Design and performance of a single photon transmission measurement for the ECAT ART

被引:0
|
作者
Watson, CC [1 ]
Jones, WF [1 ]
Brun, T [1 ]
Baker, K [1 ]
Vaigneur, K [1 ]
Young, J [1 ]
机构
[1] CTI PET Syst Inc, Knoxville, TN 37932 USA
关键词
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have implemented dual multi-slit collimator singles transmission sources on the ECAT ART comprising two sets of 12 slits having an axial pitch of twice the axial crystal ring pitch. The slits illuminating opposing detector banks are interleaved, achieving full coverage of all crystal rings as two Cs-137 point sources are simultaneously rastered axially behind the collimators. Transmission data are normalized to a slab phantom scan and corrected for scatter and cross section variation using a log-linear transformation of the attenuation factors. The value of the linear attenuation coefficient in water phantoms has been measured as 0.0946 +/- 0.0044 cm(-l) for 2 min scans employing two 6 mCi sources. A quasi-simultaneous mock scan capability permits direct acquisition and subtraction of the emission background in post-injection transmission acquisitions.
引用
收藏
页码:1366 / 1370
页数:5
相关论文
共 50 条
  • [1] Influence of single-photon-transmission scan duration measured with the ECAT ART PET-scanner
    Schaefer, A
    Kremp, S
    Hellwig, D
    Alexander, C
    Kirsch, CM
    NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2000, 39 (06): : 156 - 165
  • [2] Design and performance of 3D, single photon transmission measurement on a positron tomograph with continuously rotating detectors
    Dekemp, RA
    Jones, WF
    Beanlands, RS
    Nahmias, C
    THREE-DIMENSIONAL IMAGE RECONSTRUCTION IN RADIOLOGY AND NUCLEAR MEDICINE, 1996, 4 : 245 - 254
  • [3] DESIGN AND PERFORMANCE-CHARACTERISTICS OF ECAT POSITRON TOMOGRAPH
    PHELPS, ME
    HOFFMAN, EJ
    HUANG, SC
    KUHL, DE
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1978, 2 (05) : 648 - 648
  • [4] DESIGN AND PERFORMANCE-CHARACTERISTICS OF THE ECAT POSITRON TOMOGRAPH
    PHELPS, MF
    HOFFMAN, EJ
    HUANG, SC
    KUHL, DE
    INTERNATIONAL JOURNAL OF NUCLEAR MEDICINE & BIOLOGY, 1978, 5 (06): : 287 - 287
  • [5] ATTENUATION CORRECTION IN PET USING SINGLE-PHOTON TRANSMISSION MEASUREMENT
    DEKEMP, RA
    NAHMIAS, C
    MEDICAL PHYSICS, 1994, 21 (06) : 771 - 778
  • [6] Direct measurement of the photon statistics of a triggered single photon source -: art. no. 093601
    Treussart, F
    Alléaume, R
    Le Floc'h, V
    Xiao, LT
    Courty, JM
    Roch, JF
    PHYSICAL REVIEW LETTERS, 2002, 89 (09) : 936011 - 936014
  • [7] Room Temperature Single Photon Sources: Design, Performance, and Applications
    Barth, Michael
    Hoeckel, David
    Koch, Lars
    Schietinger, Stefan
    Schroeder, Tim
    Aichele, Thomas
    Benson, Oliver
    ADVANCED PHOTON COUNTING TECHNIQUES IV, 2010, 7681
  • [8] Evaluation of single photon transmission for the HRRT
    Knoess, C
    Rist, J
    Michel, C
    Burbar, Z
    Eriksson, L
    Panin, V
    Byars, L
    Lenox, M
    Wienhard, K
    Heiss, WD
    Nutt, R
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1936 - 1940
  • [9] Photon statistics measurement by use of single photon detection
    XIAO Liantuan
    ChineseScienceBulletin, 2004, (09) : 875 - 878
  • [10] The design of rapid turbidity measurement system based on single photon detection techniques
    Yang, Yixin
    Wang, Huanqin
    Cao, Yangyang
    Gui, Huaqiao
    Liu, Jianguo
    Lu, Liang
    Cao, Huibin
    Yu, Tongzhu
    You, Hui
    OPTICS AND LASER TECHNOLOGY, 2015, 73 : 44 - 49