Validation of an automated immunoturbidimetric assay for feline serum amyloid A

被引:10
|
作者
Waugh, Elspeth M. [1 ]
Haining, Hayley [1 ]
Harvie, James [1 ]
Ridyard, Alison E. [2 ]
Eckersall, P. David [3 ]
机构
[1] Univ Glasgow, Sch Vet Med, Vet Diagnost Serv, Glasgow, Lanark, Scotland
[2] Univ Glasgow, Sch Vet Med, Small Anim Clin Serv, Glasgow, Lanark, Scotland
[3] Univ Glasgow, Inst Biodivers Anim Hlth & Comparat Med, Glasgow, Lanark, Scotland
关键词
Serum amyloid A; Assay; Validation; Feline; Inflammation; Acute phase protein; SAA TURBIDIMETRIC IMMUNOASSAY; ACUTE-PHASE PROTEINS; A SAA; CANINE; CATS;
D O I
10.1186/s12917-022-03456-5
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Serum Amyloid A (SAA) is a major acute phase protein in cats, increasing rapidly in response to various inflammatory diseases. An automated latex-enhanced immunoturbidimetric assay for human SAA (LZ-SAA, Eiken), previously validated for use in cats, has had further major modification (VET-SAA, Eiken) for specific use in veterinary diagnostic laboratories but has yet to be validated in cats. Results Intra-assay and inter-assay CVs for the VET-SAA assay ranged from 1.88-3.57% and 3.98-6.74%, respectively. Linearity under dilution was acceptable with no prozone effect observed. Limit of detection was 1.65 mg/L and limit of quantification was 6 mg/L. Haemoglobin and triglyceride showed no adverse interference, but bilirubin produced positive bias in samples with low SAA. Comparison with the LZ-SAA assay showed significant correlation with proportional bias increasing as SAA concentration increased, likely related to differing calibration standards. SAA was significantly higher in patients with inflammatory disease compared with non-inflammatory disease, and in patients with moderate to highly elevated alpha 1-AGP compared with patients with normal alpha 1-AGP. Improvement of the assay range may be required to fully evaluate differences between disease groups at low SAA levels. Based on ROC curve analysis, at a cut-off point of 20.1 mg/L the VET-SAA assay discriminated between inflammatory and non-inflammatory disease with sensitivity of 0.93 and specificity of 0.99. Conclusions The automated VET-SAA assay is a robust, precise, and accurate method for measurement of feline SAA which can clearly identify patients with inflammatory disease. It should be a valuable biomarker for use in feline medicine.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Validation of an automated immunoturbidimetric assay for feline serum amyloid A
    Elspeth M. Waugh
    Hayley Haining
    James Harvie
    Alison E. Ridyard
    P. David Eckersall
    BMC Veterinary Research, 18
  • [2] Performance of an automated immunoturbidimetric assay for bovine serum amyloid A
    Nielsen, Lise N.
    Petersen, Mette B.
    Capion, Nynne
    Lundsgaard, Jo F. -H.
    Jensen, Asger L.
    VETERINARY CLINICAL PATHOLOGY, 2024, 53 (02) : 229 - 233
  • [3] Validation of an automated immunoturbidimetric assay for measurement of plasma D-dimer
    Bozic, M
    Stegnar, M
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, 41 (07) : 958 - 962
  • [4] Development of a new prealbumin automated immunoturbidimetric assay for serum and plasma.
    Shiojiri, M
    Adachi, H
    Hanada, T
    Kobatake, S
    Matsuura, S
    Mallory, T
    CLINICAL CHEMISTRY, 2004, 50 (06) : A64 - A64
  • [5] Development and validation of an automated and ultrasensitive immunoturbidimetric assay for C-reactive protein
    Borque, L
    Bellod, L
    Rus, A
    Seco, ML
    Galisteo-González, F
    CLINICAL CHEMISTRY, 2000, 46 (11) : 1839 - 1842
  • [6] Validation of new automated latex-immunoturbidimetric assay for measuring VWF activity
    Rimac, V
    Vojak, S. S.
    Herak, D. C.
    Zadro, R.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2015, 13 : 805 - 805
  • [7] Development of an automated immunoturbidimetric ferritin assay
    Borque, L
    Rus, A
    Bellod, L
    Seco, ML
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 1999, 37 (09) : 899 - 905
  • [8] Preliminary evaluation of an immunoturbidimetric assay and lateral flow device for the measurement of serum amyloid A in rabbits
    Lennox, Angela
    Asahi, Yukari
    Arheart, Kristopher
    Ichiyanagi, Takuo
    Cray, Carolyn
    JOURNAL OF EXOTIC PET MEDICINE, 2020, 33 (0C) : 54 - 56
  • [9] Lipoprotein(a) immunoturbidimetric assay on an routine automated analyzer
    Matsuura, S
    Adachi, H
    Ushio, Y
    Kobatake, S
    Mallory, T
    CLINICAL CHEMISTRY, 1996, 42 (06) : 71 - 71
  • [10] Analytical validation of a point-of-care test and an automated immunoturbidimetric assay for the measurement of canine C-reactive protein in serum
    Covin, Marshal A.
    Gomez, Robynne R.
    Suchodolski, Jan S.
    Steiner, Jorg M.
    Lidbury, Jonathan A.
    CANADIAN JOURNAL OF VETERINARY RESEARCH-REVUE CANADIENNE DE RECHERCHE VETERINAIRE, 2021, 85 (04): : 285 - 292