A new method for the growth of CdTe crystals for RT X-ray photon detectors in the 1-100 keV range

被引:0
|
作者
Lovergine, N [1 ]
Mancini, AM [1 ]
Cola, A [1 ]
Prete, P [1 ]
Mazzer, M [1 ]
Quaranta, F [1 ]
Tapfer, L [1 ]
机构
[1] INFM, Unita Lecce, I-73100 Lecce, Italy
来源
关键词
CdTe; H2T-VPE growth; X-ray detectors;
D O I
10.1142/9789812792013_0073
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We report on the growth of thick CdTe layers on ZnTe/(100)GaAs hybrid substrates by the novel H-2 transport vapour phase epitaxy (H2T-VPE) method. High crystalline quality (100)oriented CdTe single crystal epilayers can be fabricated under atmospheric pressure and at growth temperatures (T-D) in the 600-800 degrees C interval. Double crystal X-ray diffraction measurements performed on epilayers thicker than 30 mu m show CdTe (400) peaks with FWHM < 59 arcsec. CdTe samples grown under optimised conditions have mirror-like surfaces. Epilayers grown below 650 degrees C are p-type and low resistive, but they turn n-type above 650 degrees C, likely as a result of donor diffusion from the substrate. RT resistivities (p)similar to 10(6) Omega-cm are obtained for 675 degrees C < TD < 700 degrees C, but p decreases for higher temperatures and thinner samples. Layers grown under these conditions show RT electron concentrations in the 10(14)-10(11) cm(-3) range. The detection capability of H2T-VPE grown CdTc is demonstrated by the results of time-of-flight measurements performed at RT on Au/n-CdTe/n(+)-GaAs diode structures under reverse bias conditions.
引用
收藏
页码:416 / 421
页数:6
相关论文
共 50 条
  • [1] A novel CdTe/CZT epitaxial technology for large area RT 1-100 keV X-ray photon detectors
    Marzo, F
    Lovergine, N
    Mancini, AM
    Prete, P
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4491 - 4495
  • [2] On hydrogen transport VPE-grown CdTe epilayers for fabrication of 1-100 keV X-ray detectors
    Lovergine, N
    Cola, A
    Prete, P
    Tapfer, L
    Bayhan, M
    Mancini, AM
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 458 (1-2): : 1 - 6
  • [3] CONCEPTUAL DESIGN OF A COMPACT PHOTON SOURCE IN THE 1-100 KEV RANGE
    SILARI, M
    MILAZZO, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1994, 89 (1-4): : 33 - 37
  • [4] Vapor phase epitaxy growth of CdTe epilayers for RT x-ray detectors
    Lovergine, N
    Mancini, AM
    Prete, P
    Cola, A
    Tapfer, L
    HARD X-RAY GAMMA-RAY AND NEUTRON DETECTOR PHYSICS II, 2000, 4141 : 200 - 209
  • [6] Feasibility of intense monochromatic x-ray source in the 1-100 keV range by proton bombardment of metallic targets
    Università degli Studi, Istituto di Fisica Generale Applicata, via Celoria 16, Milan
    20133, Italy
    不详
    20131, Italy
    X-Ray Spectrom., 2 (45-51):
  • [7] FEASIBILITY OF INTENSE MONOCHROMATIC X-RAY SOURCE IN THE 1-100 KEV RANGE BY PROTON-BOMBARDMENT OF METALLIC TARGETS
    CICARDI, C
    MILAZZO, M
    MANFREDI, G
    SILARI, M
    X-RAY SPECTROMETRY, 1995, 24 (02) : 45 - 51
  • [8] A NEW TYPE OF NEUTRON SPECTROMETER IN ENERGY RANGE 1-100 KEV
    MARONI, C
    RUSSO, F
    VERONDINI, E
    NUCLEAR INSTRUMENTS & METHODS, 1969, 74 (02): : 256 - +
  • [9] Modeling CZT/CdTe X-ray Photon-Counting Detectors
    Makeev, Andrey
    Rodrigues, Miesher
    Wang, Gin-Chung
    Glick, Stephen J.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [10] Measuring polarization in the X-ray range: New simulation method for gaseous detectors
    Depaola, G. O.
    Longo, F.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 566 (02): : 590 - 597