA OF LIF
    CRITTENDEN, GC
    TOWNSEND, PD
    GILKES, J
    WINTERSGILL, MC
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1974, 7 (17) : 2410 - 2421
  • [8] Thermoluminescent response of LiF:Mg,Ti to 20 keV electrons
    Mercado-Uribe, H
    Brandan, ME
    SOLID STATE DOSIMETRY, PTS 1 AND 2, PROCEEDINGS, 2002, : 179 - 182
  • [9] Thermoluminescent response of LiF:Mg,Ti to 20 keV electrons
    Mercado-Uribe, H
    Brandan, ME
    RADIATION PROTECTION DOSIMETRY, 2002, 100 (1-4) : 179 - 182
  • [10] Thermoluminescent response of LiF:Mg,Ti to low energy electrons
    Mercado-Uribe, H
    Brandan, ME
    MEDICAL PHYSICS, 2000, 538 : 235 - 239